Description
1. Commercial Introduction
The NM 30 RFI ECG electrode (SKU 590016) is the reference disposable consumable for cardiac monitoring and electrocardiographic recording in paediatric patients and in critically ill patients who require a secure and stable contact even under difficult adhesion conditions. Its circular foam format of Ø 30 mm, its highly conductive solid gel and its silver/silver chloride (Ag/AgCl) sensor guarantee a clean trace, low impedance and without the drawbacks of liquid gels: no residue, no mess and firm adhesion from the very first moment. It is the cost-optimised solution for paediatrics departments, ICUs, emergency departments, cardiology consultations, medical transport and any high-turnover monitoring point. Supplied in an aluminium pouch of 30 units, with a minimum sales unit of a box of 900 units.
2. Technical explanation of the electrode
The electrode acts as an electrical interface between the patient’s skin and the monitor or electrocardiograph. The foam support coated with an aggressive medical-grade acrylic copolymer adhesive firmly fixes the patch to the chest, while the central column of solid conductive gel —a polymerised hydrogel with aqueous content— maintains moisture and conductivity throughout the recording, without irritating the skin: it is non-irritating, non-sensitising and non-cytotoxic. The Ag/AgCl sensor offers a stable contact potential and very low polarisation, reducing motion artefacts and ensuring a reliable electrocardiographic signal. Its central snap connection (snap/pressure type) allows direct coupling to patient cables with button terminal, the most common standard in ECG cables with forceps or alligator clips for limbs. It is a single-use, hypoallergenic, latex-free and non-sterile product, indicated for standard monitoring and short-duration electrocardiograms.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Type | Conventional disposable electrode for paediatric ECG monitoring |
| Reference / model | NM 30 RFI · SKU Cahnos 590016 (0590016) |
| Support material | Foam (medical foam) coated with aggressive acrylic copolymer, non-sensitising |
| Shape | Circular |
| Diameter | Ø 30 mm |
| Gel | Conductive solid gel (polymerised hydrogel, non-irritant, non-sensitising, non-cytotoxic) |
| Sensor | Ag/AgCl (silver/silver chloride) |
| Connection | Central pressure snap (snap type), centred gel column |
| Application | Paediatric ECG, standard monitoring and critical patients |
| Composition | Hypoallergenic, latex-free |
| Electrical performance | Meets and exceeds ANSI/AAMI EC 12:2000/(R)2005; low contact impedance |
| Use | Single-use (disposable), non-sterile |
| Presentation | Aluminium sachets of 30 units |
| Minimum sales unit | Box of 900 units (5,400 units/master carton) |
| Expiry | 36 months with sealed packaging |
| Regulatory classification | Class I medical device with CE marking (Directive 93/42/EEC, D.L. 46/97) |
Verified electrical performance (according to the family technical data sheet):
| Parameter (ANSI/AAMI EC 12) | Value |
|---|---|
| DC offset voltage | 0.12 mV |
| AC impedance at 10 Hz | 60 Ω |
| DC offset 5 s after capacitor discharge | 6.5 mV |
| AC impedance after defibrillation | 46 Ω |
| Recovery slope (30 s interval) | 0.0 mV/s |
4. Equivalences and Cross-References
The NM 30 RFI reference is the line code of the electrode marketed by Cahnos; there is no official equivalence table published by the manufacturer that cross-references this code with third-party codes. After verifying official catalogues and manufacturer websites, the following homologous references with confirmed codes have been identified. All of them are classified as comparable by category (not 1:1 interchangeable): before any substitution, the diameter, support material, gel type, sensor, connector and commercial presentation must be checked one by one.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Lessa (AB Medica Group, Spain) | Snap electrode Ø 30 mm, foam, solid conductive adhesive gel | 9983012 (official cat.) | Comparable by category (not interchangeable 1:1) | They match Ø 30 mm, foam support, solid conductive gel, snap connector (clasp) and paediatric use for short-duration resting ECG. The presentation differs: box of 600 units (20 sachets of 30) versus 900 units. Verified on the manufacturer’s official website. |
| Dormo® (Telic Group, Spain) | DORMO® SX-30 ECG Electrode Ø 30 mm, solid gel, Ag/AgCl sensor | SX-30 (SKU TEM01) | Comparable by category (not interchangeable 1:1) | They match Ø 30 mm, foam support (flexible PE), solid gel of water-soluble acrylic polymer, Ag/AgCl sensor, latex-free, class I (EU Regulation 2017/745) and ANSI/AAMI EC12 requirements. Differences: allows wear time of up to 24 h, pouch of 100 units and box of 2,500 units, with adult/paediatric/neonatal application. Verified on the manufacturer’s website. |
| Medke | Disposable Infant Foam Electrode, Solid Gel, White | BA-103 | Comparable by category (not interchangeable 1:1) | They match Ø 30 mm, foam support, solid gel, paediatric/infant use, latex-free and snap or forceps connection. The commercial presentation differs (pouches of 100 units). Code verified on the manufacturer’s official website and in its public CE/ISO 13485 data sheet. |
| Leonhard Lang (Skintact®) | F-301 Monitoring ECG Electrode, 30 mm, Paediatric, Solid Gel | F-301 | Comparable by category (not interchangeable 1:1) | They match Ø 30 mm, foam backing, solid gel (Aqua-Tac) and paediatric use. The presentation differs (30 units/sachet, box of 1,500 units) and the backing (foam/microporous depending on version). Verified on the manufacturer’s website and authorized technical distribution. |
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders. Any substitution between brands must be validated against the technical data sheet of the manufacturer issuing the code, checking one by one diameter, backing material, gel type (solid or liquid), sensor, connector type (snap versus 4 mm banana) and regulatory classification.
5. Selection tips and variants of interest
Check the connector on your patient cable: this electrode has a central press stud (snap type). If your cable ends in a 4 mm banana, consider the rectangular foam electrode of 28 x 44 mm with banana connector (NM 2844 FBI, SKU 590025) or use a snap-to-banana adapter; if you use crocodile clips, the range of tab electrodes is the appropriate alternative. Choose the diameter according to the patient: the circular format of Ø 30 mm is the reference size for paediatrics and critical patients; for very low birth weight neonates, the pre-wired radiolucent family of Ø 22 mm is available (RT 22 LWF, SKU 15079) and for adults or stress testing, the circular electrode of Ø 50 mm with high-adhesion solid gel (ST 50 RFI, SKU 0590028) or its version with liquid gel (ST 50 RLI, SKU 0590024). For recordings with possible imaging control, consider the radiolucent electrodes from the same family (SKU 0590022). Keep the bags sealed in a dry place, rotate stock on a FIFO basis and respect the 36-month expiry: the solid gel may lose conductivity if exposed to air. As it is a single-use consumable, maintain a minimum stock per monitoring station. For supplies, tenders or bids, contact the technical-commercial department of Cahnos indicating the model of your monitor fleet and the type of connector on the patient cable.