Description
1. Commercial Introduction
At Cahnos we offer you the absorbable multifilament polyglycolic acid (PGA) suture in Prime quality, reference P111GS1340-2 (SKU AMP111GS1340-2), a high-end professional solution for B2B supply to operating theatres, clinics, minor surgery practices, purchasing centres and healthcare distributors. This presentation combines a double straight needle with conventional cutting triangular point of 13 mm (2xGS13), USP 4/0 gauge (metric 1.5) and a 45 cm violet braided thread with a non-grooved surface, with a soft feel and controlled passage through tissue.
It is the ideal choice for tissue approximation, ligatures and general surgery, with predictable absorption that avoids suture removal, reliable mechanical support during the first weeks of healing and special B2B ordering conditions for hospitals and volume buyers. Synthetic absorbable medical device, sterile and single-use.
2. Technical explanation of the material and procedure
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer, presented in a braided multifilament structure coated with polycaprolactone and calcium stearate. This combination guarantees smooth passage through tissue, low suture memory and firm, secure knot seating, even in fine gauges such as USP 4/0. With its non-grooved surface, the thread offers a very regular texture that reduces the sawing effect on tissue during passage.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without causing significant tissue reaction. Complete absorption occurs between 60 and 90 days, maintaining the necessary tensile support during the critical healing phase (75% at 2 weeks, 50% at 3 weeks and 40% at 4 weeks).
The straight needle with conventional cutting triangular point, GS13 type, 13 mm, mounted in a double needle (2x) configuration, is especially effective in soft and directly accessible tissues: it allows suturing from both ends of the strand, facilitates continuous manual suturing and bilateral work without re-threading, with controlled penetration and no tearing. The violet colour ensures optimal visibility of the thread on the surgical field throughout the procedure. The product is sterilised by ethylene oxide (ETO), is single-use and cannot be resterilised.
3. Technical Specifications
3.1. Technical data sheet for reference P111GS1340-2
| Characteristic | Specification |
| Reference (SKU) | AMP111GS1340-2 |
| Factory reference | P111GS1340-2 |
| Quality / range | Prime (Premisut range) |
| Material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Coating | Polycaprolactone and calcium stearate |
| Suture type | Synthetic absorbable braided multifilament |
| Thread structure | Braided multifilament, non-grooved surface |
| USP / metric gauge | 4/0 (metric 1.5) |
| Approximate diameter | 0.15 – 0.199 mm (USP/EP standard) |
| Thread colour | Violet |
| Thread length | 45 cm |
| Swaged | Double needle (2x) |
| Needle shape | Straight |
| Needle tip | Triangular, conventional cutting |
| Needle size | 13 mm (GS 13 Prime) |
| Absorption mechanism | Hydrolysis |
| Total absorption | 60 – 90 days |
| Sterilisation | Ethylene oxide (ETO), sterile, single-use, non-resterilizable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 30 boxes |
| Regulatory standard | USP / European Pharmacopoeia (EP) |
3.2. Tensile strength retention and absorption profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Complete absorption | 60 – 90 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 4/0 | 1.5 | 0.15 – 0.199 mm (indicative according to USP/EP) |
4. Equivalences and Cross-References
Braided coated multifilament polyglycolic acid (PGA) with medium absorption (60-90 days) is one of the most widely used synthetic absorbables in surgical practice (classic ‘Dexon’ type family). The homologous reference from another manufacturer indicated below matches this suture in material, gauge, needle, length and colour and has been verified in the GUDID/FDA regulatory database. As it is from another manufacturer, it is not declared interchangeable 1:1 without validation by the technical service.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | Safil® Violet (100% polyglycolic acid, braided multifilament coated with gluconate) | C1048211 (SAFIL VIOLET 4/0 (1.5) 45CM GS13 M, verified in GUDID, DI 04038653936403) | Comparable by category (not interchangeable 1:1) | It matches in material (100% PGA), USP 4/0 gauge (EP 1.5), straight conventional cutting needle GS13 of 13 mm, 45 cm thread and violet colour. They differ in the configuration (single needle versus double needle 2x of this reference), the coating (gluconate versus polycaprolactone and calcium stearate) and the commercial presentation (usually a box of 36 units). Confirm availability on the European market before substituting. |
As homologous families of braided coated PGA within the same clinical category, without a code verified in public sources for this exact configuration (double straight needle GS13, USP 4/0, violet, 45 cm), the following are cited for guidance: Surgicryl® PGA (SMI), DemeSORB® PGA (DemeTECH), AssuCryl® PGA (Assut Sutures) and Petcryl (Dolphin Sutures); these correspondences are pending confirmation with the manufacturer or distributor before use in tenders. The copolymer range of polyglactin 910 (PGLA), such as Coated Vicryl® (Ethicon) or Novosyn® (B. Braun), shares the multifilament behaviour and medium absorption philosophy, but uses a different copolymer (90% glycolide / 10% lactide) with absorption of 56-72 days, so it is not interchangeable 1:1 material to material with this PGA suture.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. The correspondences indicated are functional comparisons verified against official sources (manufacturer catalogue and GUDID) and must be confirmed code by code before a formal substitution of references. Within the Cahnos catalogue, the reference with identical geometry configuration (double straight needle GS13, USP 4/0, violet, 45 cm) in polyglactin 910 (PGLA) is P1L11GS1340-2 (absorption of 56-72 days).
5. Selection tips and variants of interest
Select the gauge according to the tissue and the tension the suture must withstand: USP 4/0 (metric 1.5) is indicated for fine tissues under moderate tension, such as mucosa, subcutaneous tissue, delicate planes, ligatures and precision approximations in general surgery and minor ambulatory surgery. The double straight needle with a 13 mm conventional cutting triangular point is especially suitable for continuous manual suturing on directly accessible surfaces where firm, controlled penetration into soft tissues is required, allowing work from both ends of the strand without re-threading.
| Reference | Variant of interest |
| P1L11GS1340-2 | PGLA Prime, double straight needle GS13 mm, USP 4/0, violet, 45 cm (absorption 56-72 days) |
| AMP111GR1940-2 | PGA Prime, double straight needle GR19 mm conical point, USP 4/0, violet, 45 cm (atraumatic penetration) |
| AMP118GR1940 | PGA Prime, straight needle GR19 mm conical point, USP 4/0, violet, 70 cm (longer strand, single needle) |
| AMP111DS1940 | PGA Prime, ⅜ circle needle DS19 mm reverse cutting, USP 4/0, violet, 45 cm (firm tissues) |
| AMP111DS1240 | PGA Prime, ⅜ circle needle DS12 mm reverse cutting, USP 4/0, violet, 45 cm (reduced fields) |
| AMP411GS1340 | Polydioxanone (PDO) monofilament, straight needle GS13 mm, USP 4/0, violet, 45 cm (prolonged absorption 180-210 days) |
The violet colour improves thread visibility during the procedure; if maximum aesthetic discretion is preferred, the range also offers a colourless version on request. All references are supplied in boxes of 12 units, with a minimum order of 30 boxes. As this is sterile single-use material, check the integrity of the packaging before each procedure, observe the expiry date and store in a cool, dry place protected from sunlight. For special volume-based B2B terms, hospital supply contracts or equivalences from other commercial brands, contact our technical-commercial team at Cahnos.





Reviews