Description
1. Commercial Introduction
The absorbable multifilament suture of polyglycolic acid (PGA) reference 128HR3601 (SKU AM128HR3601) from the Premisut range is the professional choice for general surgery departments and procedures requiring secure, manageable closure with controlled absorption. It combines a ½ circle needle with cylindrical body and conical point (taperpoint) of 36 mm (HR 36), a colourless grooved thread of 70 cm in USP 1 gauge (metric 4) and an economical presentation in a box of 12 units, designed for hospital and clinical supply with a minimum order of 30 boxes. A synthetic, sterile, single-use medical device with an excellent quality-price ratio for the continuous supply of your centre.
2. Technical Explanation
Polyglycolic acid (PGA) is a biocompatible synthetic polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This architecture regularises the surface of the thread, favours sliding through tissue and guarantees firm, secure knot tying even in larger diameter gauges such as USP 1.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity, over a period of 60 to 90 days, maintaining tensile support during the first 2 to 4 weeks of healing while the tissue recovers its own strength. The ½ circle needle with cylindrical body and conical point (taperpoint) penetrates by separating the fibres without cutting them, minimising trauma to delicate structures. The colourless colour of the thread offers maximum aesthetic discretion and visual integration in the surgical field. The product is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference AM128HR3601 (128HR3601)
| Characteristic | Specification |
| Reference (SKU Cahnos) | AM128HR3601 |
| Factory reference | 128HR3601 |
| Brand / quality | Premisut (standard range) |
| Composition | 100% polyglycolic acid (PGA), absorbable homopolymer |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 1 (metric 4), conforming to USP/EP Pharmacopoeia |
| Thread colour | Colourless fluted |
| Thread length | 70 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Taperpoint |
| Needle size | HR 36 (36 mm), 1 needle per thread |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis (predictable and controlled rate) |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 1 (this product) | 4 | 0.400 – 0.499 mm |
| 2 | 5 | 0.500 – 0.599 mm |
| 0 | 3.5 | 0.350 – 0.399 mm |
| 2/0 | 3 | 0.300 – 0.339 mm |
The diameter of the strand complies with the official ranges of the USP Pharmacopoeia for synthetic absorbable sutures, without extending its limits.
3.4. Decoding of the official reference
Official reference 128HR3601: 128 (colourless 70 cm thread series) + HR36 (½ circle needle with cylindrical body and conical point of 36 mm) + 01 (USP 1 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or bidding.
4. Equivalences and Cross-References
Polyglycolic acid (PGA) is the most well-established medium-duration braided multifilament absorbable on the market and shares an absorption profile (60-90 days), coated braided structure and clinical behaviour with homologous families in the sector. There is no equivalence table validated by the manufacturer that cross-references this reference with other brands; the following matches are functional comparisons at category level and should not be treated as 1:1 interchangeable without prior technical validation.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| SMI (Surgicryl) | Surgicryl PGA – ½ circle needle cylindrical body conical point (round bodied taper point) HR36, USP 1 gauge | SM1140 0136 (HR36 mm, 90 cm) | Comparable by category (not interchangeable 1:1) | Confirmed in the official SMI catalogue and authorised distribution: same polymer (100% PGA), same coating (polycaprolactone and calcium stearate), same USP 1 gauge and same HR36 ½ circle needle with conical point; they differ in thread length (90 cm vs 70 cm) and colour (violet vs colourless). Validate before bidding. |
| Ethicon (Johnson & Johnson MedTech) | Coated VICRYL® – coated braided polyglactin 910, CT-1 needle 36 mm, ½ circle, conical point | J347H – USP 1, 90 cm | Comparable by category (not interchangeable 1:1) | Confirmed in the Ethicon catalogue and authorised distribution: they match in USP 1 gauge, ½ circle curvature, 36 mm needle and conical point family; they differ in polymer (polyglactin 910 vs 100% PGA), absorption profile (56-72 days vs 60-90 days), length (90 cm vs 70 cm), colour (violet vs colourless) and presentation (box of 36 units vs 12 units). |
There is no equivalence table validated by the manufacturer; consult the technical department before using in bidding. Homologous ranges in the medium-absorption braided PGA segment include Dexon II (Medtronic/Covidien), Surgicryl PGA (SMI), DemeSORB PGA (DemeTECH) and Atramat (Resorba), as well as the polyglactin 910 lines Coated VICRYL® / VICRYL® Plus (Ethicon), Safil (B. Braun), Polysorb (Medtronic) and Assucryl 910 (Assut Sutures), comparable by category by material and absorption profile; for the exact combination USP 1 / HR 36 mm conical point needle / 70 cm, their specific codes must be confirmed with the manufacturer before any substitution, verifying material, gauge, curvature, needle type and size, thread length, colour and configuration. This geometry (HR36, USP 1, 70 cm) is fully safely interchangeable only within the Premisut system itself.
5. Selection tips and variants of interest
USP 1 gauge (metric 4) is indicated for tissues requiring high initial strength: aponeurosis, fascia, muscular planes, abdominal wall and high-tension closures. The ½ circle cylindrical needle with conical point (taperpoint) of 36 mm is atraumatic and ideal for soft and delicate tissues; if you need to penetrate firmer and denser tissues (skin, sclera), consider the reverse cutting (HS) or trocar (HRT) points available in the same range.
| Reference | Variant |
| AM128HR3601 (128HR3601) | PGA USP 1, HR 36 mm needle, colourless, 70 cm (current product) |
| AM118HR3601 (118HR3601) | PGA USP 1, HR 36 mm needle, violet, 70 cm (greater visibility in the field) |
| AMP128HR3601 | PGA Prime USP 1, HR 36 mm needle, colourless, 70 cm (superior quality, same geometry) |
| AM123HR3601 (123HR3601) | PGA USP 1, HR 36 mm needle, colourless, 90 cm (longer thread length) |
| 1L28HR3601 | PGLA polyglactin 910 USP 1, HR 36 mm needle, colourless, 70 cm (absorption 56-72 days) |
| P1L28HR3601 | PGLA Prime USP 1, HR 36 mm needle, colourless, 70 cm (premium range, copolymer) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare facility. If you require a longer strength retention (180-210 days), consider the polydioxanone (PDO) family from Cahnos. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, keep the reference alongside the batch number for traceability in the operating theatre, and store in a cool, dry place, protected from sunlight. For B2B volume conditions or hospital supply agreements, contact the technical-commercial team at Cahnos.