1. Commercial Introduction
Ensure safe care for the youngest patients with this pair of paediatric AED defibrillator electrodes with limited energy, reference F7952WPL, compatible with the devices Physio-Control LIFEPAK 500 AED Biphasic (pink socket), LIFEPAK 1000, LIFEPAK EXPRESS and LIFEPAK CR Plus. It is the clinical spare part indicated for schools, nurseries, sports centres, paediatric clinics and any public access defibrillation point with children present: thanks to its external cable (lead-out) the connector remains accessible without opening the sealed pouch, so that the device remains operational in a matter of seconds. A reliable, cost-optimised solution that complies with European medical device standards, designed to keep the AED always ready for paediatric care.
2. Technical explanation of the equipment and procedure
The automatic external defibrillators (AEDs) of the LIFEPAK family analyse the heart rhythm through disposable multifunction electrodes. When this pair of paediatric electrodes is connected, the limited-energy system built into the pad itself reduces the shock that the device would deliver in adult configuration: 200 J from the defibrillator output become approximately 50 J delivered to the patient, a dose adapted to paediatric physiology according to international cardiopulmonary resuscitation guidelines. The electrodes are placed following the anatomical diagrams on the package (sternal and apical position) and, in addition to defibrillation, they allow ECG monitoring during resuscitation (up to 24 h). They incorporate low-impedance conductive adhesive gel on a medical foam backing with a conductive metal foil, ensuring a reliable signal and uniform current distribution. They are single-use: they must be replaced after each use or upon reaching the expiry date printed on the package.
3. Technical Specifications
Paediatric multifunction limited-energy electrode from the EuroDefiPads paediatric series (P series, <8 years / <25 kg), in its external cable variant (lead-out) of 120 cm.
| Field | Specification |
| Reference Cahnos (SKU) | F7952WPL |
| Compatible brand | Physio-Control (LIFEPAK equipment) |
| Compatible models | LIFEPAK 500 AED BIPHASIC with pink socket (pink socket), LIFEPAK 1000 AED, LIFEPAK EXPRESS AED, LIFEPAK CR Plus AED |
| Product type | Paediatric multifunction defibrillator electrode, QUIK-COMBO connector with pink socket |
| Population | Children under 8 years or weighing less than 25 kg (55 lb) |
| Energy | Limited energy for paediatric use: reduction from 200 J (device output) to ≈50 J delivered to the patient; supported shock range 50–100 J |
| Cable | 120 cm, external (lead-out / preconnected) |
| Functions | Defibrillation, rhythm analysis and ECG monitoring (max. 24 h) |
| Total / active area (per electrode) | 75 cm² total / 40 cm² active (paediatric series) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Backing material | Medical foam (1 mm) with conductive metal foil |
| Use | Single-use, disposable; non-sterile; latex-free and phthalate-free |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch with easy-open notch; 5 pairs per box (paediatric series) |
| Shelf life | Approx. 30 months from date of manufacture (check printed expiry date) |
| Conditions of use and storage | Operating 0 °C to +50 °C; store in a dry place between +5 °C and +35 °C; packaged withstand from −30 °C to +65 °C for a maximum of 7 days |
| Classification | Medical device class IIb (Directive 93/42/EEC and amendments) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993-10 |
| Compatibility note | The limited energy function requires a biphasic device with pink socket (LIFEPAK 500 after 2002); do not use on monophasic LIFEPAK 500 with grey socket without validation with the technical service |
4. Equivalences and Cross-References
This electrode (SKU F7952WPL) is a consumable compatible with Physio-Control/Stryker LIFEPAK defibrillators with pink socket, not an original factory spare part. The correspondence with the device manufacturer references and with other brands in the sector is as follows; any comparison between brands should be understood as comparable at category level, not 1:1 interchangeable.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Stryker / Physio-Control | Infant/Child Reduced Energy Defibrillation Electrodes (original equipment) | 11101-000016 (replacement electrode) / 11101-000017 (kit with carrying bag) | Comparable by category (not interchangeable 1:1) | Factory original for LIFEPAK 500 (pink socket), 1000, EXPRESS and CR Plus; same paediatric criterion (<8 years / <25 kg) and same attenuation (÷4); labelling, gel and packaging proprietary to Stryker. Codes confirmed on Stryker official store and in GUDID (UDI 00721902629013) |
| ZOLL Medical | Pedi·padz II (pre-connected paediatric electrodes) | 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Paediatric range 0–8 years / <25 kg with levels of 50/70/85 J selected by the AED; connector and algorithm proprietary to ZOLL (AED Plus and AED Pro); not suitable for LIFEPAK devices. Code verified in GUDID (UDI 00847946001438) |
| Philips | Infant/Child Reduced-Energy Defibrillator Pads (for HeartStart FR2/FR2+) | M3870A (also 989803108851) | Comparable by category (not interchangeable 1:1) | Reduces energy from 150 J to 50 J; connector and algorithm proprietary Philips for FR2 (M3840A/M3841A) and FR2+ (M3860A/M3861A); not compatible with LIFEPAK devices. Code verified in GUDID (UDI-DI 00884838023734) and on the official website of Philips |
There is no equivalence table validated by Physio-Control/Stryker that authorises the substitution of the original references 11101-000016 / 11101-000017 with third-party pads: the connector, the analysis algorithm and the energy management are proprietary. Any substitution must be validated against the defibrillator’s technical data sheet and IFU. Within the Cahnos catalogue there are comparable families for other equipment in the same category (paediatric AED electrodes with limited energy), such as F7950WPL (Philips), F7957WPL/F7957PL (Cardiac Science), F7959WPL (CU Medical i-PAD) or F7966WPL (Defibtech), all of them comparable by category and not interchangeable 1:1 with each other. Consult the Cahnos technical department before use in tenders.
5. Selection tips and variants of interest
- Confirm the model and the equipment connector: this pair of electrodes corresponds to the LIFEPAK 500 AED Biphasic with pink connector, LIFEPAK 1000, LIFEPAK EXPRESS and LIFEPAK CR Plus. It is not valid for the monophasic LIFEPAK 500 with grey connector or for other models without prior connector validation.
- Check the reference and expiry date: check the F7952WPL code and expiry date on the packaging; do not force the installation of an expired electrode (the gel may lose conductivity) nor reuse a pair after a procedure.
- Paediatric provision: in centres with a paediatric population, always keep a pair of these electrodes with the AED; the energy is reduced automatically when the limited-energy pad is connected, with no manual configuration.
- Variants within the Cahnos catalogue: F7952PL (same reduction 200 J→50 J with 120 cm internal cable), F7952P (standard paediatric 100 cm, without reduction, not suitable for pacing) and F7952PW (standard paediatric lead-out, without reduction). For equipment from other brands, consider F7950WPL (Philips), F7957WPL (Cardiac Science), F7959WPL (CU Medical) or F7966WPL (Defibtech), all with limited energy.
- Maintenance: manage stock rotation by expiry date (approx. 30 months) and always replace the electrode after use or expiry. For volume orders or tenders, contact the Cahnos technical department stating the exact defibrillator model.