Description
1. Commercial Introduction
Keep your defibrillator-monitor always ready to act with this pair of standard adult defibrillator electrodes with internal cable (reference Cahnos F7953). It is the ideal spare and replacement consumable for emergency services, critical care units (ICU), ambulances, hospitals and cardioprotected spaces equipped with the Philips / Agilent Hewlett-Packard and Laerdal Medical families: with a single reference it covers the most widespread defibrillator fleets of these brands (CodeMaster 100, XE, XL, XL+ and HeartStart 4000, as well as XL/XLT and MRx), simplifying the stock of resuscitation consumables.
Its configuration with integrated internal cable protects the connector and the cable inside the heat-sealed bag until the moment of use: simply open the packaging and connect the pair directly to the equipment’s electrode socket, without loose cables during assistance. Price per pair (2 units), individual heat-sealed packaging, immediate dispatch from Cahnos and specialised technical support to verify compatibility by model and connector.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the equipment analyses the cardiac rhythm, delivers the transthoracic shock, allows synchronised cardioversion and monitors the emergency electrocardiogram. This pair is indicated for adult patients (over 8 years old or weighing more than 25 kg) and connects to the electrode socket of compatible equipment via a safety anti-shock connector.
The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal transmission of the shock energy; the inner metal foil distributes the current uniformly and the 1 mm thick medical foam (FOAM) support adapts to the contours of the chest, minimising the risk of burns. The recommended placement is anterolateral (sternum-apex), following the anatomical diagrams printed on the packaging. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the sachet.
3. Technical Specifications
3.1 Technical features
| Reference Cahnos (SKU) | F7953 |
| Product type | Pair of adult multifunction defibrillator electrodes, standard, with internal cable |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Cable configuration | Integrated internal cable (direct connection to the equipment socket) |
| Total surface (per electrode) | 148 cm² |
| Active surface (per electrode) | 95 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks (50 to 360 J) |
| Functions | External defibrillation, synchronised cardioversion and ECG monitoring (max. 24 h) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector with internal cable |
| Radiotransparency | No (standard electrode; for artefact-free imaging there is the radiotransparent series of the family) |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Presentation | Price per pair (2 units); 1 pair in heat-sealed bag; 10 pairs per box in the adult series |
3.2 Compatibility with defibrillators
| Equipment model | Factory reference | Observations |
| CodeMaster 100 | M2475B | Direct connection |
| CodeMaster XE | M1724A | Direct connection |
| CodeMaster XL | M1723A / M1723B | Direct connection |
| CodeMaster XL+ | M1722A / M1722B | Direct connection |
| HeartStart 4000 | M5500B | Energy in manual adjustment |
| HeartStream / HeartStart XL | M4735A | Requires hands-free electrode cable M3508A |
| HeartStream / HeartStart XLT | M3500B | Requires hands-free electrode cable M3508A |
| HeartStart MRx | M3535A / M3536A | Requires cable M3508A; do not use in fully automatic mode |
| Laerdal AED Trainer 2 | — | Direct connection (training) |
Note: with HeartStart 4000, XL/XLT and MRx do not use in fully automatic defibrillators; the energy is adjusted manually. Always verify the correct insertion of the connector into the equipment socket before clinical use.
4. Equivalences and Cross-References
Reference F7953 is the Cahnos distribution code for this standard adult electrode with internal cable for the cited Philips / Agilent Hewlett-Packard and Laerdal Medical defibrillator families; it is not an original certified reference from the equipment manufacturer: the connector, the analysis algorithm and the energy management are proprietary. After verifying official manufacturer catalogues and regulatory databases (GUDID/FDA), the following cross-references with confirmed codes in public sources are identified. In no case does an official equivalence table validated by the manufacturer exist; any comparison between brands must be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Philips Healthcare | Adult Multifunction Pads, IEC barrel connector (platform original) | M3502A (alternative code 989803106931) | Comparable by category (not interchangeable 1:1) | Factory original for the CodeMaster 100 (M2475B), XE (M1724A), XL/XL+ (M1723A/B, M1722A/B) and HeartStart 4000 (M5500B) families via M3507A cable. Code confirmed in authorised technical distribution and in the device regulatory database (FDA authorisation K991871). It is not radiotransparent; verify connector, surface and cable before replacement. |
| Leonhard Lang | Skintact® Easibeat – adult multifunction electrodes Philips CodeMaster | DF26N | Comparable by category (not interchangeable 1:1) | Adult electrode with connector compatible with CodeMaster XE, XL, XL+ and 100; 1 pair per sachet and 10 pairs per box; superimposed gel technology. Code confirmed in the Leonhard Lang technical catalogue and in authorised distribution; with M3507A adapter cable it also covers HeartStart MRx, XL, XL+ and Efficia DFM100. Verify connector fit before replacement. |
| Cardinal Health (Covidien) | Medi-Trace Cadence adult RTS (radiotransparent) | 22661R (preconnected: 22661PC) | Comparable by category (not interchangeable 1:1) | Radiotransparent adult pair with barrel connector compatible with the Philips / Agilent-HP family (CodeMaster, HeartStream/HeartStart XL, XL+, MRx and HeartStart 4000); silver/silver chloride gradient technology and 10 pairs per box. Code confirmed in GUDID (DI 20884527022370) and manufacturer catalogue. Differs in radiotransparency, cable and geometry; verify connector and surfaces. |
| Nissha Medical Technologies (HeartSync) | 97-DEFIB F/G C100AC-PHILIPS – adult pair compatible with Philips | 6600015H (ref. C100AC-PHILIPS) | Comparable by category (not interchangeable 1:1) | Radiotransparent adult pair with internal cable (leads-in) and connector comparable to the original Philips; total surface area of 371 cm², carbon conductive element, 1.5 m cable and 36 months of service life. Code confirmed in the manufacturer’s official catalogue. Not suitable for all models: verify connector and compatibility with the exact equipment. |
| Philips Healthcare | HeartStart Adult Defibrillator Pads (AED FR2/FR2+ / ForeRunner) | 989803158211 (1 pair) / 989803158221 (5 pairs) | Comparable by category (not interchangeable 1:1) | Original for the AEDs of the HeartStart FR2/FR2+ and ForeRunner family, indicated for patients aged 8 years or more / 25 kg or more; code confirmed on the official Philips website and in GUDID (DI 00884838023710 / 00884838126299). Not radiotransparent; connector and algorithm proprietary to the same AED platform. |
There is no equivalence table validated by the manufacturer (Philips / Agilent-HP / Laerdal Medical) that authorises the replacement of the original electrodes with third-party pads: the connector, the analysis algorithm and energy management are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and IFU, checking one by one the equipment connector, energy range, target population and packaging expiry. Consult the technical department at Cahnos before use in tenders.
5. Selection tips and variants of interest
Confirm the model and connector of the equipment: check the exact model on the defibrillator nameplate (CodeMaster 100/XE/XL/XL+, HeartStart 4000, XL/XLT, MRx) and the required therapy cable (M3507A / M3508A on models that require it) before storing the spare part.
Do not confuse versions: this pair is for adult patients (over 8 years / more than 25 kg). For the paediatric population with the same equipment, use the paediatric reference with internal cable (F7953P) or with external cable (F7953PW).
Internal cable vs external cable: this variant keeps the connector protected inside the packaging until the moment of opening, ensuring a clean and stable connection; if your workflow requires leaving the pair pre-connected and ready in seconds, the external cable version of the same platform is available (F7953W).
Standard electrode vs radiolucent: this reference is not radiolucent. If the patient may require X-ray, fluoroscopy or CT during care, consider the radiolucent series of the same family (F7753 with internal cable or F7753W with external cable), which avoids image artefacts without removing the electrode.
Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the packaging expiry date; the conductive gel may dry out and reduce its conductivity after the expiry date. As it is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to expiry (FIFO method). For volume orders or tenders, contact the technical-commercial department at Cahnos indicating the exact model and serial number of your defibrillator.