Description
1. Commercial Introduction
The absorbable multifilament suture PGAR RÁPIDO in Prime quality (PREMISUT RÁPIDO range) is the reference for accelerated absorption surgical closure for hospitals, general surgery operating theatres, emergency departments and B2B purchasing centres that need a braided, flexible thread with early, programmed and fully predictable loss of strength. Its ½ circle needle with reverse cutting triangular point of 17 mm (HS17 Prime profile) guarantees clean and controlled penetration in fine and firm tissues, while the undyed 75 cm thread offers maximum discretion on the surgical field. A practical and cost-effective presentation, in a box of 12 units, designed for tissues that heal quickly.
2. Technical Explanation
Polyglycolic acid (PGAR) is a synthetic absorbable polymer spun into a braided multifilament structure, coated with polycaprolactone and calcium stearate to favour thread sliding, smooth handling and a firm, secure knot. In the RÁPIDO variant, the thread is formulated to lose tensile strength at an accelerated rate: it retains 50% of the initial resistance at 5 days and 0% at 14 days, with complete absorption by hydrolysis between 40 and 45 days. This profile makes it especially suitable for superficial tissues with rapid healing (skin closure, mucous membranes, subcutaneous tissue) and ligation of small-calibre vessels. The reverse cutting needle creates a clean entry hole by directing the cut away from the wound, preserving the skin edge and preventing the cut from enlarging in the puncture bed. Sterile single-use product, not re-sterilisable.
3. Technical Specifications
| Characteristic | Specification |
| Factory reference | P222HS1740 (Prime range) |
| Reference Cahnos (SKU) | AMP222HS1740 |
| Material | 100% polyglycolic acid (PGAR), coated braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Quality | Prime (PREMISUT RAPID range) |
| Gauge | USP 4/0 (metric 1.5 / EP 1.5) |
| Approximate thread diameter | 0.15 – 0.199 mm (according to Pharmacopoeia) |
| Thread colour | Colourless |
| Thread length | 75 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle tip | Triangular, reverse cutting |
| Needle size | HS 17 Prime (17 mm), 1 needle per thread |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial format | Box of 12 units (minimum order 30 boxes) |
Absorption and tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50% of initial strength |
| 14 days | 0% (total loss of tensile support) |
| 40 – 45 days | Complete absorption by hydrolysis (predictable and uniform rate) |
Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
4. Equivalences and Cross-References
Fast-absorbing polyglycolic acid (low molecular weight) is one of the most widespread categories of multifilament absorbable suture in the surgical market. This reference shares fast-acting absorbable synthetic material, USP 4/0 gauge, undyed 75 cm thread and early strength loss profile (50% at 5 days / 0% at 14 days) with the homologous lines of the main manufacturers. Only one code is included whose attributes have been verified in the manufacturer’s official catalogue and authorised distributors:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE® (braided polyglactin 910, undyed) | W9930 – USP 4/0, 75 cm, undyed, FS-2 needle of 19 mm, ⅜ circle, reverse cutting | Comparable by category (not interchangeable 1:1) | Same rapid absorption profile (≈42 days), USP 4/0 gauge, undyed colour and 75 cm length; they differ in needle curvature (⅜ vs ½ circle) and size (19 mm vs HS 17 mm). Code confirmed in Ethicon’s official catalogue and authorised distributors. Verify geometry in the official data sheet before substituting. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Homologous ranges in the rapid absorption segment include Safil Quick / Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category, material and absorption profile, although for the exact combination USP 4/0 / ½ circle HS 17 mm needle / 75 cm thread, their specific codes must be confirmed with each manufacturer. Standard VICRYL (absorption 56–72 days) is not comparable to this RÁPIDO version. Within the PREMISUT system itself, the direct equivalent with identical geometry, gauge and length is the standard quality version (Ref. 222HS1740, SKU AM222HS1740).
5. Selection tips and variants of interest
The USP 4/0 gauge (metric 1.5) is the reference size for skin, subcutaneous tissue and soft tissues with low-medium tension; the 17 mm ½ circle reverse cutting needle is especially recommended when a firm and precise passage through skin or mucous membranes is required without tearing the edge. Within the same PGAR RÁPIDO range, Cahnos offers complementary configurations: the standard quality variant with identical configuration (SKU AM222HS1740), the same needle and gauge in 45 cm thread (SKU AMP221HS1740), the version in USP 3/0 gauge with 75 cm thread (SKU AMP222HS1730), the variant with cylindrical needle with conical point HR 17 mm (SKU AMP222HR1740) for looser tissues, and the PGLAR Polyglactin 910 RÁPIDO line with identical configuration (SKU AMP2L22HS1730) as a material alternative within the same absorption profile. The undyed version offers maximum discretion; if greater visibility in the field is required, request the same configuration in dyed thread on demand. Sterile single-use product: check the integrity of the packaging and the expiry date before each procedure. For special B2B conditions or volume orders (minimum 30 boxes), contact the technical-commercial team of Cahnos.





Reviews
There are no reviews yet.