1. Commercial Introduction
Keep your ZOLL defibrillator always operational with this pair of adult radiotransparent defibrillator electrodes with internal cable (reference Cahnos F7751), a high-reliability consumable compatible with ZOLL Medical equipment from the PD and E series, as well as with the M&D AED 7000 AED. Designed for emergency departments, ICUs, medical transport, operating theatres and cardioprotected sites, it combines the advantages of the radiotransparent series —imaging without artefacts— with the robustness of the integrated internal lead connection, which prevents loose cables during resuscitation. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronised cardioversion, non-invasive transcutaneous pacing (external pacemaker) and emergency ECG monitoring. This adult model (patients over 8 years of age or weighing more than 25 kg) is constructed with low-impedance conductive adhesive gel, ensuring homogeneous skin contact and optimal energy transmission; the inner metal foil distributes the current evenly and the 1 mm medical foam backing conforms to the contours of the chest.
Their radiotransparent condition —in the electrode and in the cable— prevents artefacts in radiography, fluoroscopy and tomography, so the pads can remain in place during imaging examinations without interrupting monitoring or therapy. The integrated internal lead (leads-in) configuration of 100 cm provides a stable, direct connection to the device socket. They are single-use: they must be replaced after each procedure or upon reaching the expiry date printed on the packaging, and placed in the anterolateral position (sternum-apex) according to the anatomical diagrams on the pouch.
3. Technical specifications and compatibility
3.1 Technical specifications
Main features of the series of adult radiotransparent electrodes with internal lead to which this reference belongs:
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7751 |
| Product type | Pair of adult multifunction defibrillator electrodes, radiotransparent, with internal cable |
| Compatible brand | ZOLL Medical (proprietary connector) and M&D |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Radiotransparency | Yes (electrode and cable), suitable for imaging procedures |
| Cable configuration | Integrated internal cable (leads-in), 100 cm |
| Functions | Defibrillation, synchronized cardioversion, transcutaneous pacing (pacemaker) and ECG monitoring |
| Total surface per electrode | 136 cm² |
| Active surface per electrode | 94 cm² |
| Adhesive ring width | 0.7 cm |
| Supported shocks | Up to 50 shocks at 360 J |
| ECG monitoring | Max. 24 h |
| Conductive gel | Low-impedance conductive adhesive gel |
| Backing material | Medical foam (1 mm) with conductive metal foil |
| Connector | Specific connector for ZOLL equipment (verify correct fit in the device socket) |
| Use | Single-use, disposable; non-sterile; latex-free and phthalate-free |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 10 pairs per box |
| Shelf life | Approx. 30 months from date of manufacture (check printed expiry date) |
| Conditions of use and storage | Operating temperature from 0 °C to +50 °C; store in a dry place between +5 °C and +35 °C |
| Classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
Note: the values for active surface area (94 cm²), total area (136 cm²), shocks (50 to 360 J) and adhesive ring (0.7 cm) correspond to the series of adult radiotransparent electrodes with internal lead in this product family; confirm them with the technical data sheet of the supplied batch and the device manual before use in tendering.
3.2 Compatibility with defibrillators
| Equipment brand | Compatible models |
|---|---|
| ZOLL Medical | PD 1200, PD 1400, PD 1600, PD 1700, PD 2000 |
| ZOLL Medical | E Series |
| M&D | AED 7000 |
Note: according to the manufacturer’s documentation for the original ZOLL electrodes, manual mode must be used for energy adjustment with M Series, E Series and R Series equipment. Compatibility refers to the connector platform of each family; always verify the correct fit of the connector and cable of your device before clinical use.
4. Equivalences and Cross-References
Reference F7751 is a consumable compatible with the cited defibrillators, not an original spare part certified by the device manufacturer: the connector, analysis algorithm and energy management are proprietary. After verifying official manufacturer catalogues and regulatory databases (GUDID), the following cross-references with codes confirmed in public sources have been identified. Any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| ZOLL Medical | Radiotransparent Defibrillation Electrode, adult (original equipment) | 8900-004041-01 (preconnected: 8900-004045-01) | Comparable by category (not interchangeable 1:1) | Factory original radiotransparent multifunction for ZOLL R Series and X Series equipment; code confirmed in the official ZOLL store and in authorised distribution. Proprietary connector from the same platform; verify fit before replacement. |
| ZOLL Medical | Stat-Padz® II — adult multifunction electrodes (original equipment) | 8900-0802-01 | Comparable by category (not interchangeable 1:1) | Factory original for ZOLL AED Plus / AED Pro / AED 3 and M, E, R and X Series equipment; code confirmed in the official ZOLL store, in the manufacturer’s technical data sheet and in GUDID. It is not radiotransparent and the cable is inside the packaging; polymer gel, labelling and packaging proprietary to ZOLL. |
| CONMED | PadPro® RTS — adult multifunction radiotransparent electrode, ZOLL connector | 2001Z (pre-connected: 2001Z-PC) | Comparable by category (not interchangeable 1:1) | Code confirmed in authorised European and American technical distribution: adult radiotransparent with ZOLL connector, up to 50 shocks / 8 h of continuous pacing / 24 h of monitoring, 36 months of service life. Geometry, gel and labelling of CONMED’s own; same ZOLL connector platform as this reference. Verify fit before replacement. |
| Leonhard Lang | Skintact® Easibeat — adult multifunction electrode, ZOLL connector | DF28N (pre-connected: DF28NC) | Comparable by category (not interchangeable 1:1) | Code confirmed in the official Leonhard Lang catalogue and in authorised distribution: multifunction (defibrillation, pacing, cardioversion and monitoring) for ZOLL E Series, M Series, R Series, PD 1200, PD 1400, PD 1600, PD 1700 and PD 2000; 1 pair/pack, 10 packs/box. Connector, gel and materials of their own; verify the radiotransparent nature of the specific variant and connector fit before replacement. |
For the M&D AED 7000 platform, no verifiable original electrode code has been found in public sources, so its equivalents remain pending confirmation with the manufacturer. There is no manufacturer-validated equivalence table authorising the replacement of original electrodes with third-party pads: the connector, analysis algorithm and energy management of each defibrillator are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and IFU; consult Cahnos’s technical department before use in tendering.
5. Selection tips and variants of interest
Confirm the model and connector of your device: this pair corresponds to ZOLL defibrillators and monitors PD 1200, PD 1400, PD 1600, PD 1700, PD 2000 and E Series, as well as to the M&D AED 7000 AED. Verify the reference and expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after a procedure.
Choose the variant according to your workflow: the internal cable of this model provides a clean and robust connection during resuscitation; if you prefer to leave the equipment pre-connected and ready in seconds, you have the reference F7751W (same adult radiotransparent series with external cable). For standard use without radiotransparency requirement, consider F7951 (standard adult with internal cable) or F7951W (standard adult with external cable), at a lower cost. If you treat paediatric patients, you have the paediatric radiotransparent references F7751P (internal cable) and F7751PW (external cable), as well as the standard F7951P and F7951PW.
Storage and expiry: store in a dry place between +5 °C and +35 °C and observe the expiry date on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As it is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO). Other platforms in the radiotransparent category with internal cable (F7750 Philips, F7752 Physio-Control/Mindray, F7755 Nihon Kohden/Welch Allyn and F7756 Schiller) are not interchangeable with each other due to proprietary connector. For volume orders or tenders, contact the technical department of Cahnos indicating the exact model and serial number of your equipment.