Description
1. Commercial Introduction
Keep your CU Medical Systems defibrillator always ready to act with this pair of standard adult defibrillator electrodes with internal cable (reference Cahnos F7958), a replacement consumable compatible with the Paramedic CU-ER series equipment (CU-ER1, CU-ER2, CU-ER3 and CU-ER5 with ER-5 OA03 cable) and the Lifegain CU-HD1. It is the indicated solution for emergency services, ambulances, civil protection, companies, sports centres and any heart-safe space equipped with these defibrillators. The integrated internal cable configuration keeps connector and cable housed inside the heat-sealed packaging: open the pads, connect them to the equipment socket following the diagram on the sachet and the defibrillator is ready for analysis and therapy. It is sold per pair (2 units) and is ready to be integrated into your cardiopulmonary resuscitation protocols.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the equipment analyses the cardiac rhythm, delivers the biphasic transthoracic shock and enables emergency electrocardiogram monitoring. This pair is indicated for adult patients (over 8 years of age or over 25 kg) in defibrillation, synchronised cardioversion and ECG monitoring. The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal energy transmission; the inner metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest, minimising the risk of burns. The recommended placement is anterolateral (sternum-apex), according to the diagrams on the packaging. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
Features of the standard adult electrode series with internal cable to which this reference belongs:
| Parameter | Specification |
| Reference Cahnos (SKU) | F7958 |
| Product type | Pair of adult multifunction defibrillator electrodes, standard, with internal cable |
| Brand / compatible equipment | CU Medical Systems – Paramedic CU-ER series (CU-ER1, CU-ER2, CU-ER3, CU-ER5 with ER-5 OA03 cable) and Lifegain CU-HD1; check connector fit in the equipment socket |
| Original factory reference | CUA0508O (Paramedic CU-ER series operator’s manual) |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Cable configuration | Integrated internal cable (direct connection to the equipment electrode socket) |
| Total surface (per electrode) | 148 cm² (standard adult series) |
| Active surface (per electrode) | 95 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks (50 to 360 J) |
| Functions | Defibrillation, synchronized cardioversion and ECG monitoring (max. 24 h) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector or clip (depending on model), internal cable |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of manufacture; expiry date indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium bag (10 pairs per box in adult series) |
Note: the active surface values (95 cm²), total area (148 cm²), shocks (50 to 360 J) and adhesive ring (1 cm) correspond to the standard adult electrode series of this product family; confirm with the technical data sheet of the supplied batch and the equipment manual before use in tenders.
4. Equivalences and Cross-References
This electrode (SKU F7958) is a consumable compatible with CU Medical Systems defibrillators of the Paramedic CU-ER series and the Lifegain CU-HD1, not an original spare part certified by the equipment manufacturer: the connector, the analysis algorithm and energy management are proprietary to each defibrillator. As a reference for comparison, the original factory adult electrode for the Paramedic CU-ER series is identified by the code CU Medical CUA0508O, confirmed in the operator’s manual and in authorised distribution. Equivalences with other brands are comparable only at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| CU Medical Systems (equipment manufacturer) | Paramedic CU-ER standard adult electrode (factory original) | CUA0508O | Comparable by category (not interchangeable 1:1) | Original reference documented in the Paramedic series operator’s manual; same adult range and same connector platform reproduced by this data sheet. Verify expiry date and connector fit before replacement. |
| ZOLL Medical | Stat-Padz® II – adult multifunction electrode | 8900-0801-01 | Comparable by category (not interchangeable 1:1) | Adult multifunction electrode with proprietary ZOLL connector (AED Plus / AED Pro); code confirmed on the manufacturer’s website and in GUDID/FDA. Not suitable for the socket of CU Medical equipment. |
| CONMED | PadPro® RTS – adult multifunction electrode | 2001Z | Comparable by category (not interchangeable 1:1) | Same clinical function (adult multifunction electrode) with its own connector, gel and geometry; code confirmed in authorised technical distribution. |
| Leonhard Lang | Skintact® Easibeat – adult multifunction electrode | DF28N | Comparable by category (not interchangeable 1:1) | Third-party adult multifunction electrode with its own connector (ZOLL compatible variant); code confirmed in the official Leonhard Lang catalogue. Always verify compatibility with the equipment socket. |
There is no equivalence table validated by the manufacturer authorising the substitution of the original electrodes of these devices with third-party pads: the connector, analysis algorithm and energy management are proprietary to each defibrillator. Any substitution must be validated against the device’s technical data sheet and IFU; contact Cahnos technical department before using in a tender. Within the Cahnos catalogue there are other references from the same family (adult standard) intended for devices from other manufacturers, not interchangeable with each other due to differing connector and algorithm: F7950W (Philips), F7952W (Physio-Control / Bexen / Cardioline / Mindray), F7957W (Cardiac Science / GE), F7959W (CU Medical i-PAD / CMOS Drake) and F7987W (ZOLL AED).
5. Selection tips and variants of interest
Confirm the device model: verify that your defibrillator is a CU Medical Systems Paramedic CU-ER series (CU-ER1, CU-ER2, CU-ER3, CU-ER5) or a Lifegain CU-HD1, and that the internal cable connector fits correctly into the electrode port before clinical use; in case of doubt, contact Cahnos technical department providing the model and serial number. It is not compatible with the AEDs of the i-PAD family (SP1, SP2, CU-SPR, NFK200 or NF1200/NF1201), which use their own consumables. Choose the variant according to your workflow: this model with internal cable provides a clean and robust connection during resuscitation; if you prefer to leave the device pre-connected and operational in seconds, the reference F7958W (same adult standard series with external cable) is available. Do not confuse adult and paediatric versions: this pair is for adult patients (over 8 years old or over 25 kg); for paediatric patients with the same CU-ER series devices use references F7958P (internal cable) or F7958PW (external cable), and for the CU-ER1 the reduced energy version F7958PL. Check the expiry: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with damaged pouches must not be used because the gel may lose conductivity. Plan your restocking: as this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to the expiry date (FIFO). For volume orders or tenders, contact Cahnos technical department providing the exact model and serial number of your defibrillator.







Reviews