Description
1. Commercial Introduction
The multifilament absorbable suture PGLAR of Polyglactin 910 in RAPID variant (range PREMISUT RAPID) is the reference for accelerated absorption surgical closure for hospitals, general surgery operating theatres, emergency departments, ophthalmology clinics and B2B purchasing centres that need a braided, flexible thread with early and fully predictable loss of strength in the operating theatre.
Its ½ circle needle with triangular reverse cutting point of 17 mm (HS 17 profile) ensures clean and controlled penetration in fine and firm tissues, while the undyed 45 cm thread offers maximum discretion on the surgical field. A practical and cost-effective presentation, in a box of 12 units, designed for fast-healing tissues and for microsurgery and ophthalmology techniques.
2. Technical Explanation
Polyglactin 910 (PGLAR) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, spun into a braided multifilament structure coated with a grooved finish that favours thread sliding, smooth handling and a firm, secure knot.
In the RAPID variant (low molecular weight polyglactin 910), the thread is formulated to lose tensile strength in an accelerated and programmed manner: it retains 50% of the initial strength at 5 days and 0% at 14 days, with complete absorption by hydrolysis between 40 and 45 days. This profile makes it especially suitable for superficial tissues with rapid healing (skin closure, mucous membranes, subcutaneous tissue), ligation of small-calibre vessels and conjunctival suture in ophthalmic surgery.
The reverse cutting needle generates a clean entry hole by orienting the cut away from the wound, which preserves the skin edge and prevents the cut from enlarging in the puncture bed. Sterile single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics (Ref. 2L21HS1730)
| Parameter | Specification |
| Material | Polyglactin 910 (PGLAR), copolymer 90 % glycolide / 10 % L-lactide, RAPID variant |
| Thread structure | Braided coated multifilament, fluted finish |
| Quality | Standard (PREMISUT RAPID range) |
| Gauge | USP 3/0 (metric 2) |
| Thread diameter | 0.20 – 0.249 mm (indicative, according to Pharmacopoeia) |
| Thread colour | Colourless |
| Thread length | 45 cm |
| Needle shape | ½ circle (1/2, 160°) |
| Needle tip | Triangular, reverse cutting |
| Needle size | HS 17 mm |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Commercial format | Box of 12 units (minimum order 30 boxes) |
| Factory reference | 2L21HS1730 · SKU Cahnos AM2L21HS1730 |
3.2. Absorption profile and tensile strength retention
| Parameter | Value |
| Absorption type | Hydrolysis |
| Absorption time | 40 – 45 days |
| Tensile strength at 5 days | 50% of initial strength |
| Tensile strength at 14 days | 0% (total loss of tensile support) |
| Absorption rate | Predictable and uniform |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 3/0 | 2 | 0.20 – 0.249 mm (indicative, according to Pharmacopoeia) |
4. Equivalences and Cross-References
Rapid absorption polyglactin 910 (low molecular weight) is the most widespread multifilament absorbable category in the surgical market. This reference shares material, USP 3/0 gauge, undyed colour and rapid absorption profile with the homologous lines of the main manufacturers in the sector. Only one code is included whose attributes have been verified in the manufacturer’s official catalogue and authorised distributors:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (braided polyglactin 910, colourless) | W9931 – USP 3/0, FS-2 needle 19 mm, ⅜ circle, reverse cutting, 75 cm | Comparable by category (not interchangeable 1:1) | Same material, USP 3/0 gauge, undyed colour and rapid absorption profile (≈42 days); the needle curvature differs (⅜ circle vs ½ circle), the size (19 mm vs HS 17 mm) and the thread length (75 cm vs 45 cm). Verify geometry in the manufacturer’s official datasheet before substituting. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Homologous ranges in the rapid absorption segment include Safil Quick / Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category by material and absorption profile, although for the exact combination USP 3/0 / HS 17 mm needle / 45 cm their specific codes must be confirmed with the manufacturer. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version.
5. Selection tips and variants of interest
The USP 3/0 gauge (metric 2) is a common choice in ophthalmic surgery and in fine tissues where an optimal balance between strength and fineness is required: conjunctival suture, mucosal approximation, subcutaneous tissue and ligation of small-calibre vessels. The ½ circle reverse cutting needle of 17 mm is especially recommended when a firm and precise passage through the skin or mucous membranes is required without tearing the edge, typical in microsurgery and ophthalmology.
Within the same PGLAR RAPID range, Cahnos offers complementary configurations:
- AM2L22HS1730: same HS 17 needle and USP 3/0 gauge with undyed 75 cm thread (longer thread length).
- AMP2L21HS1730: same configuration (USP 3/0, HS 17, 45 cm) in Prime quality (range PREMISUT RAPID Prime).
- AM2L22HS1740: USP 4/0 gauge with HS 17 needle and 75 cm thread (finer gauge for delicate skin and mucous membranes).
- AM2L22HR1730: USP 3/0 with cylindrical needle conical point HR 17, undyed, 75 cm (for looser tissues).
- AM2L22HRT2630: USP 3/0 with trocar point needle HRT 26, undyed, 75 cm (for fibrous tissues).
- AM2L23HS4001: triangular reverse cutting needle HS 40 in USP 1 gauge (higher tension requirements).
The undyed version offers maximum discretion in the field; if greater visibility is required, request the same configuration in dyed thread on demand. Sterile single-use product: check the integrity of the packaging and the expiry date before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders (minimum 30 boxes) or equivalences from other commercial houses, contact the technical-commercial team of Cahnos.