Description
1. Commercial Introduction
The F 55 L electrode (reference Cahnos 11310) is a disposable surface ECG consumable, made of foam with conductive liquid gel, in a circular format of Ø 55 mm. It is indicated for continuous cardiac monitoring and stress testing in adults, where the demands on adhesion and signal stability are highest. Its value for money and high-turnover format (bag of 30 units, with a minimum box of 1,200) make it the reference option for cardiology departments, ergometry rooms, ICUs, primary care consultations and purchasing centres that need a reliable electrode, comfortable for the patient and with excellent cost per procedure.
2. Technical Explanation
A surface ECG electrode acts as an interface between the patient’s skin and the monitor or the stress test recorder: it captures the electrical activity of the heart and transmits it with the least possible distortion. The quality of the recording depends on three factors: adhesion (the electrode must remain fixed throughout the study), conductivity (low skin-electrode contact impedance) and stability against movement.
The foam backing (medical foam) is impermeable to liquids and very flexible: it protects the gel and the sensor, adapts to the contour of the chest and withstands the demanding conditions of a stress test, including intense sweating. The conductive liquid gel ensures rapid signal acquisition and low contact impedance from the start of the recording, reducing the risk of movement artefacts. This is a single-use, latex-free product that must be disposed of after each patient, respecting the expiry date of the packaging so as not to compromise the conductivity of the gel.
3. Technical Specifications
Main features of the F 55 L electrode (reference Cahnos 11310):
| Parameter | Specification |
| Reference Cahnos (SKU) | 11310 |
| Model / family | F 55 L (Medi Sensor Monitoring line) |
| Type | Conventional disposable electrode for monitoring and stress testing |
| Material | Foam (medical foam) with conductive liquid gel |
| Diameter | Ø 55 mm (circular format) |
| Presentation | Bag of 30 units |
| Minimum sales unit | Box of 1,200 units |
| Application | Cardiac monitoring and stress testing in adults |
| Advantages | High adhesion, precise signal, comfortable for the patient |
| Use | Clinical and hospital use; single-use product |
Note: the diameter, material and presentation data are copied from the distributor’s technical documentation; confirm with the supplied batch sheet the exact connector type (snap-type press stud or forceps) for fitting with the patient cable of the monitor fleet.
4. Equivalences and Cross-References
Reference 11310 is the Cahnos distribution code for this adult foam electrode with liquid gel; there is no equivalence table validated by the manufacturer that authorises a 1:1 cross-reference with third-party brands. The following references share the same category (disposable foam ECG electrode with wet/liquid gel and button connection for monitoring and stress testing in adults) and have a code confirmed in the GUDID regulatory database, the manufacturer’s official catalogue or authorised technical distribution. They are classified as comparable by category, not interchangeable 1:1: each field must be verified (diameter, gel type, connector position and adhesive) before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ambu (Denmark) | BlueSensor R – foam electrode with wet gel (adult) | R-00-S / R-00-S/25 (GUDID 05707480022205) | Comparable by category (not interchangeable 1:1) | Code confirmed in GUDID and Ambu’s official catalogue: foam, highly conductive wet gel, Ag/AgCl sensor and offset snap connector; suitable for stress testing and monitoring. Dimensions 57 × 48 mm (48 mm effective): smaller diameter and offset connector compared to the Ø 55 mm of this reference; validate the cable fit before substituting. |
| 3M / Solventum | Red Dot™ Monitoring Electrode with foam and adhesive gel (sticky gel) | 2560 (also V2560-5; GUDID DI 00707387789305) | Comparable by category (not interchangeable 1:1) | Code confirmed in GUDID and official catalogue: radiolucent foam with solid adhesive gel, snap connection and prolonged use validated up to 5 days. Differences: solid gel compared to the liquid gel of this reference and rectangular geometry of 4 × 3.5 cm compared to the circular format of 55 mm; verify contact surface and adhesion before substitution. |
| Cardinal Health (Kendall™) | Medi-Trace™ 530 Series Foam Electrodes (adult, diaphoretic) | 31013926 (series 530) | Comparable by category (not interchangeable 1:1) | Code confirmed in Cardinal Health’s official catalogue and authorised distribution: foam with conductive adhesive hydrogel (solid gel), snap connection, indicated for stress, Holter and diaphoretic monitoring (suggested use less than 72 h), compliant with ANSI/AAMI EC 12. Differences: solid gel and drop shape of approx. 44 mm compared to the circular liquid gel of 55 mm. |
| Leonhard Lang (Skintact®) | F-55 – monitoring electrode with foam backing (PE foam) and liquid gel | F-55 (official catalogue; 30 units/bag, 1,200 units/box) | Comparable by category (not interchangeable 1:1) | Code confirmed in Leonhard Lang’s official catalogue: PE foam, liquid gel, Ø 55 mm, large adhesive surface and same presentation of 30/1,200 units. It matches in backing, gel type, diameter and format; verify the connector position and the adhesive of the specific batch before considering any substitution. |
| Leonhard Lang (Skintact®) | FS-50 – exercise stress test electrode with foam and Aqua-Wet™ liquid gel | FS-50 (official catalogue; 30 units/bag, 1,200 units/box) | Comparable by category (not interchangeable 1:1) | Code confirmed in Leonhard Lang’s official catalogue: PE foam, Aqua-Wet™ liquid gel, snap connector, Ø 50 mm, reinforced adhesive, indicated for stress testing, emergency care and echocardiography. Differences: diameter 50 mm compared to 55 mm. |
| Vermed (Nissha Medical) | Performance Plus™ Foam Wet Gel Electrode – stress testing and Holter | A10005-60 (60 units/bag) | Comparable by category (not interchangeable 1:1) | Code confirmed in authorised technical distribution: drop-shaped foam of 38 mm (1.5 in.), wet gel with high chloride content, Ag/AgCl sensor, aggressive hypoallergenic adhesive and metallic snap connection; indicated for stress testing and Holter. Differs: geometry and diameter compared to the 55 mm circular format. |
There is no equivalence table validated by the manufacturer that authorises a 1:1 substitution between brands of monitoring and stress test electrodes: the type of backing, the type of gel (liquid or solid), the diameter and the position of the connector affect adhesion and signal quality. Any substitution must be validated against the technical data sheet and instructions for use of the electrode manufacturer and the monitoring system, checking one by one diameter, backing, gel and connector; consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
Choose the diameter according to the protocol: this Ø 55 mm electrode offers the largest adhesive surface in the family for adult patients and intense exercise conditions. If you prefer the same backing and gel in a slightly smaller format, you have the foam electrode with liquid gel of Ø 50 mm (ref. 11312, model FS 50 L) and the circular Ø 50 mm electrode for monitoring and ECG (ref. 15492, model F 50 L), both in bags of 30 units.
Liquid gel versus solid gel: the liquid gel of this reference provides very fast signal acquisition and low initial impedance, ideal for stress testing. If your department performs prolonged monitoring or recordings with extreme sweating, consider the versions of the same family with solid gel and radiotransparent snap, such as the RT 50 electrode (ref. 0590022, Ø 50 mm) or the foam one with solid gel for stress (ref. 0590028).
Variants within the Cahnos catalogue: for stress testing you also have the foam electrode with liquid gel of Ø 50 mm with high adhesiveness (ref. 0590024, model ST 50 RLI) and the Ø 55 mm electrode with liquid gel from the Top Trace line (ref. 14709, model NM 55 RLI); for prolonged ambulatory monitoring, the non-woven fabric (TNT) electrode of Ø 55 mm with liquid gel (ref. 15552, model W 601 L) is the closest breathable option. All share the same clinical indication but differ in backing, gel or diameter: adapt the choice to your monitor fleet and the type of connector on your patient cables.
Preparation and storage: clean and dry the skin before placement and avoid creams or oily residues to optimise contact impedance. Keep the bags closed in a dry, cool place, rotate stock on a FIFO basis and respect the expiry date: the liquid gel can lose conductivity if the packaging is opened or deteriorates. For volume orders, complete supplies or tenders, contact the technical-commercial department of Cahnos indicating the model of your monitor fleet and the type of connection.