Description
1. Commercial Introduction
The SibelSound 400-Supra digital clinical audiometer (ref. 33626, SKU Cahnos GM33626) is the most complete model in the SibelSound 400 series: a professional audiometry station that combines air conduction, bone conduction, masking, supraliminal tests and speech audiometry, together with the W50 audiometry software and the intercom kit, all included as standard. It is designed for ear, nose and throat (ENT) clinics, audiology services, hearing centres and occupational health units that need a reliable, fast clinical device with computerised management of results.
2. Technical explanation of the equipment
This is a two-channel audiometer based on digital signal processing (DSP), designed for quick and easy exploration of hearing thresholds and for screening tests and supraliminal tonal tests (Fowler-ABLB, Tone Decay, Weber, Lüscher, etc.). The device allows exploration via air conduction with TDH 39 headphones in the range of 125 to 8,000 Hz (or up to 20,000 Hz in high frequency with the optional HDA 200 headphones) and via bone conduction with the B71 vibrator. It incorporates manual or synchronised masking with narrow band noise (NBN), white and speech noise, SISI test, speech audiometry, musical frequencies and automatic calculation of hearing loss with diagnosis according to the COUNCIL, IMSS (Mexico), ELI, SAL, Klockhoff and MOH protocols, among others.
The communication system is one of the model’s strong points: it allows direct printing to a USB printer or connection to the PC via USB cable to the W50 software (included) to view, archive, transmit, analyse and enter audiometric tests. The software also accepts data from other audiometers, displays automatic diagnosis and is compatible with Windows 7, 8.1 and 10 in 32 and 64-bit versions. It also includes an intercom kit (headphones + microphone) with a database of 1,000 tests.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Audiometry type | 2 clinical (two channels, DSP) |
| Airway | TDH 39: 125 – 8,000 Hz |
| High frequency (optional) | HDA 200: 8,000 – 20,000 Hz |
| Bone conduction | B71 |
| Levels | Air conduction: -10 to 120 dB HL / Bone conduction: -10 to 70 dB HL / Speech: -10 to 100 dB HL / Free field: -10 to 80 dB HL |
| Masking (manual or synchronised) | NBN air conduction: -10 to 100 dB HL / NBN bone conduction: -10 to 70 dB HL / White and speech: -10 to 100 dB SPL |
| Tone | Continuous, pulsed, alternating, amplitude modulation |
| Level increments | 5 dB (400-A/400-AOM), 1 or 5 dB (Supra) |
| Accuracy | Frequency ±1% / Level ±2% dB SPL |
| Tests included | SISI, supraliminal tests (Fowler-ABLB, Tone Decay, Weber, Lüscher), speech audiometry, musical frequencies |
| Automatic diagnosis | COUNCIL, IMSS (Mexico), ELI, SAL, Klockhoff, MOH and others |
| Intercom kit | Headphones + microphone; database of 1,000 tests |
| Connectivity | USB to PC and USB printer |
| Software | W50 audiometry software (included); compatible with Windows 7, 8.1 and 10 (32/64-bit) |
| Display | Backlit alphanumeric LCD, 2 lines of 16 characters |
| Dimensions | 390 x 260 x 105 mm |
| Weight | 2.4 kg (without accessories) |
| Power supply | 100 – 240 V, 50/60 Hz |
| User manual | GB, IT (FR and ES available on request) |
Product data
| Brand | SIBELSOUND |
|---|---|
| Manufacturer reference | 33626 |
| SKU Cahnos | GM33626 |
| Sales unit | 1 pc |
| Device type | Medical device |
| Device class | II A |
| CND | Z121401 |
| GMDN | 10228 |
| EAN | 8436035522026 |
4. Equivalences and Cross-References
This product corresponds to the manufacturer’s original reference SibelSound / Ref. 33626, which Cahnos markets as GM33626. Within the same SibelSound 400 family, the following variants are available, sharing the same chassis, DSP and communication system, differing in the scope of tests:
| Model | SKU Cahnos | Reference | Features | W50 software |
|---|---|---|---|---|
| SibelSound 400-A (screening) | GM33623 | 33623 | Air conduction; intercom database of 150 tests | Optional |
| SibelSound 400-AOM (diagnostics) | GM33624 | 33624 | Air conduction + bone conduction + masking; database of 1,000 tests | Optional |
| SibelSound 400-Supra (clinical, this product) | GM33626 | 33626 | Air conduction + bone conduction + masking + supraliminal tests + SISI + speech audiometry + intercom kit | Included |
Audiometers from other manufacturers cover comparable performance ranges (e.g., Interacoustics, Amplivox or Otometrics in two-channel clinical devices), but there is no validated 1:1 equivalence table with the SibelSound 400 series; the W50 software is supplied linked to the series and, in its optional version (ref. 33629), corresponds to models marketed up to March 2023. Consult the technical department of Cahnos to confirm compatibilities or specific configurations.
5. Selection tips and variants of interest
Choose the 400-Supra if your centre performs advanced clinical audiometry: supraliminal tests (Fowler-ABLB, Tone Decay, Weber, Lüscher), SISI test, speech audiometry and musical frequencies, with the W50 software and intercom kit already included. For occupational medicine or hearing screenings with air conduction, the 400-A model (GM33623) may be sufficient; for intermediate diagnostic needs with bone conduction and masking, the 400-AOM (GM33624) is the indicated alternative. Remember that the high-frequency headphones HDA 200 are optional and extend the examination up to 20,000 Hz. For large orders you can request a quote: contact the technical-commercial team at Cahnos to receive a no-obligation offer tailored to your project.






Reviews