Description
1. Commercial Introduction
The absorbable monofilament polydioxanone (PDO/PDS) suture from the Prime line is the reference choice when the operating theatre needs prolonged tissue support, fully predictable absorption and exceptional thread handling. This presentation combines a 30 cm violet thread in USP 4/0 gauge with a double straight taper-point 19 mm needle (2xGR19), an open surgery configuration designed for precise manual suturing on accessible tissue planes.
Available in a box of 12 sterile units, it is the B2B supply solution for general surgery departments, paediatric cardiovascular surgery units, ophthalmology services and hospital purchasing centres seeking homogeneity of quality, full traceability and consistent performance in every procedure.
2. Technical Explanation
Polydioxanone is an absorbable synthetic polymer —poly(p-dioxanone)— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that ensures highly reproducible absorption kinetics between individuals, with complete absorption within 180 to 210 days: one of the longest support windows among absorbable sutures.
As a non-channelled monofilament, the thread has no wicking effect or interstitial spaces: it reduces bacterial adherence, facilitates sliding through the tissue and minimises the inflammatory response compared with braided threads. The straight needle with cylindrical body and conical point (Taperpoint type) passes through tissue separating the fibres without cutting them, making it especially suitable for soft tissues, fasciae and delicate structures where minimal trauma is desired. The double-needle configuration (one needle at each end of the thread) allows continuous suturing in both directions without rethreading, speeding up the closure of accessible tissue planes. The violet colour improves thread visibility on the surgical field. Product sterile by ethylene oxide and single-use, not resterilisable.
3. Technical Specifications
3.1. Reference tariff data
| Needle size (mm) | USP | Colour | Thread length (cm) | Reference | Box (units) |
| 2xGR19 prime | 4/0 | Violet | 30 | P414GR1950-2 | 12 |
3.2. General characteristics
| Parameter | Specification |
| Factory reference / SKU | P414GR1950-2 / AMP414GR1950-2 |
| Material | Polydioxanone (PDO/PDS), synthetic polymer of poly(p-dioxanone) |
| Thread structure | Monofilament, non-channeled |
| USP / metric gauge | 4/0 (metric 2) |
| Thread length | 30 cm |
| Thread colour | Violet (high visibility) |
| Needle shape | Straight |
| Needle body and point | Cylindrical, taperpoint conical point |
| Needle size | 2xGR19 – double straight 19 mm needle per thread |
| Absorption mechanism | Hydrolysis (180-210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
3.3. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 4 weeks | 60-70% of initial strength |
| 8 weeks | 45% of initial strength |
| Complete absorption | 180-210 days by hydrolysis, with minimal tissue reaction |
This prolonged force retention profile maintains the necessary support during the critical healing phase of slow-healing tissues, which is why it is recommended for closures of tissue planes under sustained tension and in paediatric cardiovascular surgery, where tissue growth is expected.
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market’s reference lines for long-lasting absorbable sutures (PDS-type family). For the verified configuration in this data sheet —double straight cylindrical taper-point 19 mm needle, USP 4/0 gauge, 30 cm violet thread— the following homologous reference, comparable at category level and not interchangeable 1:1, has been cross-checked against primary sources (official manufacturer catalogue and authorised distributors):
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II® (polydioxanone), double straight cylindrical needle, cylindrical point ST-4, 19 mm (2xST-4), USP 4/0, violet monofilament | Z620E | Comparable by category (not interchangeable 1:1) | Same polymer (polydioxanone), same USP 4/0 gauge, same double straight 19 mm needle configuration and same violet colour. Difference: 35 cm thread (Ethicon) versus 30 cm for this reference; absorption profile according to the Ethicon data sheet of 182-238 days versus 180-210 days for the Prime line. Confirmed in the Ethicon catalogue (EMEA) and authorised distributors. |
The other homologous families on the market —Ethicon’s PDS Plus® line, the MonoPlus® line from B. Braun (polydioxanone monofilament), SMI’s Surgicryl® PDO line and, with a different polymer (polyglyconate), the Medtronic Maxon range— must be verified code by code in their official technical data sheet before any substitution: USP gauge, thread length, colour, curvature, body, point and needle size, since small variations prevent a 1:1 equivalence. There is no equivalence table validated by the manufacturer for this specific configuration; consult the technical department before use in tenders. The brands cited belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin, subcutaneous tissue and delicate planes —like this reference—, 3/0 and 2/0 for general soft tissue, and USP 1 and above for fascia and planes under high tension. Needle geometry is also decisive: the conical point (GR) is reserved for soft and delicate tissues and separates the fibres without cutting them, the trocar point (GRT) for firm tissues of intermediate resistance, and conventional cutting (GS) for superficial fibrous tissues. The double-needle configuration provides versatility by allowing suturing in both directions.
Within the Cahnos catalogue you will find variants of interest from the PDO/PDS family:
| Reference | Variant of interest |
| P414GR1950 (SKU AMP414GR1950) | Straight taper-point GR19 needle (19 mm), USP 5/0, violet, 30 cm – single end, finer gauge |
| P414GRT3840-2 (SKU AMP414GRT3840-2) | Double straight trocar-point 38 mm needle (2xGRT 38), USP 4/0, violet, 30 cm |
| AMP418GS6040 | Straight conventional cutting GS60 needle (60 mm), USP 4/0, violet, 70 cm |
| AMP418GS6030 | Straight conventional cutting GS60 needle (60 mm), USP 3/0, violet, 70 cm |
| AMP411GS1340 | Straight conventional cutting GS13 needle (13 mm), USP 4/0, violet, 45 cm |
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and keep the reference and batch number for traceability in the operating theatre. Store in a cool, dry place below 30 ºC and protected from sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalents from other commercial brands, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.