Description
1. Connection cable for electrosurgical plates: compatible with your electrosurgical unit
The connection cable ref. 30563 is the linking accessory that allows you to connect disposable electrosurgical plates with tab to your electrosurgical unit, safely closing the monopolar return circuit. With a length of 5 metres and a universal 6.3 mm mono jack connector, it offers a wide working radius and freedom to place the plate on the patient during the procedure. It is the practical solution for operating theatres, minor surgery clinics, ambulatory surgery centres and veterinary clinics that work with disposable tab plates.
2. Technical explanation
In monopolar electrosurgery, the high-frequency current leaves through the active electrode (handle or pencil), passes through the patient’s tissue and must return to the generator via the neutral plate or dispersive electrode. This universal bipolar cable transmits that return current from the disposable tab plate to the electrosurgical unit, ensuring a stable electrical contact and a low current density in the skin-plate contact area, an essential requirement to avoid burns and ensure the correct operation of return monitoring systems. The cable incorporates a bipolar forceps for connection to disposable plates with a 22-25 mm tab and a sliding metal forceps that allows the cable to be fixed to the surgical field during the operation, keeping it tidy and out of the working area. It is latex-free and is supplied individually packaged in a sealed polyethylene bag.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Manufacturer reference | 30563 |
| SKU Cahnos | GM30563 |
| Product type | Connection cable for electrosurgical plates |
| Cable length | 5 m |
| Generator connector | Universal mono jack 6.3 mm |
| Plate connection | Bipolar forceps for disposable plates with tab (22-25 mm) |
| Fixation | Sliding metal forceps for surgical field |
| Latex-free | Yes |
| Presentation | Individual packaging in sealed polyethylene bag |
| Regulatory standard | Class I medical device in accordance with Directive 93/42/EEC |
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 pc |
| Device type | Medical device |
| Device class | I |
| CND | Z12010980 |
| GMDN | 46906 |
| EAN | 8033003263167 |
4. Equivalences and Cross-References
Reference 30563 (SKU Cahnos GM30563) uniquely identifies this connection cable for plates. No official equivalence table validated by the manufacturer has been found in reliable sources that cross-references this reference with 1:1 cables from other brands; any substitution must be validated field by field (connector, length, forceps type and generator compatibility) with the technical department of Cahnos. The compatibility declared by the manufacturer includes the following electrosurgical generators:
- Aesculap: TM, TM401
- Alexandra: D205, E300, 500D, 102/201, 2020/E, 202, MICRO180
- ALSA: Tutte MF100, MF360/A360, 380100, 180
- Berchtold: Elektrotom 200, 390, 400, 540
- Birtcher: 772, 773, 737, 737XL, 774
- Bovie: Bovie CSV, URO, BS, 400/400, 400SR, X10, X10L
- Burdick: U-7, SU-8; Cameron-Miller: 26-1290; Cobe: TS-403
- Concept: 9600, 9700, 9800, 9900; Dolley: COBII 203T, 502, 503T, 505T
- Ems Davol: System 2000, 5000; Faset: Serie COBI
- Gima, GPS, Lamidey Series Subtilec/Surgilec, Led, Martin: all models
- I Cuatro: Megeri1; Neomed: Serie 3000, 3000A, 3002, 3006, 3010, 3020
- Philips: TS-403; Ritter/Mdt: Solide State 400, 400B, 400SR
- Sidevan: Seb2; Soxil: BP 2000, X10; Thackray: T300S; Torry: Lem 300 CST
5. Selection tips and variants of interest
Check your generator plug: this cable incorporates a universal 6.3 mm mono jack connector, the most widespread standard in electrosurgical units. Make sure your disposable plates have a 22-25 mm tab, the format for which the bipolar forceps of this cable is designed. Use the sliding metal forceps to fix the cable to the surgical field and keep the working area clear. Within the electrosurgery range, Cahnos offers other complementary accessories, such as disposable plates with and without REM system, plate cables for reusable rubber plates (ref. 30575) and reusable neutral plates; to complete your system, consult our technical-commercial team. Product without public price: request B2B conditions and delivery times from our team. For large orders you can request a quote.






Reviews