Description
1. Product Presentation
The connection cable for concentric needle electrodes with 3-pole microconnector (Ref. F8995/MM) is the replacement accessory your neurophysiology service needs to keep Micromed electromyographs with ISA heads operational. With 100 cm length and shielded construction, it reliably replaces the electrode support cable, ensuring precise, interference-free transmission of very low-amplitude EMG signals. A robust, reusable cable designed for daily use in clinics, hospitals and neurodiagnostic laboratories, fully compatible with concentric needles of the F8990/F8990PD series with detachable lead wire.
2. Technical explanation of the equipment and procedure
In an electromyography (EMG) study, the concentric needle records the electrical activity of muscle fibres by inserting a fine electrode into the muscle. The electrode consists of a central wire (cathode) and an outer cannula (anode) separated by an insulating material; the captured signal is in the order of microvolts and requires shielded, low-impedance transmission to avoid electrical noise and artefacts that distort the trace.
This cable acts as the interface between the needle electrode and the equipment input: the distal end couples to the electrode connector (detachable lead wire) and the proximal end terminates in a 3-pole microconnector that connects to the amplifier input of the ISA head of Micromed systems. Its shielded construction ensures an optimal signal-to-noise ratio in both routine examinations and prolonged EMG, electroneurography (NCS) and other neurophysiological tests.
3. Technical Specifications
| Specification | Value |
|---|---|
| Product type | Reusable connection cable for concentric needle electrodes |
| Reference / SKU | F8995/MM (Cahnos distribution code) |
| Connector (equipment) | 3-pole microconnector |
| Cable length | 100 cm |
| Compatible electrodes | Concentric needles F8990 / F8990PD series with detachable lead wire |
| Equipment compatibility | Micromed equipment with ISA heads |
| Shielding | Shielded cable for stable signal transmission and low interference |
| Application | EMG, electroneurography (NCS), EEG and neurophysiological tests |
| Use | Reusable · clinical, hospital and neurodiagnostic laboratories |
| Presentation | Unit |
4. Equivalences and Cross-References
The reference F8995/MM is a proprietary distribution code, internally documented as a 100 cm cable with 3-pole microconnector for concentric needles in Micromed systems with ISA head. The third-party references shown below belong to the same functional category (patient cable for concentric needles) and have been cross-checked against primary sources (official manufacturer catalogue, official website and GUDID/FDA regulatory database); there is no manufacturer-validated equivalence table declaring them 1:1 interchangeable with this connector, so they should be treated as comparable at category level.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ambu / Neuroline | Neuroline™ concentric cable (5-pin DIN), latex-free | 1741 (100 cm) · 1742 (200 cm) | Comparable by category (not interchangeable 1:1) | Same function: reusable shielded cable for concentric needles. Differs in connector (5-pin DIN, standard 60130-9, vs. 3-pole microconnector) and electrode coupling system. Code and attributes verified in official Ambu catalogue and authorised distribution. |
| Cadwell | Reusable connection cable for concentric needles (detached lead), 5-pin DIN connector | 202462-000 (1.25 m) · 202463-000 (2.0 m) | Comparable by category (not interchangeable 1:1) | Same functional category. Differs in connector (5-pin DIN), length (125/200 cm) and coupling socket design. Code and attributes verified in the official Cadwell eStore. |
| Natus | TECA® Elite Concentric Needle Holder (cable) | X21001 | Comparable by category (not interchangeable 1:1) | Shielded cable, 36 inches (0.9 m) with 5-pin DIN connector (240°) at the equipment end and pogo-pin contact at the electrode end. Differs in connector, length and coupling system. Code and attributes verified in the Natus supplies catalogue and authorised distribution. |
| EB Neuro (original equipment) | Support cable for concentric and single-fibre needles, 5-pin DIN connector | B9720336001 (100 cm) · B9720336002 (200 cm) | Comparable by category (not interchangeable 1:1) | Shielded cable from the manufacturer of Myto2, Nemus and Nemus2 systems, with free-orientation 5-pin DIN connector. Differs in connector and electrode coupling socket. Code and attributes verified in the official EB Neuro accessories catalogue. |
| LifeSync | Reusable Cable for Concentric Needles (5-pin DIN connector) | NC-850125 (1.25 m) · NC-850200 (2.0 m) | Comparable by category (not interchangeable 1:1) | Reusable cable with 5-pin DIN connector for EMG/NCS configurations with detachable-lead concentric needle. Differs in connector and length. Verified in the official LifeSync datasheet. |
| Rochester Electro-Medical | Reusable shielded cable, 5-pin DIN, for concentric and single-fibre needle | GUDID registered device 00849593025300 (1.25 m / 4 ft) | Comparable by category (not interchangeable 1:1) | Reusable shielded 5-pole DIN cable for concentric and single-fibre needles. They differ in connector, length and regulatory registration (US market). Verified against the GUDID/FDA regulatory database. |
There is no manufacturer-validated equivalence table cross-referencing the F8995/MM reference with codes from other brands; consult the technical department before using in a tender. Before any substitution, check field by field: pin configuration (3-pole microconnector vs 5-pole DIN), cable length, electrode coupling system (separable lead) and compatibility with the ISA headbox socket of the installed electromyograph.
5. Selection tips and variants of interest
Choose the cable according to the socket on your electromyograph: Ref. F8995/MM (3-pole microconnector, 100 cm) is the specific variant for Micromed systems with ISA headboxes. If your installed base includes other electromyographs such as Cadwell, Nihon Kohden, Natus (Keypoint) or Carefusion (Medelec, Synergy, Nicolet, Viking), request the universal variant F8995/NI (5-pole DIN, 100 cm); for EB Neuro Myto2, Nemus and Nemus2 equipment, the variant F8995/EN (5-pole DIN, 100 cm).
Complete the set with the concentric needles from the F8990/F8990PD series: choose the gauge and length according to the muscle to be examined (diameters from 0.35 to 0.50 mm and lengths from 20 to 60 mm, in sterile single-use and reusable presentations), always with a separable lead wire to couple to this cable.
Maintenance of the reusable cable: clean the outer surface according to the device’s reprocessing instructions and avoid autoclave sterilisation unless expressly indicated by the manufacturer; periodically inspect the 3-pole microconnector and the shielding, and replace the cable at any sign of deterioration so as not to compromise signal quality. For institutional volumes, tenders or specific length requirements, consult our technical-commercial team at Cahnos.





Reviews