Description
1. Product Presentation
At Cahnos we offer an absorbable multifilament suture of poliglactin 910 (PGLAR) in RAPID version and Prime quality, designed for the supply of operating theatres, purchasing centres and ophthalmic surgery clinics. This reference incorporates a straight needle with triangular point and conventional cutting edge of 48 mm (GS48) and a double strand configuration of 2×75 cm in USP gauge 3/0, with colourless and non-grooved thread. Predictable mechanical behaviour, minimal tissue response and reliable knot tying, in a box presentation of 12 units for B2B hospital supply conditions.
2. Technical explanation of the material and procedure
Poliglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, structured as a multifilament braid that provides flexibility, soft feel and great knot security. Its absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the thread degrades into substances that the body metabolises without causing relevant tissue reaction.
The RAPID variant is optimised for short-term support: it retains 50% of tensile strength at 5 days and loses 100% of its resistance at 14 days, with complete absorption between 40 and 45 days. The GS48 straight needle of 48 mm, with triangular point, conventional cutting edge and non-grooved surface, facilitates penetration into firm tissues with absolute control of the path, being especially suitable for manual suturing on directly accessible surfaces, approximation stitches, superficial continuous sutures, ligatures and ophthalmic surgery procedures. The double strand format of 2×75 cm reduces handling in the operating theatre and increases yield per unit.
3. Technical Specifications
| SKU Cahnos | AMP2L22GS4830.2 |
| Tariff reference (PREMISUT) | P2L22RGS4830.2 |
| Material | Polyglactin 910 (PGLAR Prime), copolymer 90% glycolide / 10% L-lactide |
| Thread structure | Braided multifilament, non-grooved surface |
| Gauge | USP 3/0 (metric 2) |
| Thread length | 2 × 75 cm (double strand) |
| Needle shape | Straight |
| Tip type | Triangular — conventional cutting (conventional cutting) |
| Needle size | GS48 (48 mm) |
| Colour | Colourless |
| Absorption | By hydrolysis, between 40 and 45 days |
| Sterilisation | Gamma radiation; sterile, single-use |
| Quality | Prime (PREMISUT) |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
Tensile strength retention profile after implantation:
| Time point | Tensile strength retention |
| 5 days | 50% of initial strength |
| 14 days | 0% (total loss of strength) |
| Complete absorption | 40–45 days by hydrolysis |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate.
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references this specific configuration (GS48 straight needle with conventional cutting edge of 48 mm, USP gauge 3/0, colourless thread in double strand 2×75 cm) with references from other commercial houses; consult the technical department before using in tenders.
At category level, rapid-absorption poliglactin 910 corresponds to the market’s braided synthetic absorbable sutures with rapid absorption. The closest homologous family is VICRYL RAPIDE™ from Ethicon / Johnson & Johnson MedTech (poliglactin 910, same composition 90% glycolide / 10% L-lactide and same absorption profile of 40–45 days with total loss of strength at 14 days), as well as the Safil Quick and Novosyn Quick ranges from B. Braun in the segment of multifilament absorbables with rapid loss of strength. Any code from these lines should be checked against their technical data sheet and IFU before proposing substitution, since needle nomenclatures, point type (conventional cutting vs reverse cutting) and configuration (single strand vs double strand) vary by manufacturer. The cited brands belong to their respective manufacturers and are mentioned only as an equivalence reference.
5. Selection tips and variants of interest
To choose the correct suture, verify three parameters: USP gauge (this model, USP 3/0, is intended for soft tissues of thin-medium thickness that require short-term support), needle geometry (straight with conventional cutting edge for direct access sutures with good control of the path; circular variants 3/8 or 1/2 are reserved for reduced fields) and configuration (the double strand format of 2×75 cm offers greater yield per unit).
Variants of interest within the Cahnos catalogue:
- AMP2L22GS4830 (tariff P2L22GS4830): same GS48 needle, USP 3/0, colourless, in single strand of 75 cm.
- AMP2L22GS4820 / AMP2L22GS4820.2: GS48 straight needle, USP 2/0, colourless, in 75 cm and 2×75 cm.
- AMP2L22GS6030 / AMP2L22GS6030.2: straight needle GS60 (60 mm), USP 3/0, undyed, in 75 cm and 2×75 cm.
- AMP2L28GR1940: straight needle conical point GR19, USP 4/0, undyed, 70 cm, for microsurgery and more reduced fields.
As this is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply agreements or queries about equivalences from other commercial houses, contact the technical-commercial team at Cahnos.