Description
1. Commercial Introduction
Keep your Schiller family defibrillator always operational with this pair of standard adult defibrillator electrodes with internal cable (reference Cahnos F7956). It is the replacement consumable indicated for defibrillation of adult patients with the FRED Easy, FRED Easy 2G and Defigard DG 4000, DG 5000 and DG 6002 equipment, the AEDs and defibrillators/monitors most widely used in emergency services, ambulances, health centres, companies and heart-safe spaces.
Its configuration with integrated internal cable connects the pair directly and stably to the equipment’s electrode socket, without loose cables during assistance. It is sold as a pair (2 units) in an individual heat-sealed package, with immediate dispatch and specialised technical support from Cahnos to verify compatibility by model and connector before placing the order.
2. Technical explanation of the procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the equipment analyses the cardiac rhythm, delivers the transthoracic shock and enables emergency electrocardiogram monitoring. This pair is indicated for adult patients (over 8 years old or over 25 kg) and connects via the safety anti-shock connector with integrated internal cable.
The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal transmission of the shock energy; the inner metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest, minimising the risk of burns. The recommended placement is anterolateral (sternum-apex), so that both pads do not come into contact with each other. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the package.
3. Technical Specifications
3.1 Product identification
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7956 |
| Product type | Pair of adult multifunction defibrillator electrodes, standard, with internal cable |
| Brand / compatible equipment | Schiller – FRED Easy, FRED Easy 2G, Defigard DG 4000, DG 5000 and DG 6002 |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Presentation | Price per pair (2 units); heat-sealed packaging; 10 pairs per box (adult series) |
3.2 Technical features
| Parameter | Specification |
|---|---|
| Cable configuration | Integrated internal cable (direct connection to the equipment socket) |
| Total surface (per electrode) | 148 cm² |
| Active surface (per electrode) | 95 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks (50 to 360 J) |
| Functions | Defibrillation, synchronized cardioversion and ECG monitoring (max. 24 h) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector, with internal cable |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
3.3 Compatibility with defibrillators
| Equipment brand | Compatible models | Required connection |
|---|---|---|
| Schiller | FRED Easy, FRED Easy 2G, Defigard DG 4000, Defigard DG 5000, Defigard DG 6002 | Safety anti-shock connector; internal cable |
Note: the active surface values (95 cm²), total area (148 cm²), shocks (50–360 J) and adhesive ring (1 cm) correspond to the standard adult electrode series with internal cable of this product family; confirm them with the technical data sheet of the supplied batch and the equipment manual before use in tendering.
4. Equivalences and Cross-References
Reference F7956 is the Cahnos distribution code for this standard adult electrode with internal cable compatible with the Schiller FRED Easy / Defigard platform; it is not an official reference of the equipment manufacturer. The connector and energy management of each defibrillator are proprietary, so equivalences with other references are considered comparable at category level and not interchangeable 1:1. The original factory replacement from Schiller for the FRED Easy / Defigard family is identified by code 0-21-0020 (also 2.230344), consistently confirmed in authorised distribution of original Schiller spare parts; this original comes pre-connected with a 40 cm internal cable and an active surface of approximately 80 cm² per electrode, compared to the 95 cm² with internal cable of this reference.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Schiller (original equipment replacement) | Original pre-connected adult electrodes for FRED Easy, FRED Easy 2G and Defigard DG 4000/5000/6002 | 0-21-0020 (also 2.230344) | Comparable by category (not interchangeable 1:1) | Original factory replacement for the same equipment; it comes pre-connected with a 40 cm internal cable and an approximate active surface of 80 cm² per electrode, compared to the 95 cm² of this reference. Code confirmed in authorised distribution of original Schiller spare parts. |
| ZOLL Medical | Stat-Padz II – adult multifunction electrode (AED Plus / AED Pro) | 8900-0801-01 | Comparable by category (not interchangeable 1:1) | Factory original pre-connected with polymer conductive gel for ZOLL AEDs; proprietary connector, algorithm and energy management. Code confirmed on the official ZOLL website and in GUDID/FDA. Not suitable for Schiller equipment due to proprietary connector. |
| Philips | HeartStart FRx SMART Pads II (adult/child electrodes) | 989803139261 | Comparable by category (not interchangeable 1:1) | Universal electrodes for the FRx family covering adult and paediatrics via the Infant/Child Key; proprietary connector and energy management from Philips. Code confirmed in the official store; does not fit Schiller equipment. |
There is no equivalence table validated by the manufacturer that authorises the replacement of original Schiller electrodes with third-party pads: the connector, analysis algorithm and energy management are proprietary. Any replacement must be validated against the technical data sheet and instructions for use of the defibrillator; consult the technical department of Cahnos before use in tendering.
5. Selection tips and variants of interest
- Confirm the equipment model: verify that your defibrillator is a Schiller FRED Easy, FRED Easy 2G or Defigard DG 4000 / DG 5000 / DG 6002 and that the internal cable connector fits correctly into the equipment’s electrode socket before clinical use; if in doubt, contact the technical department of Cahnos stating the model and serial number.
- Do not confuse versions: this pair is for adult patients (over 8 years / over 25 kg). For paediatric population with the same equipment use reference F7956P (paediatric with internal cable) or F7956PW (paediatric with external cable).
- Internal cable vs. external cable: the internal cable configuration of this reference maintains a direct and clean connection to the equipment; if your workflow requires leaving the pair pre-connected with the connector accessible, consider the variant with external cable (F7956W).
- Standard vs. radiotransparent electrode: if the patient may require X-ray, fluoroscopy or CT during care, consider the radiotransparent series of the same family (F7756, with internal cable) to avoid image artefacts without removing the electrode.
- Check the expiry date: always check the date printed on the package and store in a dry place between +5 °C and +35 °C; do not use expired electrodes or those with damaged packaging, as the gel may lose conductivity.
- Plan your replacement: as it is a single-use consumable, maintain a minimum stock per defibrillator unit and rotate stock according to the expiry date (FIFO method). For volume orders or tenders, contact the technical department of Cahnos indicating the exact model and serial number of your equipment.