Description
1. Commercial Introduction
The Datospir Micro A (ref. 02072) is the portable digital spirometer from Sibelmed’s Micro family, developed in collaboration with the Pulmonary Function Laboratory of the Hospital de la Santa Creu i Sant Pau in Barcelona. A compact and reliable device, designed for the diagnosis of asthma, COPD and other respiratory diseases in primary care, pulmonology consultations, occupational medicine and home patient monitoring. With a factory-calibrated bidirectional turbine transducer, internal database for more than 1,000 tests and USB connectivity, it offers accurate, repeatable results ready to be integrated into the healthcare centre’s management.
2. Technical explanation of the equipment
The Datospir Micro A measures respiratory flow and volume using a bidirectional axial volumetric turbine with optoelectronic rotation sensor, which can be removed for cleaning and sterilisation. The integrated temperature sensor automatically corrects values to BTPS conditions (Body Temperature, Pressure, Saturated), ensuring measurement reliability in accordance with ERS/ATS (2005) and SEPAR recommendations. The device calculates flow-volume (F/V) and volume-time (V/T) curves and offers more than 40 spirometric parameters (FVC, FEV1, PEF, FEF25-75 %, MVV, bronchodilation and bronchoconstriction) with more than 20 international reference values for normal values. It has two operating modes: clinical mode, for using the device as a diagnostic spirometer in the consultation, and home mode, which allows the device to be programmed for monitoring asthmatic patients at home, displaying the measured PEF, FEV1 or FVC value on a colour scale. It incorporates a test quality control programme, accuracy verification with pre-recorded standard curves and export of tests to other management systems.
3. Technical Specifications
Features verified in the official Sibelmed technical documentation (Datospir Micro datasheet, ref. 02072):
| Parameter | Specification |
|---|---|
| Manufacturer / model | Sibelmed (SIBEL S.A.U.) / Datospir Micro A |
| Reference / SKU | 02072 |
| Set type | Portable digital spirometer, standalone use or with PC via software |
| Transducer | Bidirectional axial volumetric turbine, optoelectronic rotation sensor, removable and sterilisable |
| Measurement range (BTPS) – Flow | 0 to ±16 l/s |
| Measurement range (BTPS) – Volume | 0 to 10 l |
| Dynamic flow resistance | <0.6 cmH2O/l/s at 14 l/s |
| Accuracy (BTPS) – Volume | ±3 % or 50 ml (whichever is greater) |
| Accuracy (BTPS) – Flow | ±5 % or 200 ml/s (whichever is greater) |
| Accuracy – PEF | ±10 % or 170 ml/s (whichever is greater) |
| Temporal accuracy | 0.5 % |
| Volume resolution | <6 ml |
| Sampling frequency | 25 Hz |
| Display | LCD (2×16) |
| Internal database | More than 1,000 tests with graphs |
| Software | SIBELMED W20s (optional on model A; compatible with Windows Vista, 7, 8.1 and 10) |
| Connectivity | Standard USB; Bluetooth and RS232 optional; external printer via USB |
| Power supply | 2 x 1.5 V batteries (alkaline recommended) or 1.2 V NiMH battery (>2200 mAh); power below 400 mW |
| Temperature sensor | Integrated (BTPS correction) |
| Dimensions | 153.5 × 80 × 52 mm |
| Weight (without battery or accessories) | 250 g |
| Turbine service life | 1,400 disinfection cycles or 3 years |
| Tests and parameters | FVC/Bronchodilation, VC, MVV and bronchoconstriction (more than 40 parameters) |
| Normal value references | GLI/Quanjer, SEPAR 2013, ERS, Knudson, Crapo, Zapletal, Morris and others (more than 20) |
| Regulatory standard | ERS/ATS (2005) and SEPAR recommendations |
Operating modes:
| Mode | Application |
|---|---|
| Clinical mode | Use as a diagnostic spirometer in the clinic, with full configuration and quality control options |
| Home mode (patient) | Device programming for monitoring asthmatic patients at home, with a colour traffic light (PEF, FEV1 or FVC) |
4. Equivalences and Cross-References
Code 02072 corresponds to the official manufacturer reference from Sibelmed for the Datospir Micro A, confirmed in the manufacturer’s own technical data sheet. There is no manufacturer-validated equivalence table that cross-references this code with codes from other brands on a 1:1 interchangeable basis; the portable computerised spirometers in the same clinical category listed below share the same indication and general features, but differ in measurement principle, software and consumables, and are therefore classified as comparable by category:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| MIR (Medical International Research) | Spirobank II Smart | 911028E0 / 911028E1 | Comparable by category (not interchangeable 1:1) | Bidirectional digital turbine; range ±16 l/s and 0-10 l; volume accuracy ±2.5 % or 50 ml; Bluetooth 4.0 and USB 2.0; memory for 10,000 tests; WinspiroPRO software. Codes verified in the official technical data sheet of MIR. |
| ndd Medical Technologies | EasyOne Air | 2500-2 / 2500-2INT | Comparable by category (not interchangeable 1:1) | TrueFlow ultrasonic measurement with lifetime calibration; colour touchscreen; Bluetooth and USB; charging base included and memory for 10,000 tests. Different measurement principle and consumables (FlowTube). Code verified on ndd’s official website and authorised distribution. |
There is no manufacturer-validated equivalence table that cross-references reference 02072 with codes from other brands on a 1:1 interchangeable basis; consult the technical department before use in tenders. Other brands and equivalent families on the market in the same functional category (portable spirometer with software and PC connectivity) are the MIR Spirobank II, the ndd EasyOne Air and the portable spirometers from Vitalograph, COSMED and CONTEC; in all cases, verify the measurement principle, flow range, software and consumables before any replacement. Within the Datospir family itself, the compatible consumables and accessories verified in Sibelmed’s official catalogue are the bacterial/viral filter (ref. 09698, box of 100), the cardboard mouthpiece (ref. 01555, 28 ext. × 60 mm), the nasal clip (ref. 02692) and the Bluetooth module for Datospir Micro (ref. 02474).
5. Selection tips and variants of interest
- Variants of the same family: if your centre needs a PDA-type touchscreen, diagnostic firmware and an expanded database as standard, the Datospir Micro C (ref. 02073) is the higher variant of the Micro family. For maximum hygiene without reprocessing, the Aira D series with disposable transducer (ref. 09064) is the alternative; the Aira T with turbine (ref. 09065) offers the lowest consumable cost in primary care and occupational medicine.
- Recommended consumables: bacterial/viral filter ref. 09698 (box of 100) or ref. 08268 (28 mm) for each patient, cardboard mouthpieces ref. 01555 (28 ext. × 60 mm, box of 100) and nasal clip ref. 02692. For the paediatric population, attach the paediatric mouthpiece adapter (ref. 01149).
- Quality control: the ATS/ERS recommendation advises periodically checking volume accuracy with the 3-litre calibration syringe (ref. 08290), together with the verification procedure integrated into the device.
- Connectivity and data management: for wireless connection of the Micro A, the Bluetooth module (ref. 02474), the Bluetooth USB-PC adapter (ref. 01145), the charger with Ni-MH rechargeable batteries (ref. 01739) and the 5-pin mini USB cable (ref. 01682) are available. To integrate reports with the clinical history (HIS), request the W20sLink module (ref. 08279, formats PDF, XML, DICOM and CDA/HL7).





Reviews