Description
1. Commercial Introduction
The NM 55 RLI electrode (reference Cahnos 14709) is the reference consumable of the Top Trace line for cardiac monitoring and stress testing in adult patients: a disposable ECG electrode made of circular foam of Ø 55 mm with conductive liquid gel, offering one of the largest adhesive surfaces in the family. Its value for money and high-turnover format (bag of 25 units, with a minimum box of 500) make it the ideal choice for cardiology departments, ergometry rooms, sports medicine, ICU and purchasing centres that need a reliable electrode, comfortable for the patient and with excellent cost per procedure.
2. Technical Explanation
A surface ECG electrode acts as an interface between the patient’s skin and the monitor or stress test recorder: it captures the electrical activity of the heart and transmits it with the least possible distortion. The quality of the recording depends on three factors: adhesion (the electrode must remain fixed throughout the study, even with sweating), conductivity (low skin-electrode contact impedance) and stability against movement.
The foam support is flexible, light and moisture-resistant: it protects the gel and the sensor, adapts to the contour of the chest and withstands the demanding conditions of a stress test. The liquid gel, formulated with non-irritant potassium chloride, ensures rapid signal acquisition and low impedance from the start of the recording, reducing the movement artefacts typical of ergometry. The Ag/AgCl sensor and the centred press stud complete a hypoallergenic, non-sensitising and latex-free assembly, manufactured with foam coated with medical-grade acrylic copolymer.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | 14709 |
| Model / family | NM 55 RLI (Top Trace line) |
| Type | Disposable ECG electrode for monitoring and stress testing |
| Backing material | White foam with aggressive medical-grade acrylic adhesive |
| Gel | Conductive liquid gel (potassium chloride, non-irritant) |
| Sensor material | Ag/AgCl |
| Diameter | Ø 55 mm (circular format) |
| Connection type | Centred press stud |
| Application | Cardiac monitoring and stress testing in adults |
| Presentation | Bag of 25 units |
| Minimum sales unit | Box of 500 units |
| Use | Single-use, disposable; for clinical and hospital use |
| Composition | Latex-free; hypoallergenic, non-irritating, non-sensitising |
| Regulatory standard | ANSI/AAMI EC 12:2000/(R)2005 (performance requirements); medical device with CE marking |
Indicative electrical values (according to batch technical data sheet)
| Electrical parameter | Typical batch value | ANSI/AAMI EC 12 limit |
|---|---|---|
| DC offset voltage | 0.12 mV | ≤ 100 mV |
| AC impedance at 10 Hz | 60 Ω | ≤ 2000 Ω |
Note: electrical values are provided with the supplied batch certificate; confirm with the batch documentation before use in tendering.
4. Equivalences and Cross-References
Reference 14709 (NM 55 RLI) is the Cahnos distribution code for this adult foam electrode with liquid gel; there is no manufacturer-validated equivalence table authorising a 1:1 cross-reference with third-party brands. The following references share the same category (disposable foam ECG electrode or similar backing with wet/liquid gel and button connection for monitoring and stress testing in adults) and have a confirmed code in the official manufacturer catalogue, GUDID regulatory database or authorised technical distribution. They are classified as comparable by category, not interchangeable 1:1: a field-by-field check (diameter, gel type, connector position and adhesive) must be carried out before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Leonhard Lang (Skintact®) | F-55 – round foam electrode with Aqua-Wet™ liquid gel (adult) | F-55 (order code depending on market: 23050-00 / 58018 in distribution channels) | Comparable by category (not interchangeable 1:1) | Model code confirmed in the official Leonhard Lang catalogue and multiple authorised distributors: PE foam, Aqua-Wet™ liquid gel, Ø 55 mm, press stud, Ag/AgCl sensor and indicated for stress ECG, Holter and monitoring. It is the closest homologous reference as it matches in backing, gel type, diameter and indication; verify the connector position and adhesive of the specific batch, and confirm the numeric order code with the manufacturer before use in tendering. |
| Ambu (Denmark) | BlueSensor R – foam electrode with wet gel (adult) | R-00-S / R-00-S/25 (GUDID 05707480022205) | Comparable by category (not interchangeable 1:1) | Code confirmed in GUDID and official Ambu catalogue: foam, highly conductive wet gel, Ag/AgCl sensor and offset connector (not centred); suitable for stress testing and short/medium duration monitoring. Dimensions 57 × 48 mm (48 mm effective): smaller diameter and off-centre connector compared to the Ø 55 mm with centred press stud of this reference; validate the cable fit before replacing. |
| 3M / Solventum | Red Dot™ Monitoring Electrode – foam with adhesive gel (solid gel) | 2560 (also V2560-5) | Comparable by category (not interchangeable 1:1) | Reference widely documented in technical distribution and official catalogue: foam with solid adhesive gel, button connection and validated prolonged use. Differences: solid gel compared to the liquid gel of this reference and rectangular geometry of 4 × 3.5 cm compared to the circular 55 mm format; verify contact surface and adhesion before substitution. |
| Cardinal Health (Kendall™) | Medi-Trace™ 530 Series Foam Electrodes (adult, diaphoretic) | 31013926 (series 530) | Comparable by category (not interchangeable 1:1) | Code confirmed in official Cardinal Health catalogue and authorised distribution: foam with conductive adhesive hydrogel (solid gel), button connection, indicated for stress, Holter and diaphoretic monitoring, compliant with ANSI/AAMI EC 12. Differences: solid gel and drop shape of approx. 44 mm compared to the circular liquid gel of 55 mm. |
There is no equivalence table validated by the manufacturer that authorises 1:1 substitution between brands of monitoring and stress test electrodes: the type of backing, the type of gel (liquid or solid), the diameter and the position of the connector affect adhesion and signal quality. Any substitution must be validated against the technical data sheet and instructions for use of the electrode and monitoring system manufacturer, checking one by one diameter, backing, gel and connector; consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
Choose the diameter according to the protocol: this Ø 55 mm electrode offers one of the largest adhesive surfaces in the family, ideal for adult patients and intense exercise conditions. If you prefer the same backing and gel in a slightly smaller format, you have the foam electrode with liquid gel of Ø 50 mm (ref. 0590024, model ST 50 RLI, box of 600 units) or the FS 50 L of Ø 50 mm (ref. 11312).
Liquid gel versus solid gel: the liquid gel of this reference provides very fast signal acquisition and low initial impedance, ideal for stress testing. If your department performs prolonged monitoring (Holter) or recordings with extreme sweating, consider the versions of the same family with solid gel, such as the ST 50 RFI electrode of Ø 50 mm (ref. 0590028) or the radiotransparent RT 50 RFI (ref. 0590022).
Check the connector: this electrode features a centred snap button. Check that the patient cable of your ergometer or monitor is compatible with button connection before placing the order; if your fleet uses forceps or other types of connectors, consult the range of electrodes with tab and the connection adapters from Cahnos.
Preparation and storage: clean and dry the skin before application and avoid creams or oily residues to optimise contact impedance. Keep the bags closed in a dry, cool place (5–32 ºC), rotate stock on a FIFO basis and respect the expiry date: liquid gel may lose conductivity if the packaging is opened or deteriorates. For volume orders, complete supplies or tenders, contact the technical-commercial department of Cahnos indicating the model of your monitor fleet and the type of connection.