1. Commercial Introduction
Keep your defibrillator always operational with this pair of adult radiotransparent defibrillator electrodes with internal cable (reference Cahnos F7753), the ideal replacement and replenishment consumable for emergency services, critical care units, ambulances, operating theatres, haemodynamics and cardioprotected areas with equipment from the Agilent Hewlett-Packard and Laerdal Medical families. With a single reference it covers the most widespread defibrillator-monitor fleets of these brands —CodeMaster 100, XE, XL and XL+, HeartStart 4000, XL/XLT and MRx—, simplifying the stock of resuscitation consumables. Price per pair, individually heat-sealed packaging, immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the equipment analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronised cardioversion, non-invasive transcutaneous pacing (external pacemaker) and emergency ECG monitoring. This adult model, indicated for patients over 8 years of age or weighing more than 25 kg, is constructed with low-impedance conductive adhesive gel, which ensures homogeneous contact with the skin and optimal energy transmission; the inner metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest.
Its radiotransparent condition —in the electrode and in the cable— avoids artefacts in radiography, radioscopy and tomography, so the pads can be kept in place during imaging examinations without interrupting monitoring or therapy. The configuration with integrated internal cable (100 cm) provides a direct, stable and clean connection to the equipment’s electrode socket, with no loose elements during resuscitation. The pads are placed in the anterolateral position (sternum-apex) according to the anatomical diagrams on the sachet and are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
Main features of the series of adult radiotransparent electrodes with internal cable to which this reference belongs:
| Parameter | Specification |
| Reference Cahnos (SKU) | F7753 |
| Product type | Pair of adult multifunction defibrillator electrodes, radiotransparent, with internal cable |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Radiotransparency | Yes (electrode and cable), suitable for imaging procedures |
| Cable configuration | Integrated internal cable (leads-in), 100 cm |
| Functions | External defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring |
| Total surface per electrode | 136 cm² (adult series) |
| Active surface per electrode | 94 cm² (adult series) |
| Adhesive ring width | 0.7 cm |
| Supported shocks | Up to 50 shocks at 360 J |
| ECG monitoring | Up to 24 hours |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector compatible with Agilent HP / Laerdal equipment |
| Composition | Latex-free and phthalate-free · Single-use |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium bag |
| Sales format | 10 pairs per box (adult series) |
| Shelf life | Approx. 30 months from production date (see expiry on the packaging) |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
3.1 Compatibility with defibrillators
| Equipment model | Factory reference | Observations |
| CodeMaster 100 | M2475B | Direct connection |
| CodeMaster XE | M1724A | Direct connection |
| CodeMaster XL | M1723A / M1723B | Direct connection |
| CodeMaster XL+ | M1722A / M1722B | Direct connection |
| HeartStart 4000 | M5500B | Energy in manual adjustment |
| HeartStream / HeartStart XL | M4735A | Requires hands-free electrode cable M3508A |
| HeartStream / HeartStart XLT | M3500B | Requires hands-free electrode cable M3508A |
| HeartStart MRx | M3535A / M3536A | Requires cable M3508A; do not use in fully automatic mode |
| Laerdal AED Trainer 2 | — | Direct connection (training) |
Note: with HeartStart 4000, XL/XLT and MRx do not use in fully automatic defibrillators; the energy is adjusted manually. Always verify the correct insertion of the connector into the equipment socket before clinical use.
4. Equivalences and Cross-References
Reference F7753 is a consumable compatible with the cited Agilent Hewlett-Packard and Laerdal Medical defibrillators, not an original spare part certified by each equipment manufacturer: the connector, the analysis algorithm and the energy management are proprietary. After verifying official manufacturer catalogues and regulatory databases (GUDID/FDA), the following cross-references with codes confirmed in public sources are identified. Any comparison between brands must be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Philips Healthcare | Adult Multifunction Pads, IEC barrel connector (original equipment) | M3502A (alternative code 989803106931) | Comparable by category (not interchangeable 1:1) | Factory original for the CodeMaster 100 (M2475B), XE (M1724A), XL/XL+ (M1723A/B, M1722A/B), HeartStart 4000 (M5500B), XL (M4735A), XLT (M3500B) and MRx (M3535A/M3536A) families. Code confirmed in authorised technical distribution and in the device regulatory database (FDA authorisation K991871). Not radiotransparent (standard version); verify connector, surface and cable before replacement. |
| Philips Healthcare | HeartStart Adult Defibrillator Pads (AED FR2/FR2+ / ForeRunner) | 989803158211 (1 pair) / 989803158221 (5 pairs) | Comparable by category (not interchangeable 1:1) | Original for the AEDs of the HeartStart FR2/FR2+ and ForeRunner family, indicated for patients aged 8 years or more / 25 kg or more; code confirmed on the official Philips website and in GUDID (DI 00884838023710 / 00884838126299). Not radiotransparent; connector and algorithm proprietary to the same AED platform. |
| Cardinal Health (Covidien) | Medi-Trace Cadence adult RTS (radiotransparent) | 22661R (preconnected: 22661PC) | Comparable by category (not interchangeable 1:1) | Adult radiotransparent pair with barrel connector compatible with the Philips/Agilent-HP family (HeartStream/HeartStart XL, XL+, MRx and CodeMaster); silver/silver chloride gradient technology and 10 pairs per box. Code confirmed in GUDID (DI 20884527022370) and manufacturer catalogue. Verify connector and surfaces before replacement. |
| Nissha Medical Technologies (HeartSync) | 97-DEFIB F/G C100AC-PHILIPS (D311) | 6600015H (ref. C100AC-PHILIPS) | Comparable by category (not interchangeable 1:1) | Adult radiotransparent pair with internal cable (leads-in) and connector comparable to the original Philips; total surface area of 371 cm², carbon conductive element, 1.5 m cable and 36-month service life. Code confirmed in the manufacturer’s official catalogue. Not suitable for all models: verify connector and compatibility with the exact equipment. |
There is no equivalence table validated by the manufacturer that authorises the replacement of original electrodes with third-party pads: the connector, the analysis algorithm and the energy management of each defibrillator are proprietary. Any replacement must be validated against the technical data sheet and instructions for use of the defibrillator, checking one by one the equipment connector, the energy range, the target population and the expiry date of the packaging. Consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
- Confirm the model and connector of the equipment: verify on the defibrillator plate the exact model (CodeMaster 100/XE/XL/XL+, HeartStart 4000, XL/XLT, MRx) and check the required therapy cable (M3507A / M3508A on models that require it) before storing the spare part.
- Operating mode: on HeartStart 4000, XL/XLT and MRx the energy is adjusted manually; do not use this pair in fully automatic defibrillators. On the Laerdal AED Trainer 2 it can be used for training with direct connection.
- Choose radiotransparent when imaging tests are anticipated: if the patient may require radiography, radioscopy or CT during care, this version avoids artefacts without removing the electrode. If imaging is not required, consider the standard version of the same platform (F7953), at lower cost.
- Variants of the same family on Cahnos: paediatric radiotransparent version with internal cable (F7753P) for patients under 8 years or 25 kg; radiotransparent version with external cable (F7753W) if you prefer the pre-connected pair ready for emergencies. Reference F7750 covers the AED families Philips (ForeRunner/FR) and is not interchangeable with this one due to proprietary connector.
- Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the expiry date on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO).