Description
1. Commercial Introduction
Keep your defibrillator-monitor always ready to act with this pair of multifunction defibrillation electrodes for adult, radiotransparent and with external cable (lead-out), reference FIAB F7753W. It is the ideal replacement and replenishment consumable for emergency services, critical care units (ICU), ambulances, operating theatres, haemodynamics and cardioprotected areas equipped with devices from the Agilent Hewlett-Packard and Laerdal Medical families: CodeMaster 100, XE, XL, XL+ and HeartStart 4000, as well as the XL/XLT and MRx series with the corresponding therapy cables. Manufactured by FIAB SpA (Vicchio, Florence, Italy), a leading European manufacturer of defibrillation electrodes.
Its configuration with external cable (cable outside the bag, approximately 120 cm) allows the pair to be pre-connected to the device and ready in seconds, without opening or handling the pouch at the critical moment. It is sold in full boxes and on quotation: contact the technical-commercial department of Cahnos for conditions, availability and verification of compatibility by model and connector.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronised cardioversion and monitors the emergency electrocardiogram. This pair is indicated for adult patients (over 8 years old or weighing more than 25 kg) and connects via a safety anti-shock connector to the electrode socket of compatible devices.
The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal transmission of the shock energy; the inner metal foil distributes the current uniformly and the medical foam (FOAM) backing, 1 mm thick, adapts to the contours of the thorax, minimising the risk of burns. Its radiotransparent condition —in the electrode and the cable— avoids artefacts in radiography, radioscopy and tomography, so the pads can remain in place during imaging without interrupting monitoring or therapy. The recommended placement is anterolateral (sternum-apex), following the anatomical diagrams on the packaging. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the pouch.
3. Technical Specifications
3.1 Manufacturer’s specifications
| Characteristic | Value |
|---|---|
| Reference FIAB | F7753W |
| Packaging | 10 pairs per box |
| Units per box (web) | 10 |
3.2 Technical features of the series
Main features of the series of adult radiotransparent electrodes with external cable to which this reference belongs:
| Parameter | Specification |
|---|---|
| Reference FIAB (SKU) | F7753W |
| Product type | Pair of multifunction defibrillation electrodes for adult, radiotransparent, with external cable (lead-out) |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Radiotransparency | Yes (electrode and cable), suitable for imaging procedures |
| Cable configuration | External cable (pre-connectable / lead-out), approximate length 120 cm |
| Total surface per electrode | 136 cm² (adult series) |
| Active surface per electrode | 94 cm² (adult series) |
| Adhesive ring width | 0.7 cm |
| Supported shocks | Up to 50 shocks at 360 J |
| Functions | External defibrillation, synchronised cardioversion and ECG monitoring |
| ECG monitoring | Up to 24 hours |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector compatible with Agilent HP / Laerdal equipment |
| Use | Single-use, disposable; latex-free and phthalate-free |
| Primary packaging | 1 pair in heat-sealed bag; 10 pairs per box (adult series) |
| Shelf life | Approx. 30 months from production date (see expiry on the packaging) |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
Note: the values for active surface (94 cm²), total area (136 cm²), discharges (50 to 360 J) and adhesive ring (0.7 cm) correspond to the series of adult radiotransparent electrodes with external cable of this product family; confirm with the technical data sheet of the supplied batch and the device manual before use in tendering.
3.3 Compatibility with defibrillators
| Equipment model | Factory reference | Observations |
|---|---|---|
| CodeMaster 100 | M2475B | Direct connection |
| CodeMaster XE | M1724A | Direct connection |
| CodeMaster XL | M1723A / M1723B | Direct connection |
| CodeMaster XL+ | M1722A / M1722B | Direct connection |
| HeartStart 4000 | M5500B | Energy in manual adjustment |
| HeartStream / HeartStart XL | M4735A | Requires hands-free electrode cable M3508A |
| HeartStream / HeartStart XLT | M3500B | Requires hands-free electrode cable M3508A |
| HeartStart MRx | M3535A / M3536A | Requires cable M3508A; do not use in fully automatic mode |
| Laerdal AED Trainer 2 | — | Direct connection (training) |
Note: with HeartStart 4000, XL/XLT and MRx do not use in fully automatic defibrillators; the energy is adjusted manually. Always verify the correct insertion of the connector into the device socket before clinical use.
4. Equivalences and Cross-References
Reference F7753W is a consumable compatible with the cited Agilent Hewlett-Packard and Laerdal Medical defibrillators, not an original spare part certified by each device manufacturer: the connector, the analysis algorithm and the energy management are proprietary. The following cross-references from other manufacturers are identified as comparable at category level (same barrel connector platform), with code confirmed in public sources; in no case is there an official equivalence table validated by the manufacturer and they should not be considered 1:1 interchangeable.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level |
|---|---|---|---|
| Philips Healthcare | Adult Multifunction Pads, IEC barrel connector (platform original) | M3502A (alternative 989803106931) | Comparable by category (not interchangeable 1:1) |
| Philips Healthcare | HeartStart Adult Defibrillator Pads (AED FR2/FR2+ / ForeRunner) | 989803158211 (1 pair) / 989803158221 (5 pairs) | Comparable by category (not interchangeable 1:1) |
| Leonhard Lang | Skintact® Easibeat – adult multifunction electrodes Philips CodeMaster | DF26N | Comparable by category (not interchangeable 1:1) |
| Cardinal Health (Covidien) | Medi-Trace Cadence adult RTS (radiotransparent) | 22661R (preconnected: 22661PC) | Comparable by category (not interchangeable 1:1) |
| Nissha Medical Technologies (HeartSync) | 97-DEFIB F/G C100AC-PHILIPS – adult pair compatible with Philips | 6600015H (ref. C100AC-PHILIPS) | Comparable by category (not interchangeable 1:1) |
There is no equivalence table validated by the manufacturer (Agilent-HP / Laerdal Medical) that authorises the replacement of original electrodes with third-party pads: the connector, the analysis algorithm and the energy management are proprietary. Any replacement must be validated against the technical data sheet and instructions for use (IFU) of the defibrillator, checking one by one the device connector, the energy range, the target population and the expiry date of the packaging. Consult the technical department of Cahnos before use in tendering.
5. Selection tips and variants of interest
Confirm the model and connector of your device: check on the defibrillator plate the exact model (CodeMaster 100/XE/XL/XL+, HeartStart 4000, XL/XLT, MRx) and the required therapy cable (M3507A / M3508A on models that require it) before storing the spare part.
External cable vs internal cable: this variant (F7753W) leaves the connector accessible outside the bag to have the equipment pre-connected and ready in seconds (ideal in ambulances and cardio-protected spaces); if you prefer a placement without loose cables, the radiotransparent version with internal cable of the same platform (F7753) is available.
Choose radiotransparent when imaging tests are anticipated: if the patient may require X-ray, fluoroscopy or CT during care, this series avoids artefacts without removing the electrode. If imaging is not required, the standard non-radiotransparent version with external cable from the same family (F7953W) may be considered.
Paediatric population: for patients under 8 years of age or weighing less than 25 kg, use the radiotransparent paediatric version of the same platform (F7753P).
Other platforms in the radiotransparent category with external cable: F7750W (Philips), F7751W (ZOLL Medical / M&D), F7752W (Physio-Control / Bexen / Mindray) and F7755W (Nihon Kohden / Welch Allyn / Dräger / MRL), not interchangeable with each other due to proprietary connector.
Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the expiry date on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to expiry (FIFO method). For volume orders or tenders, contact the technical-commercial department of Cahnos indicating the exact model and serial number of your defibrillator.





Reviews
There are no reviews yet.