Description
Single-use REM electrosurgical plate — Ref. 30495
The return plate Ref. 30495 (SKU GM30495) is the disposable neutral electrode that completes the monopolar circuit of your electrosurgical unit when the generator requires a split plate with contact control (REM system). With a non-woven fabric backing and solid gel adhesive, it is designed for use on adult and paediatric patients, ensuring a uniform distribution of the return current and a high level of safety against thermal burns during the procedure. It is supplied in a box of 100 units, each individually packaged.
Technical explanation
In monopolar electrosurgery, the high-frequency current emitted by the active electrode (handle or pencil) passes through the patient’s tissue and must return to the generator via the neutral plate, also called the dispersive or return electrode. If the skin-plate contact is poor or the conductive surface is insufficient, the current density at the return point increases and, with it, the risk of burns. This plate incorporates the REM (Return Electrode Monitoring) system: its surface is divided into two sectors, allowing the generator to continuously monitor contact impedance and block emission in the event of partial detachment, dry gel or any failure in contact quality. The non-woven fabric backing, combined with the solid gel, promotes adaptation to the body contour and a uniform, stable contact.
The plate is supplied without cable: it requires the connection cable for disposable plates with tab Ref. 30563 (SKU GM30563), a universal 5 m bipolar cable with a 6.3 mm mono jack connector fitted with a bipolar forceps for 22-25 mm tabs.
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 30495 |
| Sales unit | Box of 100 units |
| Device type | Medical device (Medical device) |
| Device class | II B |
| CND | K02010202 |
| GMDN | 58494 |
| EAN | 8054242442557 |
Technical specifications
| Plate type | Disposable single-use return electrode (neutral plate) |
|---|---|
| Safety system | REM (Return Electrode Monitoring), split plate |
| Support | Non-woven fabric |
| Adhesive gel | Solid gel |
| Population | Adult and paediatric |
| Cable | Not included; requires cable Ref. 30563 (GM30563, bipolar forceps for 22-25 mm tabs, universal 6.3 mm jack connector, 5 m) |
| Presentation | Box of 100 units, individually packaged |
Equivalences and cross-references
Reference 30495 (SKU Cahnos GM30495) is the manufacturer’s original catalogue designation for this non-woven REM return plate. No reliable sources have been found for a 1:1 equivalence table validated by the manufacturer cross-referencing this reference with codes from other brands; any substitution must be validated field by field (REM configuration, conductive surface, gel type and connector compatibility with the installed generator). Within the catalogue itself, this plate is part of the family of disposable return electrodes with tab connection that uses the cable Ref. 30563 (GM30563), which serves the disposable tab plates of Refs. 30490-30495. Related references from the same REM family include the 3M split plates (Refs. 30496, 30497 and 30498) and the complementary connection cables (Refs. 30546 and 30547, for other generator configurations).
Selection tips and variants of interest
Always verify that your electrosurgical unit has a socket for a plate with REM control and that the cable connector is compatible with the plug of your generator (the GM30563 cable incorporates a universal 6.3 mm mono jack and is compatible with numerous electrosurgical devices on the market: Aesculap, Bovie, Martin, Philips, Erbe and others). As this is a cable-free plate, complete the set with the connection cable Ref. 30563, reusable between patients, which optimises the cost per procedure. If your activity requires hospital consumption volume, the box of 100 units is the ideal format for surgical scheduling. Always apply the plate to dry, intact, hair-free skin, on a well-irrigated muscular area away from bony prominences, pacemakers or metal implants; check the integrity of the individual packaging before each use. Single-use product, not reusable: request B2B conditions and supply times from our technical-commercial team. For large orders you can request a quote.