Description
1. Commercial Introduction
Ensure safe and effective cardiac care for the smallest patients with this pair of radiotransparent paediatric defibrillator electrodes with external cable (reference F7751PW), a high-reliability consumable compatible with ZOLL defibrillators and monitors from the PD, E, M and R series. Designed for emergency services, paediatric ICU, medical transport, operating theatres and haemodynamics, it combines the advantages of the radiotransparent series (artefact-free imaging) with the operability of the pre-connected external cable, which keeps the equipment always ready to act. A cost-optimised solution, compliant with European medical device standards and latex- and phthalate-free.
2. Technical explanation of the equipment and procedure
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the equipment analyses the rhythm, delivers therapy and monitors the response. This paediatric model allows transthoracic external defibrillation, synchronised cardioversion, non-invasive transcutaneous cardiac pacing and ECG monitoring, adapted to paediatric physiology (children under 8 years or under 25 kg). Its radiotransparent condition —electrode and cable— avoids artefacts in imaging studies, so it can remain in place during X-ray-guided procedures without interfering with diagnosis. The external cable variant allows the connector to remain accessible and the pad ready, reducing preparation time in emergencies. The low-impedance conductive adhesive gel on medical foam with a metallic film ensures uniform current distribution and a reliable signal. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging, and placed according to the anatomical diagrams (sternal and apical position).
3. Technical Specifications
Radiotransparent paediatric multifunction electrode from the EuroDefiPads paediatric series, in its external cable (pre-connected) variant.
| Field | Specification |
|---|---|
| Reference Cahnos (SKU) | F7751PW |
| Product type | Pair of paediatric multifunction defibrillator electrodes, radiotransparent, with external cable (pre-connected) |
| Indicated population | Children under 8 years or weighing less than 25 kg |
| Radiotransparency | Yes: radiotransparent electrode and cable, for artefact-free imaging |
| Cable | External (pre-connected), to leave the equipment ready in emergencies |
| Functions | Defibrillation, synchronised cardioversion, non-invasive external pacing and ECG monitoring (max. 24 h) |
| Supported shocks | 50–100 J (paediatric series) |
| Total / active area (per electrode) | 75 cm² total / 40 cm² active |
| Conductive gel | Low-impedance conductive adhesive gel |
| Backing material | Medical foam (1 mm) with conductive metal foil |
| Connector | Specific connector for ZOLL equipment (verify correct fit in the device socket) |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 5 pairs per box (paediatric series) |
| Use | Single-use, disposable; latex-free and phthalate-free |
| Shelf life | Approx. 30 months from date of manufacture (check printed expiry date) |
| Conditions of use and storage | Operating temperature from 0 °C to +50 °C; store in a dry place between +5 °C and +35 °C |
| Classification | Medical device class IIb (Directive 93/42/EEC and amendments) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993-10 |
Compatibility (equipment models)
| Equipment manufacturer | Compatible models |
|---|---|
| ZOLL Medical | PD 1200, PD 1400, PD 1600, PD 1700, PD 2000 |
| ZOLL Medical | E Series, M Series, R Series |
Technical note: according to the manufacturer’s documentation for the original ZOLL electrodes, with M Series, E Series and R Series equipment the manual mode must be used: energy levels are not adjusted automatically and the device does not recognise them as intelligent paediatric electrodes. Always verify correct connector insertion before clinical use.
4. Equivalences and Cross-References
This electrode (SKU F7751PW) is a consumable compatible with the aforementioned ZOLL equipment, not an original factory spare part. The correspondence with the equipment manufacturers’ references is as follows; any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| ZOLL Medical | Pedi·Padz II — multifunction paediatric electrodes (original equipment) | 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Factory original for ZOLL equipment (AED Plus and AED Pro with intelligent paediatric capability; M/E/R Series in manual mode). Hydrogel with tin conductive element, pre-connected, paediatric range 0–8 years / under 25 kg with energy levels of 50/70/85 J. Code confirmed in the official ZOLL store, authorised distributors and GUDID (UDI 00847946001438). |
| Stryker / Physio-Control | Infant/Child Reduced Energy Defibrillation Electrodes (for LIFEPAK with pink connector) | 11101-000016 | Comparable by category (not interchangeable 1:1) | Paediatric electrode with energy reduction (200 J → ≈50 J) for the LIFEPAK 500, 1000, EXPRESS and CR Plus AEDs; proprietary QUIK-COMBO connector and algorithm, not suitable for ZOLL equipment. Code confirmed in the official Stryker store and in GUDID (UDI 00721902629013). |
There is no equivalence table validated by ZOLL Medical that authorises the replacement of the original Pedi·Padz II electrodes with third-party pads: the connector, analysis algorithm and energy management are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and IFU. Consult the Cahnos technical department before use in tenders.
5. Selection tips and variants of interest
Confirm the model and connector of the equipment: this pair corresponds to the ZOLL defibrillators and monitors PD 1200, PD 1400, PD 1600, PD 1700, PD 2000, E Series, M Series and R Series. Verify the reference and expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention.
Choose the variant according to the workflow: the external cable of this model allows the equipment to be pre-connected and ready in seconds; if you prefer an internal cable, the reference F7751P is available (same paediatric radiotransparent series). For standard use without a radiotransparency requirement, consider F7951PW (standard pre-connected paediatric) or F7951P (standard paediatric with internal cable). The adult version of this same radiotransparent series is F7751W; for equipment from other manufacturers, F7752PW is available (Physio-Control/Medtronic LIFEPAK and Mindray BeneHeart).
Maintenance and stocking: manage stock rotation by expiry date (approx. 30 months) and always keep a spare pair next to the equipment in centres with a paediatric population. For volume orders or tenders, contact the Cahnos technical department indicating the exact model and serial number of the defibrillator.






Reviews