Description
Reusable silicone electrosurgical plate with REM — Ref. F7930
We present the reusable silicone electrosurgical plate with REM system, a solution designed for professionals who need a reliable, durable and reusable return electrode for their monopolar electrosurgery. Its compact design of 200 x 150 mm with 250 cm² of surface area offers an excellent balance between contact safety, patient comfort and cost per use compared with disposable plates, with the advantage of being able to be cleaned and reused in multiple procedures.
It is the usual choice in operating theatres, minor surgery clinics, dermatology, dentistry and veterinary clinics that work with high-frequency electrosurgical equipment and wish to reduce medical waste and the recurring cost of single-use electrodes.
Technical explanation
In monopolar electrosurgery, the high-frequency current flows from the active electrode (pencil) through the patient to a return plate (neutral electrode) that closes the circuit. If the contact of that plate with the skin is not correct, current density concentrates on a reduced surface and a burn may occur on the patient.
The REM system (Return Electrode Monitoring) is the return electrode monitoring technology associated with Valleylab-type generators and other equipment with split-plate control: the generator continuously monitors contact impedance and blocks emission with an alarm if it detects poor contact. This plate, designed to work with the REM system, has a 4 mm female socket: it is supplied without cable, so it is connected to the electrosurgical unit via the reusable plate cable most suitable for each device (see table of compatible cables). Universal bipolar cables with double 4 mm plug (ref. F7922/24) or Valleylab type with REM (ref. F7923) allow its use with numerous electrosurgical units, including Valleylab, Bowa, Conmed and Erbe equipment (depending on connector configuration).
As it is reusable, it must be cleaned after each use, visually inspected (no cuts, cracks or deterioration of the conductive surface) and continuity checked periodically before each procedure; it must be replaced at any sign of degradation.
Technical specifications
| Characteristic | Specification |
|---|---|
| Reference (SKU) | F7930 |
| Product type | Reusable silicone electrosurgical plate (return electrode) |
| Material | Medical-grade conductive silicone, latex-free |
| Dimensions | 200 x 150 mm |
| Surface (area) | 250 cm² |
| Monitoring system | REM (Return Electrode Monitoring), for generators with return electrode control |
| Cable | Without cable (cable-free design; requires connection cable, not included) |
| Connection | 4 mm female socket |
| Type of use | Reusable (multiple uses) |
| Presentation | Price per unit |
| Compliance | Medical device; confirm CE marking and classification in the manufacturer’s original data sheet (EU Regulation 2017/745) |
Connection cables for reusable plates compatible with the 4 mm socket of this plate (Cahnos catalogue):
| Reference | Description |
|---|---|
| F7922 | Universal cable, 6.3 mm mono jack connector, 500 cm |
| F7922/4 | Cable with 4 mm plug, 500 cm |
| F7922/24 | Cable with double 4 mm plug, 500 cm |
| F7923 | Valleylab type cable (ESU with REM), 500 cm |
| F7923/F | Valleylab type cable (ESU without REM), 500 cm |
| F7924 | Special 6.3 mm mono jack cable for Erbe (ACC, ICC, VIO), 500 cm |
Equivalences and cross-references
This category of reusable silicone return electrodes is marketed internationally under the generic name of reusable silicone neutral plates with REM connector. Among the homologous families at category level are the reusable silicone plates from the surgical electrode range of manufacturers such as Megadyne (capacitive range Mega Soft / Mega 2000, based on capacitive coupling principle and flexible surface not 1:1 interchangeable with a conductive silicone plate), the reusable silicone plates from Bovie/Apyx and the reusable silicone electrodes with REM connector from Asian manufacturers such as YKD Medical (reusable plate model TS111, pending verification of dimensions and configuration against official source), as well as reference BMH-EDC0001A (reusable silicone return plate with distal REM connector, seen at an authorised distributor: 300 x 180 mm and integrated 3 m cable, pending confirmation and not 1:1 interchangeable due to differences in dimensions, surface and absence of 4 mm socket). In all cases, before replacement, verification must be carried out field by field: material, dimensions, active surface, connector compatibility and generator monitoring system.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
Selection tips and variants of interest
- Non-REM version: if your electrosurgical unit does not have return electrode monitoring, consider the reusable silicone plate of 200 x 150 mm (266 cm²) without REM system, with the same 4 mm socket (ref. F7915), from the same technical family.
- Cable selection: for equipment with REM system preferably use the Valleylab-type cable with REM (F7923) or the double 4 mm cable (F7922/24); for equipment without REM, use cable F7923/F or F7922/4. Always confirm the connector type of your generator before choosing the cable.
- Generator compatibility: verify that your electrosurgical unit accepts reusable plates with a 4 mm connection and REM system (Valleylab, Bowa, Conmed, Erbe types in international configurations with the appropriate cable).
- Maintenance: clean the plate after each use with the products recommended by the manufacturer, dry it completely and store it without folds; check the conductive surface before each procedure and replace the plate if it shows cracks, cuts or loss of conductivity.
- Placement: apply the plate to healthy, dry and well-shaved skin, as far as possible from the surgical site and avoiding areas with hair, scars or bony prominences to ensure uniform contact.
- Institutional purchases: for operating theatre equipment, veterinary clinics or volume orders with special B2B conditions, contact our technical-commercial department.




Reviews