Description
1. Commercial Introduction
Keep your defibrillator always ready to act with this pair of standard adult defibrillator electrodes with internal cable (reference Cahnos F7952), a replacement consumable compatible with equipment from the Physio-Control LIFEPAK, Bexen Cardio REANIBEX, Cardioline ELIFE 700 and Mindray BeneHeart families. It is the indicated solution for emergency services, ambulances, hospitals, health centres, companies and any heart-safe space equipped with these devices.
Its configuration with integrated internal cable keeps the connector protected in the packaging until the moment of use: simply open the sachet, place the pads according to the anatomical diagrams and connect to the device, with no loose cables during resuscitation. It is sold per pair (2 units), with immediate dispatch and specialised technical support from Cahnos to verify compatibility by model and connector before placing the order.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronized cardioversion, non-invasive transcutaneous pacing and emergency electrocardiogram monitoring. This pair is indicated for adult patients (older than 8 years or more than 25 kg).
The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal energy transmission; the internal metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest, minimising the risk of burns. The recommended placement is anterolateral (sternum-apex), according to the diagrams on the packaging. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | F7952 |
| Product type | Pair of standard multifunction adult defibrillator electrodes with internal cable (leads-in) |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Cable configuration | Integrated internal cable, length 120 cm |
| Total surface (per electrode) | 148 cm² |
| Active surface (per electrode) | 95 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | From 50 to 360 J (up to 50 shocks) |
| Functions | Defibrillation, synchronized cardioversion, non-invasive transcutaneous pacing and ECG monitoring |
| ECG monitoring | Max. 24 h |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (1 mm thick) |
| Connection type | Anti-shock safety connector or clip (depending on the equipment model) |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Class IIb medical device (Directive 93/42/EEC and amendments) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of manufacture; expiry date indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium bag; 10 pairs per box (adult series) |
3.1 Compatibility with defibrillators
| Equipment manufacturer | Compatible models |
| Physio-Control | LIFEPAK 9 (with QUIK-COMBO adapter), LIFEPAK 10, LIFEPAK 12, LIFEPAK 15, LIFEPAK 20, LIFEPAK 20E, LIFEPAK 500 AED (with grey socket and pink connector), LIFEPAK 1000 |
| Bexen Cardio | REANIBEX-200, REANIBEX-300, REANIBEX-500, REANIBEX-700, REANIBEX-800 |
| Cardioline | ELIFE 700 |
| Mindray | BENEHEART D1 (Public / Pro), BENEHEART D3, BENEHEART D6 |
Note: compatibility refers to the connector platform of each family. Always verify the correct fit of the connector into the socket of your device before clinical use.
4. Equivalences and Cross-References
This electrode (SKU F7952) is a consumable compatible with the cited defibrillators, not an original spare part certified by each equipment manufacturer: the connector, analysis algorithm and energy management are proprietary to each defibrillator. After verifying official manufacturer catalogues, authorised technical distribution and regulatory databases (GUDID), the following cross-references with code confirmed in public sources have been identified. Any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Stryker / Physio-Control | EDGE System Electrodes with QUIK-COMBO Connector (original for the device, internal cable / not pre-connected) | 11996-000091 | Comparable by category (not interchangeable 1:1) | Factory original for LIFEPAK 12, 15 and 20; cable inside the packaging (not pre-connected), same internal cable configuration and same clinical function (defibrillation, pacing, cardioversion and ECG). Code verified in GUDID/FDA (UDI 00883873784945). Verify connector, gel and labelling before replacement. |
| Leonhard Lang | Skintact® Easibeat – adult multifunction electrode, Physio-Control connection platform | DF20N | Comparable by category (not interchangeable 1:1) | Multifunction (defibrillation, pacing, cardioversion and monitoring) for LIFEPAK 9, 10, 12, 15, 20, 500 and 1000; 1 pair per pouch and 10 pouches per box. Code confirmed in official Leonhard Lang documentation and authorised technical distribution. Own connector, gel and materials; verify fit before replacement. |
| Bexen Cardio | Original adult electrodes for Reanibex 200 / 300 / 500 / 700 / 800 (not pre-connected) | TKL0311A | Comparable by category (not interchangeable 1:1) | Factory original replacement for the REANIBEX series, from the same QUIK-COMBO type connection platform; non-pre-connected presentation (internal cable). Code confirmed in authorised distribution. Verify connector fit before replacement. |
| Mindray | Adult Multifunction Electrode Pads (Basic) for BeneHeart / MR60 series | 115-040517-00 | Comparable by category (not interchangeable 1:1) | Mindray original for BeneHeart D1, D3, D6 defibrillators and MR60 series; adult >25 kg, up to 50 shocks at 360 J. Code confirmed in Mindray’s official pad guide and authorised distribution. Mindray proprietary connector and algorithm; verify compatibility with the device socket. |
| Cardioline | Single-use defibrillation electrode (2 pcs) for ELIFE 700 | 8730420 | Comparable by category (not interchangeable 1:1) | Factory original replacement for the Cardioline ELIFE 700 defibrillator; supplied as a pair. Code confirmed in authorised distribution. Verify connector and device energy management before replacement. |
There is no equivalence table validated by the manufacturer that authorises the replacement of the original electrodes of these devices with third-party pads: the connector, analysis algorithm and energy management are proprietary to each defibrillator. Any replacement must be validated against the device’s technical data sheet and IFU; consult the Cahnos technical department before use in tenders.
5. Selection tips and variants of interest
Confirm the device model: this pair corresponds to Physio-Control LIFEPAK 9 (with QUIK-COMBO adapter), 10, 12, 15, 20, 20E, 500 AED (grey socket and pink connector) and 1000 defibrillators, as well as Bexen Cardio REANIBEX, Cardioline ELIFE 700 and Mindray BeneHeart D1/D3/D6. Always verify the connector fit and the expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention: the gel may lose conductivity and compromise the effectiveness of the shock.
Choose the variant according to your workflow: the internal cable of this model (F7952) provides a clean and robust connection during resuscitation; if you prefer to leave the device pre-connected and ready in seconds, you have the version with external lead-out cable from the same platform (F7952W, 120 cm) or the long cable variant F7952/200 (200 cm). For image-guided procedures (radiography, fluoroscopy or CT), consider the radiotransparent versions F7752 (internal cable) or F7752W (external cable). For paediatric patients (under 8 years or 25 kg), use the paediatric references F7952P and F7952PW (or F7752P / F7752PW in radiotransparent version).
Storage and stock: store in a dry place between +5 °C and +35 °C and respect the expiry date on the packaging (approximate shelf life of 30 months). As this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO). For volume orders or tenders, contact the Cahnos technical-commercial department indicating the exact model and serial number of your devices; we will confirm compatibility before formalising the order.