Description
1. Commercial Introduction
The disposable foam ECG electrode with offset snap connection (reference 33309, code FIAB F2017) is a high-turnover consumable designed for cardiac monitoring in adult patients. Its round format of Ø 50 mm, conductive liquid gel and silver chloride sensor ensure a stable signal from the first seconds of recording, while the offset snap press stud facilitates a firm fit with the patient cable and better anatomical adaptation. It is supplied in a box of 1,000 units (20 bags of 50 units), making it the ideal solution for cardiology departments, ergometry units, sports medicine, ICUs and purchasing centres that need a reliable electrode that is comfortable for the patient and offers excellent cost per procedure.
2. Technical Explanation
A surface ECG electrode acts as an interface between the patient’s skin and the monitor or recorder: it captures the electrical activity of the heart and transmits it with the least possible distortion. The quality of the recording depends on three factors: adhesion, conductivity (low skin-electrode contact impedance) and stability against movement.
The polyethylene foam (PE foam) backing is lightweight, flexible and moisture-resistant: it protects the gel and sensor, adapts to the contour of the chest and maintains its integrity during short-term monitoring recordings, stress tests and Holter studies. The conductive liquid gel ensures rapid signal acquisition with low initial impedance, reducing motion artefacts typical of ergometry. All electrodes incorporate a silver chloride (Ag/AgCl) sensor and an offset snap press stud connection, which provides a secure fit with standard patient cables. This is a single-use product for clinical and hospital use, which must be disposed of after each patient and stored in its closed packaging in a cool, dry place.
3. Technical Specifications
Main features of the disposable foam ECG electrode with offset snap connection (reference 33309 / FIAB F2017):
| Parameter | Specification |
|---|---|
| Reference (SKU) | 33309 |
| FIAB code | F2017 |
| Type | Disposable surface ECG electrode |
| Material | PE-Foam (polyethylene foam) |
| Size | 50 mm (round format) |
| Gel | Liquid |
| Connection | Off-centre snap press stud |
| Sensor | Silver chloride (Ag/AgCl) |
| Patient | Adult |
| Application | Short- and medium-term monitoring, stress ECG (ergometry) and Holter study |
| Presentation | Box of 1,000 units (20 bags of 50 units) |
| Sales unit | Box of 1,000 units |
| Use | Single-use, disposable |
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | Box of 1,000 units |
| Device type | Medical device |
| Device class | I |
| CND | C020501 |
| GMDN | 35035 |
| EAN | 8054953341224 |
4. Equivalences and Cross-References
Reference 33309 corresponds to the Cahnos distribution code (SKU GM33309) for this foam electrode with liquid gel and offset snap connector; code F2017 is the manufacturer’s reference from FIAB. There is no equivalence table validated by the manufacturer that authorises a 1:1 cross-reference with third-party brands: substitution must be verified field by field (diameter, gel type, connector position and adhesive) before any change.
By category, the following are identified as comparable (not interchangeable 1:1): foam electrodes with wet gel and offset snap connector such as the Ambu BlueSensor™ R (R-00-S / R-00-S/25), the FS-50 foam electrode with Aqua-Wet™ liquid gel from Leonhard Lang (Skintact®) and the foam electrodes with adhesive gel Red Dot™ from 3M/Solventum (2560, V2560-5). All share application in monitoring and stress testing in adults, but differ in geometry, connector position or gel formulation, so the fit of the patient cable to the monitor fleet must be validated before substituting.
5. Selection tips and variants of interest
Check the connector: this electrode features an offset snap press stud. Confirm that the patient cable of your monitor, ergometer or Holter recorder is compatible with snap connection before placing the order; if your fleet uses crocodile clips or banana connectors, consult the range of tab electrodes and connection adapters from Cahnos.
Variants within the Cahnos catalogue: for liquid gel and Ø 50 mm with centred snap, you have the FS 50 L electrode (ref. 11312, box of 1,200 units) or the high-adhesion ST 50 RLI (ref. 0590024, box of 600 units); for monitoring and resting ECG, the F 50 L (ref. 15492); and for a larger adhesive surface, the F 55 L with Ø 55 mm (ref. 11310). If your department performs prolonged monitoring or recordings with heavy sweating, consider the solid gel versions from the same family, such as the ST 50 RFI (ref. 0590028). All share clinical indication but differ in gel, diameter or connector position: adapt the choice to your department’s protocol and the connection type of your cables.
Preparation and storage: clean and dry the skin before placement and avoid creams or oily residues to optimise contact impedance. Keep the bags closed in a cool, dry place, rotate stock on a FIFO basis and respect the expiry date: the liquid gel may lose conductivity if the packaging is opened or damaged. For volume orders, full allocations or tenders, contact the technical-commercial department of Cahnos, indicating the model of your monitor fleet and the connection type.