Paediatric AED limited-energy defibrillator electrode 150 J → 50 J (Ref. F7950PL) – Compatible replacement for Philips HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A), functional substitute for Philips M3870A

62,50 €

Paediatric AED limited-energy defibrillator electrode (Ref. F7950PL), compatible replacement electrode for Philips HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A) defibrillators. Pair of disposable multifunction pads for patients under 8 years or under 25 kg, with electronics that attenuate the discharge from 150 J to 50 J (SMART Biphasic therapy) and functionally replace the original factory Philips M3870A electrode. Low-impedance conductive adhesive gel, shock-proof connector, latex-free and phthalate-free. Sold per pair; check the package expiry date before use.

International shipping. Import duties and taxes are payable by the recipient.

Reference
F7950PL

Description

1. Commercial Introduction

Ensure your Philips defibrillator is paediatric-ready with this pair of limited-energy defibrillation electrodes, designed as a compatible replacement for the HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A) models. It is the ideal solution for schools, nurseries, sports centres, paediatric clinics, ambulances and any public-access defibrillation point where children are present: when this paediatric electrode is connected, the device delivers a safe shock adapted to the physiological characteristics of the smallest patients.

With reference Cahnos F7950PL, this consumable replicates the function of the original factory paediatric electrode Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads), maintaining the same connection type and the same limited-energy concept (150 J → 50 J). Single-use product, available for dispatch and with specialised technical support from Cahnos.

2. Technical explanation of the equipment and procedure

The automated external defibrillator (AED) analyses the heart rhythm through the multifunction electrodes connected to the patient and, if it detects a shockable rhythm (ventricular fibrillation or ventricular tachycardia), delivers a biphasic shock. In paediatric patients — under 8 years of age or weighing less than 25 kg — applying the energy intended for adults can be harmful to the child’s myocardium. Therefore, this electrode is limited-energy: the original factory Philips M3870A electrode contains electronics that attenuate the device’s shock from 150 J to 50 J, a dose suitable for paediatric resuscitation in accordance with international CPR guidelines (SMART Biphasic therapy).

The pads are placed following the anatomical diagrams on the packaging and allow, in addition to defibrillation, heart rhythm monitoring during care (max. 24 h). They incorporate low-impedance conductive adhesive gel that ensures optimal skin contact and efficient current transmission. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging.

3. Technical Specifications

Parameter Specification
Reference Cahnos (SKU) F7950PL
Product type Paediatric multifunction limited-energy defibrillation electrode, 1 pair (2 pieces), disposable
Compatible brand Philips (HeartStart / HeartStream / HeartStar)
Compatible models HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A)
Indicated population Children under 8 years or weighing less than 25 kg (55 lb)
Energy Limited for paediatric use; the original factory Philips electrode attenuates the shock from 150 J to 50 J (SMART Biphasic therapy)
Functions External defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring (max. 24 h)
Total surface (per electrode) 75 cm² (paediatric series)
Active surface (per electrode) 40 cm² (paediatric series)
Adhesive ring width 0.7 cm
Supported shocks 50–100 J (paediatric series)
Conductive gel Low-impedance conductive adhesive gel
Conductive material Metal foil
Backing material Medical foam (1 mm thick)
Connection type Safety anti-shock connector (Philips)
Composition Latex-free and phthalate-free
Use Single patient, non-reusable
Sterility Non-sterile
Presentation 1 pair in heat-sealed polyethylene and aluminium bag
Commercial format 5 pairs per box (paediatric series)
Regulatory standard Medical device class IIb (Directive 93/42/EEC); ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993
Shelf life 24–30 months from production date depending on batch (see labelling)
Expiry As indicated on the packaging; do not use after the expiry date
Operating temperature 0 °C to +50 °C
Storage Dry place, between +5 °C and +35 °C; do not use if the packaging is open or damaged

Technical note: the compatibility data, target population and limited energy correspond to the manufacturer’s technical documentation for the paediatric series; before placing the order, confirm the exact model of your device.

4. Equivalences and Cross-References

The original factory reference that this compatible replacement substitutes is the paediatric electrode Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads, for HeartStart FR2/FR2+), verified in the GUDID/FDA regulatory database (UDI-DI 00884838023734; FDA class III; GMDN: paediatric external defibrillation electrode, single-use; non-sterile). Philips does not publish an official equivalence table authorising pad interchange between brands: the connector, energy attenuator and analysis algorithm are proprietary. The references listed below share the same type (paediatric limited-energy electrode for AED) and are classified as comparable by category, not interchangeable 1:1.

Manufacturer / Brand Family / Model Reference code / SKU Equivalence level Technical differences / Remarks
Philips Infant/Child Reduced-Energy Defibrillation Pads (HeartStart FR2 / FR2+) M3870A (also 989803108851) Comparable by category (not interchangeable 1:1) Original factory reference that this compatible replacement substitutes for the same FR2/FR2+ devices; attenuates energy from 150 J to 50 J (SMART Biphasic therapy). Verified in GUDID/FDA (UDI-DI 00884838023734).
Philips HeartStart OnSite / Home / HS1 Infant/Child SMART Pads Cartridge M5072A Comparable by category (not interchangeable 1:1) Paediatric cartridge with 150→50 J attenuator, cartridge format and proprietary connector for the OnSite/HS1 series; does not fit FR2/FR2+. Verified in GUDID/FDA (UDI-DI 00884838127784).
ZOLL Medical Pedi-Padz II (AED Plus / AED Pro) 8900-0810-01 Comparable by category (not interchangeable 1:1) Same paediatric range (0–8 years / <25 kg) with energy automatically adjusted to paediatric levels (50/70/85 J); connector, algorithm and management electronics exclusive to ZOLL. Verified in GUDID/FDA (UDI-DI 00847946001438).
Cardiac Science (ZOLL) Powerheart Pediatric Pads (series 9200/9300) 9730-002 Comparable by category (not interchangeable 1:1) Integrated attenuator to ~50 J for Powerheart G3 and GE Responder AED; proprietary connector. Verified in GUDID/FDA (UDI-DI 00812394020935).
Defibtech Lifeline DDU-1XX Pediatric Rescue Pads DDP-200P Comparable by category (not interchangeable 1:1) Paediatric reduced-energy electrodes with integrated attenuator for children under 8 years or <25 kg; proprietary Defibtech connector. Verified in GUDID/FDA (UDI-DI 00815098020430).

There is no equivalence table validated by Philips that cross-references the M3870A with third-party electrodes for HeartStart FR2/FR2+: the connector, energy attenuator and analysis algorithm are proprietary. Any comparison with other brands should be understood as comparable at category level, not interchangeable 1:1, verifying one by one the device connector, energy range, target population and packaging expiry date before any substitution. Consult with the technical department of Cahnos before using in a tender.

5. Selection tips and variants of interest

Confirm the device model: this electrode corresponds to Philips AEDs HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A). Check the model on the device label before ordering; not valid for devices with a different connector (OnSite/HS1, FR3 in standby, etc.).

Check the reference and expiry date: verify on the packaging the code F7950PL and the expiry date; do not force installation of an expired electrode, as the gel may lose conductivity and compromise the shock.

Paediatric provision: in centres with a child population always keep a pair of paediatric electrodes with the AED, periodically check that the device is operational with the pads in date and immediately replace the pair after any real use.

Variants within the Cahnos catalogue: F7950WPL (same compatible paediatric family Philips with type W connector, external lead-out cable to leave the device pre-connected), F7952PL (paediatric limited-energy electrode for Physio-Control LIFEPAK 500/1000 with pink connector), F7957PL (Cardiac Science Powerheart / GE Responder AED) and F7959PL (CU Medical i-PAD NF1200/NF1201), all limited-energy.

Maintenance: manage stock rotation by expiry date (24–30 months depending on batch) and always replace the electrode after use or upon expiry. For bulk orders or tenders, contact the technical department of Cahnos quoting the exact model and serial number of the device.

Additional information

Product type

Paediatric multifunction limited-energy defibrillation electrode, 1 pair (2 pieces), disposable

Compatibility

Philips (HeartStart / HeartStream / HeartStar)

Compatible models

HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A)

Manufacturer reference

Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads; GUDID UDI-DI 00884838023734)

Indicated population

Under 8 years or weight below 25 kg (55 lb)

Energy

Limited energy: attenuation from 150 J to 50 J (SMART Biphasic therapy); supported shocks 50–100 J

Functions

External defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring (max. 24 h)

Total surface

75 cm² per electrode (paediatric series)

Active surface

40 cm² per electrode (paediatric series)

Adhesive ring width

0.7 cm

Supported shocks

50–100 J (paediatric series)

Conductive gel

Low-impedance conductive adhesive gel

Conductive material

Metal foil

Backing material

Medical foam (1 mm thick)

Connector

Safety anti-shock connector (Philips)

Composition

Latex-free and phthalate-free

Use

Single patient, non-reusable

Presentation

1 pair in heat-sealed polyethylene and aluminium bag; 5 pairs per box (paediatric series)

Commercial format

5 pairs per box (paediatric series)

Regulatory standard

Medical device class IIb (Directive 93/42/EEC); ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993

Shelf life

24–30 months from production date per batch

Expiry

As indicated on the packaging; do not use after the expiry date

Storage

Dry place, between +5 °C and +35 °C; do not use if the packaging is open or damaged

Connection

Safety anti-shock connector (Philips)

Sterilisation

Non-sterile

Operating temperature

0 °C to +50 °C