Description
1. Commercial Introduction
The radiotransparent multifunction electrode F 50 S BLUE (reference Cahnos 15843) is a disposable ECG electrode made of circular foam, 50 mm in diameter, with solid conductive gel and a radiotransparent snap connector, designed to provide reliable and stable signal transmission in both continuous monitoring and conventional electrocardiograms. Its radiotransparent nature makes it the reference consumable for cardiology departments, ICUs, operating theatres, recovery rooms and MRI units, where it can remain in place during imaging without interrupting patient monitoring and without generating artefacts. Supplied in a bag of 30 units and with a minimum sale unit of a box of 120 units, it is the high-turnover solution for hospitals, clinics and healthcare companies that need an electrode with excellent adhesion and low cost per patient.
2. Technical Explanation
This electrode is a Class I medical device with CE marking and is verified according to the voluntary standard ANSI/AAMI EC 12:2000/(R)2005, which defines the performance requirements for disposable ECG electrodes. It is manufactured with a foam coated with an aggressive, non-sensitising acrylic copolymer, which maximises adhesion even under conditions of perspiration and stress, maintaining a low skin-electrode interface impedance. The solid conductive gel is presented in a central column and retains its aqueous content in a polymerised hydrogel formulation: it is non-irritating, non-sensitising and non-cytotoxic to the skin, and all components are latex-free. Its radiotransparent construction, including the snap connector, avoids artefacts in imaging studies (X-ray and MRI), so the electrode can remain in place without the need to remove monitoring. It is indicated for standard and continuous monitoring, ECG recording, Holter and stress testing in difficult-adhesion environments: ICU/CCU, operating theatre, recovery room, cardiac rehabilitation and dental surgery.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | 15843 |
| Trade name | F 50 S BLUE – Multifunction radiotransparent electrode |
| Type | Conventional multifunction foam and solid gel electrode |
| Backing material | Foam (medical foam) |
| Gel | Solid conductive gel (polymerised hydrogel) |
| Diameter | Ø 50 mm |
| Connection type | Radiotransparent snap (press stud), 1.55 mm |
| Radiotransparency | Yes: suitable for X-ray and MRI scans (without imaging artefacts) |
| Adhesive | Medical-grade acrylic copolymer, aggressive and non-sensitising |
| Composition | Non-irritant, non-sensitising, non-cytotoxic; latex-free |
| Use | Single-use, disposable, non-sterile |
| Presentation | Bag of 30 units |
| Minimum sales unit | Box of 120 units |
| Applications | Continuous monitoring, ECG, Holter and stress testing |
| Regulatory classification | Class I medical device with CE marking (Directive 93/42/EEC) |
| Reference standards | ANSI/AAMI EC 12:2000/(R)2005 |
| Expiry | 36 months in sealed packaging |
| Storage | Store in a cool, dry place in sealed packaging until use |
4. Equivalences and Cross-References
Reference 15843 (F 50 S BLUE) is the Cahnos distribution code for this radiotransparent monitoring electrode; it is a standard consumable with a snap connector, not an official reference from the monitoring system manufacturer. After verifying official manufacturer catalogues and regulatory databases (GUDID/FDA), the following cross-references with confirmed codes in public sources are identified. Any comparison between brands should be understood as comparable at category level, not interchangeable 1:1: same function (adult ECG monitoring with snap connection), but with verifiable differences in geometry, gel or radiotransparency that require prior validation.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| 3M / Solventum | Red Dot™ Monitoring Electrode with foam and sticky gel (solid adhesive gel) | 2560 (also V2560-5) | Comparable by category (not interchangeable 1:1) | Code confirmed in the official Solventum/3M catalogue and in GUDID (DI 00707387789305): foam electrode, solid adhesive gel, snap connection and high-performance adhesive with validated use up to 5 days. It differs in geometry: rectangular format of 4 x 3.5 cm versus the circular Ø 50 mm. Its GUDID record does not include safety information for MRI and its IFU warns that the metal snap connector is not tested for MRI: verify radiotransparency and compatibility with your system before replacing. |
| Cardinal Health | Kendall™ Medi-Trace™ 530 Series Foam Electrodes (adult, diaphoretic) | 31013926 (series 530) | Comparable by category (not interchangeable 1:1) | Code confirmed in authorised technical distribution (McKesson, Davis Medical) and in the manufacturer’s datasheet: foam electrode with conductive adhesive hydrogel (solid gel), snap connection, ANSI/AAMI EC 12 compliance, use in stress, Holter and diaphoretic monitoring with suggested use time of less than 72 h. Drop shape of approximately 44 mm; radiotransparency is not confirmed for this specific SKU. Verify geometry, gel and radiotransparency of the batch before replacement. |
| Ambu | BlueSensor™ R (adult, foam, snap) | R-00-S / R-00-S/5 | Comparable by category (not interchangeable 1:1) | Code confirmed in Ambu’s official catalogue and authorised technical distribution: foam electrode with Ag/AgCl sensor and offset snap connector, approx. 57 x 48 mm. Key difference: highly conductive wet gel versus the solid gel of this reference, and off-centre connector versus the centred stud. Suitable for stress testing and monitoring; check gel type and connector fit before replacement. |
| Leonhard Lang (Skintact®) | Skintact™ FS-RGC – Radiotransparent foam electrode with carbon stud | FS-RGC/10 | Comparable by category (not interchangeable 1:1) | Code confirmed in the manufacturer’s catalogue and authorised technical distribution: foam electrode with solid adhesive Aqua-Tac gel (adhesive hydrogel), radiotranslucent carbon stud and labelled MR-conditional for X-ray and MRI, without removing the electrode. It differs in geometry: rectangular format of 41 x 32 mm versus the circular Ø 50 mm. It is the closest radiotransparent alternative in concept; check contact surface and snap connector fit (10 mm stud / standard clip) before replacement. |
| ConMed | TotalTrace™ RTL Adult Foam General Purpose ECG Electrode (radiolucent carbon stud) | 3700C-050 | Comparable by category (not interchangeable 1:1) | Code confirmed in ConMed’s official catalogue and authorised technical distribution: low-profile foam electrode with adhesive conductive gel (solid gel) and radiolucent carbon stud (RTL) that improves visualisation on X-rays. It differs in geometry (smaller footprint format, not circular Ø 50 mm) and in that its radiotransparency is documented for X-rays; MRI compatibility of the specific batch must be verified before replacement. |
There is no equivalence table validated by the manufacturer that authorises 1:1 substitution between brands of monitoring electrodes: geometry, gel type, radiotransparency and active surface area affect signal quality and adhesion. Any substitution must be validated against the technical data sheet and instructions for use from the electrode manufacturer and the monitoring system, checking one by one diameter, gel, connector and radiotransparency; consult the technical department at Cahnos before use in tenders.
5. Selection tips and variants of interest
Check your system’s connector: this electrode uses a snap-type press stud (1.55 mm). Check that your monitor’s clip or cable is compatible with snap connection before placing the order; if your installed base uses a 4 mm banana connector, consider an electrode from the same family with that configuration.
Choose radiotransparent when imaging tests are anticipated: if the patient may require X-ray or MRI during monitoring, this version avoids artefacts without removing the electrode. If imaging is not required, the standard version of the same family with liquid gel (reference 0590024, ST 50 RLI) or with solid gel for stress testing (0590028, ST 50 RFI) may be considered, adapting the format to the department’s protocol.
Variants of the same family at Cahnos: radiotransparent electrode with radiotranslucent stud and box of 600 units (0590022, RT 50 RFI), rectangular 28 x 44 mm with solid gel and banana connector (0590025, NM 2844 FBI), rectangular 28 x 44 mm for resting ECG (16115, NM 2844 RFS), oval 51 x 33 mm (0590026), 45 x 42 mm format with solid gel and Ag/AgCl sensor (0590027, NM 3642 OFI) and paediatric version Ø 30 mm (0590016, NM 30 RFI). All share foam and gel technology, but differ in size, geometry and connection type: check your system’s cable and clip before selecting.
Storage and expiry: store in a dry, cool place, in the closed packaging until use, and respect the stated expiry date (36 months in closed packaging); the solid gel may lose conductivity if exposed to air. As it is a single-use consumable, maintain a minimum stock per monitoring station and rotate stock (FIFO). For volume orders or tenders, contact the technical-commercial department at Cahnos stating the monitor model and connector type of your installed base.





Reviews