Description
1. Commercial Introduction
The Prime range multifilament absorbable suture of polyglycolic acid (PGA), reference P112DR2620, is the trusted choice for the professional seeking a synthetic absorbable thread with predictable behaviour, excellent handling and maximum knot security in tissue approximation, ligatures and general surgery. Its ⅜ circle needle with cylindrical body and 26 mm (DR 26) Taperpoint conical point guarantees atraumatic and controlled penetration in soft and friable tissues, while the 75 cm violet grooved braided thread offers optimal visibility on the surgical field and smooth sliding during suturing.
It is a synthetic, absorbable, sterile, single-use medical device with an excellent quality-price ratio for the B2B supply of operating theatres, clinics, purchasing centres and distributors of surgical material. Commercial presentation in a box of 12 units, with a minimum order of 30 boxes to guarantee continuous supply to the healthcare centre.
2. Technical explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This configuration provides a regular surface that facilitates the passage of the thread through tissue, reduces the sawing effect, offers smooth handling and guarantees firm and secure knot tying, even in medium working gauges such as USP 2/0.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days and independent of the patient’s enzymatic activity. This profile maintains the necessary tensile strength during the first weeks of healing (75% at 2 weeks, 50% at 3 and 40% at 4), while the tissue recovers its own resistance.
The DR 26 mm needle, with cylindrical body and Taperpoint conical point, passes through tissue separating the fibres without cutting effect, minimising injury to the tissue bed; it is therefore especially suitable for delicate tissues, soft tissue approximation and ligatures. The violet colour improves visibility of the thread on the surgical field. The product is supplied sterile by ethylene oxide, is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet for reference P112DR2620
| Parameter | Specification |
|---|---|
| Brand / Range | PGA Prime |
| Factory reference | P112DR2620 |
| SKU Cahnos | AMP112DR2620 |
| Material | 100 % Polyglycolic Acid (PGA) synthetic absorbable |
| Coating | Polycaprolactone and calcium stearate |
| Thread structure | Braided multifilament with grooved finish |
| Gauge | USP 2/0 (metric 3) |
| Approximate diameter | 0.30 – 0.349 mm (indicative according to USP/EP) |
| Thread colour | Fluted violet (colourless version available in the range) |
| Thread length | 75 cm |
| Needle shape | 3/8 circle (approx. 135°) |
| Needle body | Cylindrical |
| Needle tip | Taperpoint conical |
| Needle size | DR 26 (26 mm), 1 needle per suture |
| Absorption mechanism | By hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Regulatory standard | USP / European Pharmacopoeia (EP) |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units (minimum order 30 boxes) |
3.2. Tensile strength retention profile after implantation
| Time point | Tensile strength retention |
|---|---|
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60 – 90 days (hydrolysis) |
Complete absorption by hydrolysis between 60 and 90 days, with minimal tissue reaction, excellent tensile strength and optimal knot tying, with a totally predictable and controlled absorption rate.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 2/0 | 3 | 0.30 – 0.349 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Coated braided multifilament polyglycolic acid (PGA) is the most widely used medium-duration synthetic absorbable in the surgical market, with an absorption profile of 60 to 90 days. Homologous families by composition (PGA homopolymer coated with polycaprolactone and calcium stearate), USP 2/0 gauge and violet colour include: SMI Surgicryl® PGA (Belgium), B. Braun Safil® (violet PGA), DemeTECH DemeSORB® PGA (USA), Atramat® PGA (International Farmacéutica) and the PGA lines from Resorba®; the historic Dexon II family from Medtronic/Covidien has become of limited availability.
Having verified the official documentation of the main homologous manufacturers (official SMI catalogue and GUDID/FDA regulatory database for Atramat), no reference has been found that meets the exact configuration of this suture —⅜ circle needle with 26 mm (DR 26) Taperpoint conical point, USP 2/0 gauge, 75 cm violet thread—: the 26 mm taper point presentations available on the market are of ½ circle curvature (e.g. SMI Surgicryl® PGA, order 11300126, USP 2/0 / EP 3, HR 26 mm ½ circle, violet, 75 cm, verified in official catalogue; Atramat® offers 26 mm conical point needles in ½ and 5/8 circle, verified in GUDID) or differ in needle size, so none meets the 1:1 interchangeability tolerance.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. The indicated equivalences are functional category comparisons, not 1:1 interchangeable: before substituting this reference in a specification, each one must be verified individually in the corresponding manufacturer’s technical data sheet: material, USP gauge, needle size (26 mm), ⅜ curvature, point family (Taperpoint conical), thread length (75 cm), colour and coating. The copolymer ranges of polyglactin 910 (PGLA), such as Vicryl® (Ethicon), Safil® 910 / Novosyn® (B. Braun) or Polysorb® (Medtronic), are chemically different from PGA homopolymer (absorption of 56 to 72 days versus 60-90 days) and should not be treated as 1:1 interchangeable without validation from the pharmacy service or the technical department. The cited brands belong to their respective manufacturers and are mentioned only as an equivalence reference.
5. Selection tips and variants of interest
The USP 2/0 gauge (metric 3) offers an ideal balance between initial strength and handling for fine and medium tissues, ligatures, subcutaneous planes and tissue approximation in general surgery. The ⅜ circle needle with cylindrical body and 26 mm Taperpoint conical point is especially recommended in delicate and friable tissues where the cutting effect is to be avoided; if the tissue is firm (skin, fascia, aponeurosis), consider a reverse cutting needle (DS) from the same family. Consult the Cahnos catalogue variants to adjust gauge, quality and geometry:
| Reference | Variant of interest |
|---|---|
| AM112DR2620 | Standard PGA USP 2/0, ⅜ cylindrical DR 26 mm needle, violet, 75 cm (economical alternative) |
| AMP112DR2630 | PGA Prime USP 3/0, DR 26 mm needle, violet, 75 cm (finer gauge) |
| AMP112DS2620 | PGA Prime USP 2/0, ⅜ reverse cutting triangular point DS 26 mm needle, violet, 75 cm (firm tissues) |
| AMP112HR2620 | PGA Prime USP 2/0, ½ circle cylindrical HR 26 mm needle, violet, 75 cm (another curvature) |
| P1L12DR2620 | Polyglactin 910 (PGLA) Prime USP 2/0, DR 26 mm needle, violet, 75 cm (absorption 56-72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes. As this is a sterile single-use product, always check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiries about additional gauges, lengths and colours, contact the technical-commercial team at Cahnos.