Description
1. Commercial Introduction
Ensure safe and effective cardiopulmonary resuscitation in the smallest patients with this pair of standard paediatric defibrillator electrodes with internal cable (reference Cahnos F7953P). It is the ideal replacement and replenishment consumable for emergency services, paediatric intensive care units (PICU), ambulances and cardioprotected spaces that care for children with equipment from the Philips, Agilent Hewlett-Packard and Laerdal Medical families: a single reference covers the most widespread hospital fleets of defibrillator-monitors (CodeMaster 100, XE, XL, XL+ and HeartStart 4000), simplifying the consumables stock. Price per pair, individually heat-sealed packaging, immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the device analyses the rhythm, delivers the shock and monitors the electrocardiographic response. This paediatric model allows transthoracic external defibrillation, synchronised cardioversion and ECG monitoring (up to 24 hours), adapted to paediatric physiology.
In patients under 8 years of age or weighing less than 25 kg, it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode incorporates a reduced active surface of 40 cm² per electrode and must be used with the device’s paediatric mode switching or with the reduced energy setting (50–100 J) indicated in the defibrillator manual. The variant with integrated internal cable (leads-in, 100 cm) offers a direct and clean connection to the device connector, avoiding loose elements during care. The low-impedance conductive adhesive gel on 1 mm medical foam with metal foil ensures stable adhesion, uniform current transmission and a reliable signal. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging, being placed according to the anatomical diagrams on the sachet.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | F7953P |
| Product type | Paediatric multifunction defibrillator electrode pair, standard, with internal cable |
| Indicated population | Paediatric: <8 years / <25 kg |
| Cable configuration | Integrated internal cable (leads-in), 100 cm |
| Radiotransparency | No (standard version) |
| Total surface per electrode | 75 cm² |
| Active surface per electrode | 40 cm² |
| Adhesive ring width | 0.7 cm |
| Supported shocks | Up to 50 shocks at 100 J |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | 1 mm thick medical foam |
| Connection type | Safety anti-shock connector compatible with Philips / Agilent HP / Laerdal equipment |
| ECG monitoring | Up to 24 hours |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium pouch, with laser pre-cut and easy opening |
| Sales format | 5 pairs per box (paediatric series) |
| Shelf life | 30 months from the date of production (see expiry date on the packaging) |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Composition | Latex-free and phthalate-free · Single-use |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993 |
3.1 Compatibility with defibrillators
| Device model | Factory reference | Observations |
| CodeMaster 100 | M2475B | Direct connection |
| CodeMaster XE | M1724A | Direct connection |
| CodeMaster XL | M1723A / M1723B | Direct connection |
| CodeMaster XL+ | M1722A / M1722B | Direct connection |
| HeartStart 4000 | M5500B | Paediatric function and energy in manual setting |
| HeartStream / HeartStart XL | M4735A | Paediatric function and energy in manual setting |
| HeartStream / HeartStart XLT | M3500B | Paediatric function and energy in manual setting |
| HeartStart MRx | M3535A / M3536A | With hands-free electrode cable M3508A |
| Laerdal AED Trainer 2 | — | Direct connection (training) |
Note: with HeartStart 4000, XL/XLT and MRx do not use in fully automatic defibrillators; the paediatric function and energy are adjusted manually. Always verify the correct insertion of the connector into the device socket before clinical use.
4. Equivalences and Cross-References
Reference F7953P corresponds to a standard paediatric multifunction electrode with internal cable for the Philips CodeMaster (HP/Agilent) and Laerdal Medical defibrillator families indicated above. After verifying manufacturer sources and regulatory databases (GUDID and official catalogues), the following cross-references with confirmed code are identified. In no case is there an official equivalence table validated by the manufacturer for this reference; any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Philips Healthcare | Paediatric multifunction radiotransparent electrodes (plug-in connection) | M3719A / 989803125411 | Comparable by category (not interchangeable 1:1) | Original manufacturer reference for the same compatibility family (M1722A/B, M1723A/B, M1724A, M2475B, M3500B, M4735A, M5500B, M3535A/M3536A). Code confirmed in GUDID and official Philips website. Its datasheet directs it to infants <10 kg, with short cable and conductive surface of 50 cm²; verify size, surface and connector before replacement. |
| Cardinal Health (Covidien) | Medi-Trace Cadence paediatric RTS (radiotransparent) | 22661P | Comparable by category (not interchangeable 1:1) | Paediatric radiotransparent pair (barrel connector compatible with Philips HP) for the CodeMaster XL+ (M1722A/B, M1723A/B), HeartStream XLT (M3500B), HeartStart XL (M4735A), MRx and HeartStart 4000 (M5500B) equipment family. Code confirmed in GUDID (DI 20884527022356) and in authorised distribution catalogues; verify connector, cable and surfaces before use. |
| Nissha Medical Technologies (HeartSync) | Paediatric defibrillation pad for Philips | 6600066H (PEDIATRIC-PHILIPS, D321) | Comparable by category (not interchangeable 1:1) | Paediatric radiolucent pair for the HeartStart XL/T, XL+, MRx and Intrepid family; total surface of 78.4 cm² with tin and tin-copper conductive element; packaged per pair in heat-sealed bag. Code confirmed in the manufacturer’s official catalogue; not suitable for use with AED: verify compatibility with the exact model of the device. |
There is no equivalence table validated by the manufacturer for this reference; consult with the Cahnos technical department before using in tendering.
5. Selection tips and variants of interest
- Select according to the patient: this paediatric reference is for children under 8 years of age or weighing less than 25 kg. For patients over 25 kg use the adult version of the same connector family (F7953).
- Choose the cable configuration according to your workflow: the internal cable of this reference offers clean placement with direct connection to the device. If you prefer to leave the device pre-connected and ready in seconds, you have the standard version with external cable (F7953PW).
- Radiotransparency: if X-rays or CT scans are anticipated during care, consider the radiotransparent series with internal cable (F7753P) or with external cable (F7753PW), which avoid image artefacts without removing the electrode.
- Verify the device configuration: on the HeartStart 4000, XL/XLT and MRx series models, the paediatric function and energy are adjusted manually; always check connector compatibility (and M3508A cable where applicable) before storing the spare.
- Storage: keep in a dry place between +5 °C and +35 °C and respect the expiry date indicated on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date.
- Single-use: do not reuse; one pair per patient and clinical event. Immediate availability from Cahnos with technical assistance for compatibility verification by model and connector.