Description
1. Reference bone densitometry for your centre
The DEXA bone densitometer with pencil beam, model Inus D, is an axial dual-energy X-ray absorptiometry (DEXA) system designed for the diagnosis and follow-up of osteoporosis and other bone mineral density pathologies. It can scan the lumbar spine L1-L4, the dual femur and the forearm, and adds value-added functions such as FRAX® fracture risk and body fat analysis.
Designed for practices, diagnostic clinics and hospitals with limited space or moderate patient volume, the Inus D combines the diagnostic gold standard (DEXA) with a reduced footprint thanks to its side panels and folding head/foot sections, and a low radiation dose for the patient and operator by using a point emitter (pencil beam).
2. Technology and scanning procedure
The Inus D system uses DEXA technology with a pencil beam: a fine dual-energy X-ray beam continuously scans the study area (Non-Stop Scan mode) while a point detector records the beam attenuation. Bone mineral density parameters are calculated from this attenuation: BMD (bone mineral density), BMC (bone mineral content), area and tissue thickness, expressed as the clinical indices T-score and Z-score, in accordance with WHO diagnostic criteria.
Lumbar spine and femur scanning follows the worldwide gold standard for the diagnosis of osteoporosis. The patient is positioned using a cross laser alignment system, and the region of interest (ROI) is determined automatically to improve reproducibility between scans. The system also includes specific paediatric and orthopaedic functions, and generates complete reports with high-resolution images.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| System / X-ray source | DEXA (dual-energy X-ray absorptiometry) |
| Scan type | Pencil beam, Non-Stop Scan cycle |
| Scan sites | Lumbar spine L1-L4, dual femur, forearm; FRAX® and body fat analysis (LWS) |
| Fast scan time | L1-L4 approx. 1.5 min; femur approx. 1 min |
| Normal scan time | L1-L4 approx. 3 min; femur approx. 2 min |
| Accuracy | < 1 % |
| Captured parameters | BMD, BMC, area, tissue thickness; T-score and Z-score; automatic ROI |
| Calibration | Automatic via measurement system and daily control phantom (QC) |
| Patient positioning | Automatic laser aid (cross alignment) |
| Image storage | DICOM, PACS compatible |
| Operating system | Windows 7 / 10 (PC included) |
| Monitor | LCD, 1280 x 1024 pixels |
| Printer | Colour (inkjet), included |
| Dimensions | 1900-2030 x 800-810 x 1210-1220 mm (head and foot extended); 1260 x 810 x 1220 mm with sections folded |
| Weight | 130.5-131 kg |
| Power supply | AC 220-240 V, 50/60 Hz |
| Operating temperature | 15-40 °C; humidity 20-80 % |
| Electrical safety | Class I, Type B |
| Additional modes | Paediatric and orthopaedic |
Delivery contents:
- DEXA scanning unit with folding panel stretcher.
- Complete workstation: PC with software, LCD monitor, colour printer, mouse and keyboard.
- Equipment table/column.
- Installation, commissioning and operating instructions.
4. Equivalences and Cross-References
Below are DXA systems from other manufacturers with a comparable technical profile (pencil beam technology and spine, femur and forearm scanning). The data have been cross-checked against each manufacturer’s official catalogues; however, there is no equivalence table validated by the manufacturer; consult the technical department before using in a tender.
| Manufacturer / Brand | Family / Model | Code / Reference | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Osteosys | DEXXUM T / 3 | DEXXUM T (catalogue model reference) | Comparable by category (not interchangeable 1:1) | Central pencil-beam DXA with AP spine, femur and forearm sites, FRAX and abdominal fat thickness; CV reproducibility ≤ 1.5% (the Inus D declares < 1%); triple-section folding table; numeric SKU not publicly published: confirm with authorised distributor. |
| DMS Imaging (Medilink) | MEDIX90 | MEDIX90 (catalogue model reference) | Comparable by category (not interchangeable 1:1) | Enhanced pencil-beam DXA (Digital Fast Beam), with whole-body application in addition to spine, femur and forearm; shorter scan times; classified as Class IIb in the EU; numeric SKU not publicly published: confirm with authorised distributor. |
| Norland (Swissray) | XR-46 | XR-46 (model verified in FDA/GUDID registration K992125) | Comparable by category (not interchangeable 1:1) | Central pencil-beam DXA densitometer for hip, spine, forearm and whole body; automatic calibration in 77 steps; the FRAX® function is not confirmed on this model: verify with the manufacturer. |
In the high-end segment with fan-beam technology are, among others, the GE HealthCare Lunar (Prodigy/iDXA) and Hologic Horizon systems, comparable at category level for central densitometry, but with a different beam geometry, higher scan speed, higher acquisition cost and different room requirements: they are not 1:1 interchangeable with the Inus D.
There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender.
5. Selection tips and variants of interest
- For screening and moderate volume: the Inus D (reference 25.41-3100) is an entry-level solution for complete central densitometry with diagnostic DEXA quality; its compact size and folding panels make it suitable for rooms with limited dimensions.
- As a peripheral screening complement: the ultrasound densitometer Inus S (reference 25.41-1300, available at Cahnos) offers rapid heel measurements in about 15 seconds, ideal for primary care points, health fairs or as prior triage.
- High patient volume: if your centre expects a large number of daily studies or needs advanced whole-body analysis, consider high-end fan-beam systems (GE Lunar, Hologic) or the MEDIX90 from DMS Imaging, assuming greater investment and space.
- Digital integration: verify DICOM/PACS compatibility with your hospital network and the interface language before installation; Cahnos includes installation, commissioning and instruction in use.
- Radiological protection: complement the installation with shielding, aprons and lead-lined protectors from the radiology range for staff.





Reviews