Description
1. Introduction
The Neuropacer nerve locator (ref. F1745) is a disposable, sterile, self-powered nerve stimulator developed to identify exposed motor nerves during microsurgery and protect them from any accidental damage. It is especially indicated for plastic surgery and head, neck and hand surgery, where preservation of nerve branches is critical for functional and aesthetic outcome.
It is supplied complete with active electrode and subcutaneous reference needle, ready for use in the operating theatre; each box contains 10 units in individual packaging. A reliable and cost-effective solution for surgical departments that need immediate availability of a nerve locator in every procedure.
2. Technical explanation of the equipment
Neuropacer generates an alternating and pulsating current which, applied through the active electrode onto the tissue, causes a visible contraction of the muscle innervated by the motor nerve to be located. The subcutaneous reference needle is placed near the surgical field and acts as a return electrode, closing the electrical circuit in a controlled and homogeneous manner.
The device incorporates a safety system that allows it to be switched on only immediately before use, preventing accidental activations, and its short-circuit current is limited to 0.6 mA (±10%) for effective and safe stimulation. It is classified as a medical device class IIa (Directive 93/42/EEC) with applied parts type BF according to EN 60601-1, and is sterilised by ethylene oxide.
3. Technical Specifications
| Characteristic | Technical data |
|---|---|
| Product type | Nerve locator / nerve stimulator, sterile, single-use, disposable |
| Manufacturer reference | F1745 (Neuropacer) |
| Terms of use | Continuous operation |
| Maximum short-circuit current | 0.6 mA ± 10% |
| Power supply | 2 D357H 1.5 V batteries (included) |
| Device length | 150 ± 2 mm |
| Device diameter | 18 ± 0.5 mm |
| Active electrode angle (probe) | 30° ± 15% |
| Active electrode tip diameter | 0.2 ± 0.05 mm |
| Electrode | Active (probe), stainless steel |
| Reference needle | Subcutaneous (earth), stainless steel; cable + needle length 725 ± 30 mm |
| Body material | Plastic |
| Sterilisation | Ethylene oxide (EO) |
| Regulatory classification | Medical device class IIa (Directive 93/42/EEC); EN 60601-1, internal power supply, applied parts type BF |
| Presentation | 10 units in a cardboard box, each in individual heat-sealed tyvek-PE packaging |
| Storage and use | 0 °C to +50 °C; relative humidity 20% to 80% |
| Expiry | 48 months from the date of production |
4. Equivalences and Cross-References
Neuropacer F1745 is the manufacturer’s original reference for this sterile single-use nerve stimulator with active electrode and reference needle. There is no official equivalence table validated by the manufacturer cross-referencing this reference with other models; the third-party references listed below belong to the same product category (disposable, sterile, battery-powered nerve locators for identifying exposed motor nerves in head, neck, hand and plastic surgery) and have been verified in regulatory databases (GUDID) and manufacturers’ catalogues. As they differ in design, output intensities, response indicator or presentation, they are labelled as comparable by category: not interchangeable 1:1 without prior technical validation.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Aspen Surgical (Bovie) | Neuro-Pulse™ Surgical Nerve Locator, red LED, box of 10 | 0003 | Comparable by category (not interchangeable 1:1) | Single-use, sterile, battery-powered, for identifying exposed motor nerves in head, neck, hand, plastic and facial surgery; 3 output levels (0.5/1/2 mA at 3 V DC) and LED indicator. Confirmed in GUDID (DI 10840113234728) and Aspen Surgical. |
| Aspen Surgical (Bovie) | Neuro-Pulse™ Nerve Locator, yellow LED, box of 10 | 0003Y | Comparable by category (not interchangeable 1:1) | Same family as 0003 with yellow LED and 3 output levels (0.5/1/2 mA at 3 V DC). Confirmed in Aspen Surgical and McKesson catalogues. |
| Checkpoint Surgical | Guardian® Nerve Stimulator | 9095 | Comparable by category (not interchangeable 1:1) | Single-use sterile nerve stimulator for locating and identifying exposed motor nerves and assessing excitability; individual-use presentation. Confirmed in GUDID (B13090950) and manufacturer’s website. |
| Checkpoint Surgical | Gemini® Bipolar Nerve Stimulator | 9092 | Comparable by category (not interchangeable 1:1) | Single-use sterile bipolar stimulator: bipolar geometry versus Neuropacer’s active electrode + subcutaneous reference needle; not interchangeable 1:1. Confirmed in GUDID (B13090920) and manufacturer’s website. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
Correct indication: Neuropacer is designed for intraoperative identification of exposed motor nerves in open surgery or microsurgery (head, neck, hands, plastic surgery). It is not a stimulator for regional anaesthesia or nerve blocks with an insulated needle: those techniques require specific stimulating needles connected to an anaesthesia neurostimulator.
Before replacing with another brand: check the candidate’s technical data sheet for the type of stimulation (pulsating alternating current as opposed to adjustable output), the number and level of intensities, the presence of an active electrode and reference needle (or unipolar/bipolar design), the presentation (single unit or box of 10) and the sterility of the individual packaging.
Use and storage: single-use device: do not reuse or re-sterilise. Store it at 0–50 °C with relative humidity of 20% to 80%, in its original packaging until use; the shelf life is 48 months from the date of production. For institutional orders, tenders or continuous supply contracts, contact our technical-commercial department to confirm compatibility and a formal quotation.