Description
1. Commercial Introduction
Keep your defibrillator always operational with this pair of radiotransparent adult defibrillator electrodes with internal cable (Ref. F7759), a highly reliable consumable compatible with the equipment CU Medical Systems i-PAD NF 1200 and CMOS Drake Life 400 Futura, as well as with the defibrillators of the Saver One line. Designed for emergency services, ICU, ambulances, health centres, operating theatres and cardioprotected points, it simplifies stock management with a single reference per platform. Its integrated internal cable configuration maintains a clean and stable connection throughout resuscitation, and its radiotransparent construction allows imaging control without removing the patches. Available with immediate dispatch from Cahnos and specialised technical assistance for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronised cardioversion, non-invasive transcutaneous pacing (external pacemaker) and emergency ECG monitoring. The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal transmission of the shock energy; the internal metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest.
Being radiotransparent —electrode and cable—, they do not generate artefacts in radioscopy, tomography or intraoperative control, allowing the healthcare team to assess the rib cage, pacemaker or vascular accesses without removing the patches. The recommended placement in adults is anterolateral (sternum-apex), according to the anatomical diagrams on the packaging. They are single-use: after resuscitation they must be discarded and replaced, always respecting the expiry date printed on the pouch.
3. Technical specifications and compatibility
3.1 Technical specifications
Main characteristics of the series of radiotransparent adult electrodes (internal/external cable) to which this reference belongs:
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7759 |
| Product type | Pair of adult multifunction defibrillator electrodes, radiotransparent, with internal cable |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Radiotransparency | Yes (electrode and cable), suitable for imaging procedures |
| Cable configuration | Integrated internal cable (leads-in) |
| Functions | External defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring |
| Total surface per electrode | 136 cm² |
| Active surface per electrode | 94 cm² |
| Adhesive ring width | 0.7 cm |
| Supported shocks | 50 to 360 J |
| ECG monitoring | Max. 24 h |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (1 mm thick) |
| Connection type | Anti-shock safety connector or clip (depending on the equipment model) |
| Use | Single-use, disposable; latex-free and phthalate-free |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 10 pairs per box |
| Shelf life | Approx. 30 months from production date (expiry on packaging) |
| Operating temperature | 0 °C to +50 °C |
| Storage | Dry place, between +5 °C and +35 °C |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
Technical note: the values for active surface (94 cm²), total area (136 cm²), shocks (50–360 J) and adhesive ring (0.7 cm) correspond to the series of radiotransparent adult electrodes of this product family; confirm with the technical data sheet of the supplied batch and the device manual before use in tendering.
3.2 Compatibility with defibrillators
| Equipment brand | Compatible models | Observations |
|---|---|---|
| CU Medical Systems | i-PAD NF 1200 (semi-automatic) · i-PAD NF 1201 (automatic, verify) | Proprietary connector for the NF 1200 platform |
| CMOS Drake | AED Life 400 Futura (Futura model) | Proprietary connector; the device detects authenticity and validity of the pads |
| Saver One line (Cahnos) | Models of the Saver One platform | Verify model and connector fit before ordering |
Note: compatibility refers to the connector platform of each family. Always verify the correct fit of the connector and cable of your device before clinical use.
4. Equivalences and Cross-References
Reference F7759 is a consumable compatible with the cited defibrillators, not an original spare part certified by each equipment manufacturer: the connector, authenticity/validity detection and energy management are proprietary. After verifying official manufacturer catalogues and equipment manuals, the following cross-references with code confirmed in public sources are identified. Any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| CU Medical Systems (equipment manufacturer) | i-PAD SAVER NF1200 / NF1201 Adult Electrode Pads (original equipment) | CUA0903PF | Comparable by category (not interchangeable 1:1) | Code confirmed in the official CU Medical Systems catalogue and in authorised distribution: original adult pair for the i-PAD NF 1200 and NF 1201 devices, with proprietary connector; the radiotransparent nature of the original has not been verified. Verify expiry and connector fit before replacement. |
| CMOS Drake (equipment manufacturer) | Disposable adhesive adult shock pads – Life 400 Futura (original equipment) | LT49131 · LT72471 | Comparable by category (not interchangeable 1:1) | Original references documented in Appendix A (accessories list) of the official user manual of the AED Life 400 Futura; proprietary connector and authenticity/validity detection system of the pads in the device. Verify expiry and connector fit before replacement. |
There is no equivalence table validated by CU Medical Systems or CMOS Drake that authorises the replacement of original electrodes with third-party pads: the connector, originality/validity detection and energy management of each defibrillator are proprietary. Any replacement must be validated against the technical data sheet and instructions for use of the defibrillator; consult the technical department of Cahnos before using in a tender.
5. Selection tips and variants of interest
Confirm the model and connector of your equipment: this pair corresponds to the CU Medical Systems i-PAD NF 1200 (and NF 1201, verify) and CMOS Drake Life 400 Futura defibrillators, as well as the Saver One line. Check the reference and expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention.
Choose the variant according to your workflow: the internal cable of this model provides a clean and robust connection during resuscitation; if radiolucency is not required, consider the standard version of the same platform (F7959), at a lower cost. If you attend to paediatric patients, paediatric references F7959P (CMOS Drake Futura) and F7958P (CU Medical Systems) are available, with automatic energy reduction when connecting the paediatric pair.
Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the expiry date on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As it is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO). For volume orders or tenders, contact the technical department of Cahnos indicating the exact model and serial number of your equipment.