Description
1. Introduction
The ZOLL AED PLUS is one of the lightest and most proven automated external defibrillators (AEDs) on the market: at just 3.1 kg, with a design that allows any rescuer, even without prior medical training, to intervene safely in the event of cardiac arrest. This unit is supplied as a complete kit with protective case, disposable lithium battery type 123 and 1 pair of adult Stat-Padz II electrodes, the essential equipment to have the defibrillation point operational from the very first moment.
It is the recommended solution for workplaces, schools, gyms, public administrations, residential communities and emergency services that need an accessible, reliable AED with minimal maintenance.
2. Technical explanation of the equipment
The AED PLUS automatically analyses the patient’s cardiac rhythm through the multifunction electrodes and determines whether defibrillation is necessary. Its analysis algorithm evaluates electrode connection and the electrocardiogram, and only recommends a shock for shockable rhythms: ventricular fibrillation with an average amplitude greater than 100 microvolts and wide-complex ventricular tachycardia with rates greater than 150 bpm in adults or 200 bpm in children.
Energy is selected automatically in a pre-programmed manner (120 J, 150 J and 200 J) with a rectilinear biphasic waveform, ZOLL’s proprietary technology that adjusts the current to the patient’s impedance. The unit performs complete automatic checks: battery capacity, electrode connection, electrocardiogram, charge and discharge circuits, microprocessor hardware and software, CPR circuit and CPR-D sensor, and CPR audio circuit. All patient connections are electrically isolated, ensuring the safety of both the rescuer and the patient.
3. Technical Specifications
3.1. General specifications of the unit
| Field | Specification |
|---|---|
| Manufacturer | ZOLL Medical Corporation |
| Set type | Semi-automatic automated external defibrillator (AED) |
| Dimensions (height x width x depth) | 13.3 cm x 24.1 cm x 29.2 cm |
| Weight | 3.1 kg (one of the lightest on the market) |
| Waveform | Biphasic rectilinear |
| Energy selection | Automatic preprogrammed: 120 J, 150 J, 200 J |
| Weight-bearing time | Less than 10 seconds with new batteries |
| Charge retention time | 30 seconds |
| Patient safety | All patient connections electrically isolated |
| Automatic check | Configurable from 1 to 7 days (default value: every 7 days); monthly check at full energy (200 J) |
| Battery included | Disposable lithium battery type 123 (pack of 10 cells, non-rechargeable) – ref. 8000-0807-01 |
| Included electrode | 1 adult pair Stat-Padz II HVP (ref. 8900-0801-01) |
| Kit contents | AED PLUS unit + protective cover + disposable battery + adult Stat-Padz II electrode |
| Presentation | Price per unit of equipment |
3.2. CPR guidance (Real CPR Help, with CPR-D-Padz electrodes)
| Parameter | Value |
|---|---|
| Frequency metronome | Variable between 60 and 100 compressions per minute (CPM) |
| Compression depth | From 3/4” to 3.5”; from 1.9 to 8.9 cm |
| Sensor | CPR-D sensor integrated in the electrode |
| Feedback | Voice prompts to adjust the depth and rate of compressions |
3.3. Shockable rhythms
| Rhythm | Criterion |
|---|---|
| Ventricular fibrillation | Average amplitude greater than 100 microvolts |
| Wide-complex ventricular tachycardia (adults) | Rates greater than 150 bpm |
| Wide-complex ventricular tachycardia (children) | Rates greater than 200 bpm |
3.4. Compatible electrodes (ZOLL family)
| Reference | Product | Related items |
|---|---|---|
| 8900-0801-01 | Stat-Padz II adult electrode (1 pair) | Included in this kit |
| 8900-0800-01 | CPR-D-Padz adult (one-piece) with Real CPR Help | Alternative with integrated CPR sensor |
| 8900-0802-01 | Stat-Padz II adult electrode (box of 12 pairs) | Same family, multipack format |
| 8900-0810-01 | Pedi-Padz II paediatric electrode (1 pair) | Supplementary for children under 8 years of age or under 25 kg |
For information on the sensitivity and specificity of the ECG analysis algorithm, refer to the AED Plus Administrator Guide.
Technical documentation: View AED Plus catalogue | View AED PLUS technical specifications
4. Equivalences and Cross-References
Exact reference equivalence (consumables included): the Stat-Padz II electrode included in this kit corresponds to ZOLL’s official factory construction 8900-0801-01 (Stat-Padz II Electrode, Single), confirmed in ZOLL’s official store (ZOLL Webstore) and through the manufacturer’s authorised distributors. The official battery for the AED PLUS is the pack of 10 lithium type 123 batteries, ref. 8000-0807-01 (with listed alternative 1008-0107-01), also confirmed in official manufacturer sources.
Comparison with other brands: there is no equivalence table validated by ZOLL Medical Corporation that would allow the AED PLUS unit or its electrodes to be replaced by products from another manufacturer: the connector, the analysis algorithm and the Real CPR Help system are proprietary to ZOLL. Any comparison with other leading AEDs on the market (e.g. Philips HeartStart, Stryker/Physio-Control LIFEPAK or Defibtech) should be understood as comparable at category level, not interchangeable 1:1: they differ in waveform, algorithm, electrodes and accessories, although they share the same general clinical indication for cardiopulmonary resuscitation with biphasic defibrillation. Any specific cross-reference must be confirmed with Cahnos’s technical department before use in tenders.
5. Selection tips and variants of interest
- Kit contents: ideal for making the defibrillation point operational instantly; the included Stat-Padz II electrodes cover the adult population and children over 8 years of age (approx. 25 kg).
- Paediatric population: if the environment cares for children (schools, sports centres), add a pair of Pedi-Padz II (8900-0810-01): the AED Plus detects their connection and automatically switches to paediatric energy levels.
- Real-time CPR guidance: if your programme requires compression guidance, consider the CPR-D-Padz electrode (8900-0800-01) with Real CPR Help technology, compatible with this unit.
- Maintenance: replace the electrodes after each use or upon expiry (check the date on the packaging; service life of 18 to 24 months) and replace the battery pack (8000-0807-01) when the unit indicates; the configurable automatic check allows status monitoring without manual intervention.
- Variants of the same unit at Cahnos: SER0A0PL001 (with CPR-D-Padz electrode), SER0A0PL002 (with Stat-Padz II + Pedi-Padz II), SER0A0PL003 (with CPR-D-Padz + Pedi-Padz II) and SER0A0PL000SE (case and battery only, without electrodes).
- Training: for CPR training programmes, the ZOLL AED PLUS TRAINER 2 (8008-0050-10) replicates the unit’s operation without delivering real shocks; always use training electrodes, never clinical ones. For public tenders that require equivalences, request documentary confirmation from our technical department before placing the order.