We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

Portable digital spirometer Datospir Micro A (Sibelmed) – Bidirectional turbine, clinical and home mode, USB connection and W20s software | Ref. 02072

800,00 €

The portable digital spirometer Datospir Micro A (ref. 02072) from Sibelmed incorporates a bidirectional turbine transducer and complies with ERS/ATS (2005) and SEPAR recommendations for the diagnosis of asthma, COPD and other respiratory pathologies. It operates in clinical or home mode, with an internal database for more than 1,000 tests with graphs, a manoeuvre quality control programme, integrated temperature sensor with BTPS correction and USB connection (Bluetooth optional). Ideal for primary care, pulmonology and occupational medicine; compatible with SIBELMED W20s software.

International shipping. Import duties and taxes are payable by the recipient.

Datasheet (opens in a new tab)

Reference
02072

Description

1. Commercial Introduction

The Datospir Micro A (ref. 02072) is the portable digital spirometer from Sibelmed’s Micro family, developed in collaboration with the Pulmonary Function Laboratory of the Hospital de la Santa Creu i Sant Pau in Barcelona. A compact and reliable device, designed for the diagnosis of asthma, COPD and other respiratory diseases in primary care, pulmonology consultations, occupational medicine and home patient monitoring. With a factory-calibrated bidirectional turbine transducer, internal database for more than 1,000 tests and USB connectivity, it offers accurate, repeatable results ready to be integrated into the healthcare centre’s management.

2. Technical explanation of the equipment

The Datospir Micro A measures respiratory flow and volume using a bidirectional axial volumetric turbine with optoelectronic rotation sensor, which can be removed for cleaning and sterilisation. The integrated temperature sensor automatically corrects values to BTPS conditions (Body Temperature, Pressure, Saturated), ensuring measurement reliability in accordance with ERS/ATS (2005) and SEPAR recommendations. The device calculates flow-volume (F/V) and volume-time (V/T) curves and offers more than 40 spirometric parameters (FVC, FEV1, PEF, FEF25-75 %, MVV, bronchodilation and bronchoconstriction) with more than 20 international reference values for normal values. It has two operating modes: clinical mode, for using the device as a diagnostic spirometer in the consultation, and home mode, which allows the device to be programmed for monitoring asthmatic patients at home, displaying the measured PEF, FEV1 or FVC value on a colour scale. It incorporates a test quality control programme, accuracy verification with pre-recorded standard curves and export of tests to other management systems.

3. Technical Specifications

Features verified in the official Sibelmed technical documentation (Datospir Micro datasheet, ref. 02072):

Parameter Specification
Manufacturer / model Sibelmed (SIBEL S.A.U.) / Datospir Micro A
Reference / SKU 02072
Set type Portable digital spirometer, standalone use or with PC via software
Transducer Bidirectional axial volumetric turbine, optoelectronic rotation sensor, removable and sterilisable
Measurement range (BTPS) – Flow 0 to ±16 l/s
Measurement range (BTPS) – Volume 0 to 10 l
Dynamic flow resistance <0.6 cmH2O/l/s at 14 l/s
Accuracy (BTPS) – Volume ±3 % or 50 ml (whichever is greater)
Accuracy (BTPS) – Flow ±5 % or 200 ml/s (whichever is greater)
Accuracy – PEF ±10 % or 170 ml/s (whichever is greater)
Temporal accuracy 0.5 %
Volume resolution <6 ml
Sampling frequency 25 Hz
Display LCD (2×16)
Internal database More than 1,000 tests with graphs
Software SIBELMED W20s (optional on model A; compatible with Windows Vista, 7, 8.1 and 10)
Connectivity Standard USB; Bluetooth and RS232 optional; external printer via USB
Power supply 2 x 1.5 V batteries (alkaline recommended) or 1.2 V NiMH battery (>2200 mAh); power below 400 mW
Temperature sensor Integrated (BTPS correction)
Dimensions 153.5 × 80 × 52 mm
Weight (without battery or accessories) 250 g
Turbine service life 1,400 disinfection cycles or 3 years
Tests and parameters FVC/Bronchodilation, VC, MVV and bronchoconstriction (more than 40 parameters)
Normal value references GLI/Quanjer, SEPAR 2013, ERS, Knudson, Crapo, Zapletal, Morris and others (more than 20)
Regulatory standard ERS/ATS (2005) and SEPAR recommendations

Operating modes:

Mode Application
Clinical mode Use as a diagnostic spirometer in the clinic, with full configuration and quality control options
Home mode (patient) Device programming for monitoring asthmatic patients at home, with a colour traffic light (PEF, FEV1 or FVC)

4. Equivalences and Cross-References

Code 02072 corresponds to the official manufacturer reference from Sibelmed for the Datospir Micro A, confirmed in the manufacturer’s own technical data sheet. There is no manufacturer-validated equivalence table that cross-references this code with codes from other brands on a 1:1 interchangeable basis; the portable computerised spirometers in the same clinical category listed below share the same indication and general features, but differ in measurement principle, software and consumables, and are therefore classified as comparable by category:

Manufacturer / Brand Family / Model Reference code / SKU Equivalence level Technical differences / Remarks
MIR (Medical International Research) Spirobank II Smart 911028E0 / 911028E1 Comparable by category (not interchangeable 1:1) Bidirectional digital turbine; range ±16 l/s and 0-10 l; volume accuracy ±2.5 % or 50 ml; Bluetooth 4.0 and USB 2.0; memory for 10,000 tests; WinspiroPRO software. Codes verified in the official technical data sheet of MIR.
ndd Medical Technologies EasyOne Air 2500-2 / 2500-2INT Comparable by category (not interchangeable 1:1) TrueFlow ultrasonic measurement with lifetime calibration; colour touchscreen; Bluetooth and USB; charging base included and memory for 10,000 tests. Different measurement principle and consumables (FlowTube). Code verified on ndd’s official website and authorised distribution.

There is no manufacturer-validated equivalence table that cross-references reference 02072 with codes from other brands on a 1:1 interchangeable basis; consult the technical department before use in tenders. Other brands and equivalent families on the market in the same functional category (portable spirometer with software and PC connectivity) are the MIR Spirobank II, the ndd EasyOne Air and the portable spirometers from Vitalograph, COSMED and CONTEC; in all cases, verify the measurement principle, flow range, software and consumables before any replacement. Within the Datospir family itself, the compatible consumables and accessories verified in Sibelmed’s official catalogue are the bacterial/viral filter (ref. 09698, box of 100), the cardboard mouthpiece (ref. 01555, 28 ext. × 60 mm), the nasal clip (ref. 02692) and the Bluetooth module for Datospir Micro (ref. 02474).

5. Selection tips and variants of interest

  • Variants of the same family: if your centre needs a PDA-type touchscreen, diagnostic firmware and an expanded database as standard, the Datospir Micro C (ref. 02073) is the higher variant of the Micro family. For maximum hygiene without reprocessing, the Aira D series with disposable transducer (ref. 09064) is the alternative; the Aira T with turbine (ref. 09065) offers the lowest consumable cost in primary care and occupational medicine.
  • Recommended consumables: bacterial/viral filter ref. 09698 (box of 100) or ref. 08268 (28 mm) for each patient, cardboard mouthpieces ref. 01555 (28 ext. × 60 mm, box of 100) and nasal clip ref. 02692. For the paediatric population, attach the paediatric mouthpiece adapter (ref. 01149).
  • Quality control: the ATS/ERS recommendation advises periodically checking volume accuracy with the 3-litre calibration syringe (ref. 08290), together with the verification procedure integrated into the device.
  • Connectivity and data management: for wireless connection of the Micro A, the Bluetooth module (ref. 02474), the Bluetooth USB-PC adapter (ref. 01145), the charger with Ni-MH rechargeable batteries (ref. 01739) and the 5-pin mini USB cable (ref. 01682) are available. To integrate reports with the clinical history (HIS), request the W20sLink module (ref. 08279, formats PDF, XML, DICOM and CDA/HL7).

Additional information

Manufacturer

Sibelmed (SIBEL S.A.U.)

Model

Datospir Micro A

Reference

02072

Product type

Portable digital spirometer

Transducer

Bidirectional axial volumetric turbine with optoelectronic rotation sensor, removable and sterilisable

Volume accuracy

±3 % or 50 ml (whichever is greater)

Flow accuracy

±5 % or 200 ml/s (whichever is greater)

PEF accuracy

±10 % or 170 ml/s (whichever is greater)

Display

LCD (2×16)

Database

More than 1,000 tests with graphs

Connectivity

Standard USB; Bluetooth and RS232 optional

Software

SIBELMED W20s

Power supply

2 x 1.5 V batteries (alkaline recommended) or 1.2 V NiMH battery (>2200 mAh); power below 400 mW

Dimensions

153.5 × 80 × 52 mm

Weight

250 g (without battery or accessories)

Turbine service life

1,400 disinfection cycles or 3 years

Tests

FVC/Bronchodilation, VC, MVV and bronchoconstriction (more than 40 parameters)

Regulatory standard

ERS/ATS (2005) and SEPAR recommendations

Presentation

Unit with carrying case, cardboard mouthpiece, plastic mouthpiece, nose clip, USB cable and 2 AAA batteries

Flow range

0 to ±16 l/s (BTPS)

Volume range

0 to 10 l (BTPS)

Sampling frequency

25 Hz

Modes of use

Clinical mode and home mode (patient)

Tests and parameters

FVC/Bronchodilation, VC, MVV and bronchoconstriction (more than 40 parameters)

Reviews

There are no reviews yet.

Be the first to review “Portable digital spirometer Datospir Micro A (Sibelmed) – Bidirectional turbine, clinical and home mode, USB connection and W20s software | Ref. 02072”

Your email address will not be published. Required fields are marked *