Description
1. Introduction
The Paediatric SpO₂ pulse oximeter sensor with 9-pin connection (Ref. 18632) is the reference consumable for services that monitor blood oxygenation in paediatric patients with equipment compatible with Nellcor technology. Its 9-pin connector (DB9) ensures direct and secure integration with most bedside, transport and multiparameter monitors that use this standard, with a patient profile comparable to the original Nellcor D-20 reference (10–50 kg).
Presented in a box of 25 single-use units per patient, it is designed for hospitals, paediatric clinics, emergency departments and paediatric intensive care units that need to maintain a constant rotation of sensors with an adjusted cost per unit, immediate availability and maximum hygiene against cross-contamination.
2. Technical explanation
Pulse oximetry non-invasively measures the functional oxygen saturation of arterial haemoglobin (SpO₂) and pulse rate by spectrophotometry: the sensor emits red light (660 nm) and infrared light (905 nm) through the vascular bed and detects differential absorption according to the degree of blood oxygenation. The signal is transmitted to the monitor via the 9-pin connector (DB9), which follows the electrical scheme of the Nellcor/Oxismart standard.
This sensor is specifically designed for the paediatric population, with a size profile and adhesive fixation system adapted to that population, allowing comfortable and stable placement during continuous monitoring. As it is a single-use component per patient, it eliminates the risk of cross-contamination and simplifies service logistics in high-turnover units.
3. Technical Specifications
| Sensor type | SpO₂ pulse oximeter sensor for paediatric patients |
| Compatibility | Paediatric Nellcor systems (Nellcor/Oxismart protocol) |
| Connection | 9-pin (DB9) |
| Patient population | Paediatric (profile comparable to original reference Nellcor D-20, 10–50 kg) |
| Applications | Monitoring of blood oxygen saturation (SpO₂) and pulse rate |
| Contents | Box of 25 units |
| SpO₂ measurement range | 35–100 % (reference value of the Nellcor DB9 family; confirm in batch IFU) |
| SpO₂ accuracy | ≤ 3 % in range 70–100 % (ISO 9919 / ISO 80601-2-61, reference value) |
| Pulse rate range | 30–250 ppm (reference value; confirm in batch IFU) |
| Wavelengths | Red light 660 nm / infrared 905 nm (reference value) |
| Material | Latex-free |
| Sterility | Non-sterile (single-use product per patient) |
| Storage | Store in a dry place, protected from direct sunlight and extreme temperatures |
Note: the range and accuracy values marked as reference come from technical data sheets of the same family of Nellcor sensors with DB9 connector; always confirm them on the labelling and IFU of the supplied batch.
4. Equivalences and Cross-References
SKU 18632 is Cahnos’s own management code. The following references have been cross-checked against authorised distribution catalogues, public product documentation and compatible sensor manufacturers; none is interchangeable 1:1 with this sensor without prior verification, since the electrical scheme and communication protocol must be confirmed with the specific monitor installed at the centre.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Nellcor (Covidien/Medtronic) | Oxisensor II – paediatric disposable adhesive sensor | D-20 | Comparable by category (not interchangeable 1:1) | Original sensor from the protocol manufacturer for paediatric patients (10–50 kg according to data sheet), DB9 9-pin connector and Oxismart technology; coding and electronics specific to the Nellcor standard. Confirm protocol (Oxisensor/OxiMax) with the monitor. |
| Nellcor (Covidien/Medtronic) | Nellcor OxiMax – paediatric adhesive sensor | MAX-P (MAX-P-I) | Comparable by category (not interchangeable 1:1) | Paediatric sensor from the original manufacturer with OxiMax memory chip and 9-pin D-Sub connector; requires a monitor compatible with OxiMax technology. Verify protocol compatibility with each monitor before replacing. |
| Medke (manufacturer of compatible sensors) | Medke P1219 – paediatric disposable sensor comparable to D-20 | P1219 | Comparable by category (not interchangeable 1:1) | Paediatric sensor compatible with Nellcor (15–40 kg according to the manufacturer’s data sheet), latex-free, 0.9 m PVC cable and DB9 connector; verified on the manufacturer’s official website (CE / ISO 13485 / FDA). Weight profile different from the reference; confirm suitability with the centre. |
There is no equivalence table validated by the manufacturer that cross-references this sensor with codes from other brands in an interchangeable 1:1 manner; consult the technical department before using in tenders. The ear clip sensors of the Nellcor family are another type of product, with a different geometry and application point, and must not be confused with this paediatric sensor.
5. Selection tips and variants of interest
Choice by population: within the same family of 9-pin Nellcor sensors, the adult sensor (SKU 18631), paediatric (SKU 18632, this product), infant (SKU 18633) and adult/newborn (SKU 18634) are available. If your service works with reusable sensors, consider the paediatric forceps compatible with Nellcor technology (Ref. AB78315040097) or, for neonates, the disposable Y-type wrap-around sensor (Ref. AB78315040078).
Compatibility verification: although the connector is 9-pin (DB9), the Nellcor protocol has variants (Oxisensor/OxiMax). Always indicate the brand and model of the monitor to confirm exact compatibility before placing the order.
Hygiene and storage: as it is a single-use sensor per patient, it eliminates the risk of cross-contamination; do not reuse or reprocess the sensor. Store in a dry place, protected from direct sunlight and extreme temperatures.
Cost-effectiveness: the box of 25 units reduces the cost per sensor and ensures continuous stock in high-turnover units. Consult our technical-commercial team for special B2B conditions by volume and supply contracts.





Reviews