Description
1. Commercial Introduction
Keep your Weinmann MEDUCORE defibrillator or CardiAid CT0207 AED always ready for action with this pair of standard adult defibrillator electrodes with external cable (reference Cahnos F7964W). It is the replacement consumable indicated for emergency services, ambulances, health centres, companies and any cardioprotected space equipped with these devices.
Thanks to its external cable configuration, the pair can be left ready with the connector accessible: at the critical moment, simply remove the packaging protector and place the pads following the anatomical diagrams on the pouch, minimising preparation time. It is supplied as a pair (2 units) in an individual heat-sealed package, ready to be integrated into your centre’s cardiopulmonary resuscitation protocols.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock and enables emergency electrocardiogram monitoring. This pair is indicated for adult patients (over 8 years of age or weighing more than 25 kg) and connects to the electrode socket of Weinmann MEDUCORE Easy / Standard and CardiAid CT0207 devices via its external cable configuration with anti-shock connector.
The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal energy transmission; the inner metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest, minimising the risk of burns. The recommended placement is anterolateral (sternum-apex), according to the diagrams on the packaging. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7964W |
| Brand / compatible equipment | Weinmann MEDUCORE Easy and MEDUCORE Standard; CardiAid CT0207 AED (RS/RF); check connector fit in the device socket |
| Product type | Pair of adult multifunction defibrillator electrodes, standard, with external cable |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Cable configuration | External cable, approximate length 120–130 cm in the series |
| Total surface (per electrode) | 148 cm² |
| Active surface (per electrode) | 95 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks (50 to 360 J) |
| Functions | Defibrillation, synchronized cardioversion and ECG monitoring (max. 24 h) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector or clip (depending on model), with external cable |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Shelf life | Approx. 30 months from the date of manufacture; expiry date indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Presentation | Price per pair of electrodes (2 units); 1 pair in heat-sealed polyethylene and aluminium pouch (10 pairs per box in adult series) |
Note: the values for active surface area (95 cm²), total area (148 cm²), shocks (50 to 360 J) and adhesive ring (1 cm) correspond to the standard adult electrode series with external cable in this product family; confirm them with the technical data sheet of the supplied batch and the device manual before use in tendering.
4. Equivalences and Cross-References
This electrode (SKU F7964W) is a consumable compatible with Weinmann MEDUCORE Easy / Standard defibrillators and CardiAid CT0207, not an original spare part certified by the device manufacturer: the connector, analysis algorithm and energy management are proprietary to each defibrillator. The factory original spare parts for these devices are listed below as a comparative reference.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Weinmann (original spare part) | Adult defibrillation electrodes MEDUCORE Standard² (self-adhesive, disposable) | WM45418 | Comparable by category (not interchangeable 1:1) | Factory original spare part for the MEDUCORE Standard² series; according to the authorised technical distribution data sheet it has an electrode surface area of 86 cm² and a 1.3 m cable (or 0.6 m depending on version), compared with the 95 cm² active surface area and 120–130 cm cable of this compatible spare part. Code consistently confirmed across multiple authorised Weinmann spare parts distributors. Verify the exact device version (Standard, Standard² or Easy) before replacing. |
| Weinmann (original spare part) | Adult electrodes for MEDUCORE Easy | WM40116 | Comparable by category (not interchangeable 1:1) | Factory original spare part for the MEDUCORE Easy defibrillator; code confirmed in authorised technical distribution. Cable length, surface area and geometry specific to the Easy version; check connector, expiry date and fit in the device socket before any replacement. |
| CardiAid (original spare part) | Adult defibrillation electrodes for AED CT0207 (semi-automatic RS and automatic RF) | CA-10ES | Comparable by category (not interchangeable 1:1) | Factory original spare part for the CardiAid CT0207 defibrillator; pre-connected presentation with proprietary CardiAid connector. Code confirmed in authorised distribution (CardiAid and technical distributors). Devices manufactured from 2019 onwards incorporate an updated connector that must be verified before purchase. |
There is no equivalence table validated by the manufacturer that authorises the replacement of original Weinmann or CardiAid electrodes with third-party pads: the connector, analysis algorithm and energy management are proprietary to each defibrillator. Any replacement must be validated against the technical data sheet and instructions for use of the device; consult the technical department of Cahnos before use in tendering.
Within the Cahnos catalogue there are other references in the same family (standard adult with external cable) intended for equipment from other manufacturers, not interchangeable with each other as the connector and algorithm differ: F7959W (CU Medical / CMOS Drake), F7966W (Defibtech), F7967W (Schiller), F7969W (Mediana HeartOn), F7984W (Instramed) and F7952W (Physio-Control / Bexen / Cardioline / Mindray). The paediatric version for the same Weinmann device is F7964PW.
5. Selection tips and variants of interest
- Confirm the device model: verify that your defibrillator is a Weinmann MEDUCORE Easy, MEDUCORE Standard (or Standard²) or a CardiAid CT0207 (RS/RF) and that the external cable connector fits correctly in the device socket before clinical use; if in doubt, contact the technical department of Cahnos stating the model and serial number.
- Do not confuse the adult and paediatric versions: this pair is for adult patients (over 8 years / over 25 kg). For the paediatric population with the same Weinmann device, use reference F7964PW (paediatric with external cable).
- External cable: this configuration leaves the connector accessible for plugging the pair into the device at the precise moment, reducing preparation time in cardioprotected spaces and ambulances.
- Check the expiry date: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with a damaged pouch must not be used because the gel may lose conductivity.
- Plan your replenishment: as this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to expiry date (FIFO). For volume orders or tenders, contact the technical department of Cahnos stating the exact model and serial number of your defibrillator.