Description
1. Commercial Introduction
Ensure safe and effective cardiopulmonary resuscitation in the smallest patients with this pair of standard paediatric defibrillator electrodes with internal cable (reference Cahnos F7961P), a replacement consumable compatible with Nihon Kohden defibrillators from the Cardiolife TEC-2200, TEC-2300 and TEC-5500 series with JC-765V cable. It is the ideal solution for emergency departments, paediatric intensive care units (PICU), ambulances and any clinical setting that cares for children: a single pair of disposable pads keeps the equipment ready to act in cardiorespiratory arrest. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock and monitors the emergency electrocardiogram. In paediatric patients (under 8 years of age or weighing less than 25 kg) it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode incorporates a reduced active surface area of 40 cm² per electrode and must be used with the device’s paediatric mode switching or with the reduced energy configuration indicated in the defibrillator manual.
The integrated internal cable configuration provides a stable, direct connection to the device’s electrode socket, with no loose cables during resuscitation. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm medical foam (FOAM) guarantee a uniform distribution of current over the infant chest. The recommended placement is anterolateral (sternum-apex position), maintaining a minimum separation between the two pads. They are single-use: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical specifications and compatibility
| Reference Cahnos (SKU) | F7961P |
| Brand / compatible equipment | Nihon Kohden – Cardiolife TEC-2200 Series, TEC-2300 Series and TEC-5500 Series |
| Cable type | Internal cable JC-765V (direct connection to the electrode socket) |
| Product type | Standard multifunction paediatric electrode pair with internal cable |
| Presentation | Price per pair (2 units); 5 pairs per box (paediatric series) |
| Parameter | Specification |
| Target population | Paediatric (< 8 years / < 25 kg) |
| Cable configuration | Integrated internal cable (direct connection to the equipment electrode socket) |
| Total surface (per electrode) | 75 cm² |
| Active surface (per electrode) | 40 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 discharges (energy 50–100 J in paediatric configuration; verify according to the device indications) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector with internal cable |
| ECG monitoring | Up to 24 h |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C |
Note: the values for active surface area (40 cm²), total area (75 cm²), discharges (50–100 J) and adhesive ring (1 cm) correspond to the standard paediatric electrode series with internal cable in this product family; confirm them with the technical data sheet of the supplied batch and the device manual before use in tendering.
Compatibility with defibrillators
| Equipment brand | Compatible models | Required cable |
| Nihon Kohden | Cardiolife TEC-2200 Series | Internal cable JC-765V |
| Nihon Kohden | Cardiolife TEC-2300 Series | Internal cable JC-765V |
| Nihon Kohden | Cardiolife TEC-5500 Series | Internal cable JC-765V |
Note: always verify the correct fit of the connector in the device’s electrode socket before clinical use.
4. Equivalences and Cross-References
Reference F7961P is the Cahnos distribution code for this electrode compatible with Nihon Kohden; it is not an official reference of the equipment manufacturer. The connector and energy management of the Cardiolife TEC defibrillators are proprietary to Nihon Kohden, so equivalences with other brands are comparable only at category level and are not interchangeable 1:1. After verifying authorised distribution catalogues, the following cross-references with code confirmed in public sources have been identified. In no case is there an official equivalence table validated by the manufacturer for this reference.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Nihon Kohden | Disposable defibrillation electrodes adult/child/infant (original P-740K) | P-740K | Comparable by category (not interchangeable 1:1) | Factory original with 1.5 m cable, for adults and children; confirmed in authorised distributors of original spare parts. Covers AED-2100, AED-3100 and TEC-5600, TEC-7700 and TEC-8300 families; not verified for the TEC-2200/2300 series with internal JC-765V cable. Connector and labelling proprietary to Nihon Kohden; verify paediatric size and fit before replacement. |
| Nihon Kohden | Disposable AED electrodes adult/child/infant (original P-711) | P-711 (order code H329) | Comparable by category (not interchangeable 1:1) | Factory original with 1.5 m cable, for adult, child and infant; confirmed in authorised distributors. Aimed at the Cardiolife AED and TEC-8300K families; not verified for the TEC-2200/2300/5500 series with JC-765V cable. Verify compatibility before replacement. |
No verifiable original Nihon Kohden paediatric electrode specific to the TEC-2200/2300 series with internal JC-765V cable has been found in public sources; the manufacturer publishes electrodes with paediatric capability aimed at its AED and TEC-8300 families (P-711/H329, P-713/H330 for infant and P-740K). There is no equivalence table validated by the manufacturer that authorises the replacement of original electrodes with third-party pads: the connector and energy management of each defibrillator are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and instructions for use; consult the Cahnos technical department before use in tendering.
5. Selection tips and variants of interest
Confirm the equipment model: verify that your defibrillator is a Nihon Kohden Cardiolife TEC-2200, TEC-2300 or TEC-5500 and that the internal cable connector JC-765V fits correctly into the electrode socket of the equipment before clinical use; if in doubt, contact the technical department of Cahnos stating the model and serial number.
Do not confuse the paediatric and adult versions: this pair is for patients under 8 years of age or weighing less than 25 kg. For adult patients of the same connector family, Nihon Kohden offers the standard adult version with internal cable of the same series.
Energy in paediatric mode: always select the paediatric or reduced energy setting (50–100 J) indicated in the equipment manual; do not manually modify energy levels without protocol indication.
Variants of the same Cahnos family: version with external cable for Nihon Kohden (F7961PW, for TEC-5500, TEC-5600 and TEC-7600), standard multi-brand version with internal cable (F7955P) or with external cable (F7955PW), and radiotransparent version with internal cable (F7755P) if X-rays or CT scans are anticipated during care. In the internal cable family for other equipment, not interchangeable with each other due to proprietary connector: F7981P (GE Responder), F7986P (Mediana D500), F7985P (Instramed CardioMax) and F7951P (ZOLL).
Storage and expiry: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with damaged packaging must not be used because the gel may lose conductivity. As this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to the expiry date. For volume orders or tenders, contact the technical department of Cahnos stating the exact model and serial number of your equipment.