Description
1. Introduction
The PGLAR polyglactin 910 RAPID multifilament absorbable suture (PREMISUT RAPID range) is a high-performance surgical reference designed for professionals who need a braided, flexible thread with early and fully predictable strength loss in the operating theatre. It is supplied mounted with a ½ circle cylindrical body taper point needle of 48 mm (HR 48), USP size 2, and a 90 cm colourless ribbed thread, in a box of 12 units aimed at hospitals, ophthalmology clinics and B2B purchasing organisations.
2. Technical explanation
This suture is made from polyglactin 910 in its low molecular weight variant (RAPID), an absorbable synthetic copolymer composed of 90% glycolide and 10% L-lactide that accelerates strength and mass loss compared to conventional polyglactin 910. Its braided multifilament structure with ribbed finish and coating provides very smooth handling, secure knot tying and controlled sliding.
Absorption occurs by hydrolysis, i.e. polymer degradation in contact with bodily fluids, over a predictable period of 40 to 45 days, with minimal tissue reaction. The tensile strength retention profile is 50% at 5 days and 0% at 14 days, making it particularly indicated for tissues that require only short-term support: skin and mucosa approximation, ligatures, episiotomies, paediatric surgery and ophthalmic surgery. The ½ circle taper point needle penetrates soft tissues by separating the fibres without tearing them. The product is supplied sterile by gamma radiation, is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Characteristic | Specification |
|---|---|
| Material | Polyglactin 910 (PGLAR), copolymer 90% glycolide / 10% L-lactide, low molecular weight variant (RAPID) |
| Thread structure | Braided coated multifilament, fluted finish |
| Gauge | USP 2 (metric 5) |
| Thread colour | Colourless |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Taperpoint |
| Needle size | HR 48 mm |
| Thread length | 90 cm |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days |
| Tensile strength retention | 50 % at 5 days / 0 % at 14 days |
| Sterilisation | Gamma radiation |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Factory reference | 2L23HR4802 |
| SKU Cahnos | AM2L23HR4802 |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 5 days | 50 % |
| 14 days | 0 % |
Complete absorption by hydrolysis within 40 to 45 days, with minimal tissue reaction, excellent initial tensile strength and optimal knot tying.
4. Equivalences and Cross-References
Rapid-absorption polyglactin 910 (low molecular weight) is one of the most widely used rapid-absorption synthetic absorbable materials on the market. The reference analysed is comparable at category level with other manufacturers’ braided polyglactin 910 RAPID sutures, both in composition and in absorption profile (≈42 days) and tension retention (≈50% at 5 days).
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson) | VICRYL RAPIDE (polyglactin 910) | W9962 | Comparable by category (not interchangeable 1:1) | Same rapid-absorption material, 90 cm braided colourless thread and ½ circle cylindrical body needle, but USP size 2-0 as opposed to USP 2 of the reference analysed; verify the size on the manufacturer’s official datasheet before substituting. |
| Ethicon (Johnson & Johnson) | VICRYL RAPIDE (polyglactin 910) | W9978 | Comparable by category (not interchangeable 1:1) | Same material and rapid absorption profile, 90 cm colourless thread and 48 mm ½ circle taper point cylindrical needle (geometry matching the HR 48), but USP size 2-0 as opposed to USP 2 of the reference analysed; confirm the size with the manufacturer. |
| Ethicon (Johnson & Johnson) | VICRYL RAPIDE (polyglactin 910) | VR945 | Comparable by category (not interchangeable 1:1) | Same RAPID line, 90 cm colourless thread, 36 mm ½ circle taper point needle and USP size 2-0; the needle geometry (36 mm) differs from the 48 mm HR. Verify needle measurement on the official datasheet. |
There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender. The industry benchmark homologous ranges are the rapid-absorption braided polyglactin 910 sutures from B. Braun (Safil Quick type line), as well as the absorbable braided lines from Medtronic and the HR series from Asian suture manufacturers, whose specific codes for USP size 2 with HR 48 needle must be confirmed with each manufacturer.
5. Selection tips and variants of interest
To choose the most appropriate suture, consider the wound support time required by the tissue and the necessary size according to the anatomical site. USP size 2 with a 48 mm HR needle is a common choice for skin and mucosa approximation, ligatures and ophthalmic surgery, with early strength loss that avoids suture removal in many indications:
- AM2L23HR4002: PGLAR RAPID USP 2 with 40 mm HR needle, 90 cm colourless thread (same family and size, shorter needle for restricted fields).
- AMP2L23HR4802: PGLAR RAPID HR 48 USP 2 in PRIME quality, 90 cm colourless thread (same geometry, premium range).
- AM2L23HR4810: PGLAR RAPID HR 48 in USP size 0, 90 cm colourless thread (same needle, smaller size).
- AM223HR4802: rapid polyglycolic acid suture (PGAR) HR 48 USP 2 (alternative material within the rapid absorption category).
- 1L23HR4802: conventional polyglactin 910 (absorption 56-72 days and longer strength retention) when wound healing requires support beyond 2 weeks.
- PDO (polydioxanone) family: for prolonged strength retention (180-210 days) in tissues subjected to sustained tension.
All references are supplied in a box of 12 units. As it is a sterile single-use product, check the integrity of the packaging before each procedure. For special B2B conditions and volume orders, contact our technical-commercial team.