Description
1. Commercial Introduction
The absorbable multifilament polyglactin 910 (PGLA) suture, reference 1L13HR4810 from the Premisut range, is the reference surgical closure solution for professionals who demand a flexible, resistant thread with fully predictable absorption in the operating theatre. Its ½ circle needle with cylindrical body and conical point of 48 mm (HR 48) passes through soft tissues separating the fibres without cutting them, which minimises trauma and makes it especially suitable for tissue approximation, ligatures and ophthalmic surgery. It is a synthetic absorbable medical device, sterile and single-use, with an excellent quality-price ratio for B2B supply to hospitals, ophthalmic clinics and purchasing centres, guaranteeing continuous supply to the healthcare facility.
2. Technical explanation of the product and procedure
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through tissue, very smooth handling, comfortable sliding and firm, secure knot tying. Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile maintains wound support during the first critical weeks of healing: it retains 75% of tensile strength at 2 weeks, 50% at 3 weeks and 30% at 4 weeks, while the tissue recovers its own resistance. The ½ circle needle, cylindrical body and conical point penetrates without cutting, minimising trauma to soft and friable tissues, while the violet colouration of the thread guarantees optimal visibility on the surgical field. The product is supplied sterile by ethylene oxide (EO), single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference 1L13HR4810
| Characteristic | Specification |
| Reference (SKU) | 1L13HR4810 |
| Range / Quality | Premisut, Normal quality |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Colour | Violet |
| Gauge | USP 0 (metric 3.5) |
| Thread length | 90 cm |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Conical (taper point) |
| Needle size | 48 mm (HR 48) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 56-72 days |
| Sterilisation | Ethylene oxide (EO); single-use, non-reesterilizable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Total absorption | 56-72 days, by hydrolysis |
Excellent tensile strength and knot tying, with a fully predictable absorption rate and minimal tissue reaction.
3.3. Presentation and ordering conditions
| Value | Value |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Format | Sterile, single-use |
4. Equivalences and Cross-References
Polyglactin 910 is one of the most widely used synthetic absorbable materials on the European market. This reference is clinically comparable, in terms of composition, absorption profile and tension retention, to the braided multifilament sutures of the main manufacturers:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (braided coated polyglactin 910, violet) |
| B. Braun | Safil (braided polyglactin 910, violet) |
| Medtronic / Covidien | Absorbable braided polyglactin 910 range |
| SMI | Surgicryl (braided polyglactin 910, violet) |
Within the Cahnos catalogue, the references closest to this configuration (HR 48 mm needle, USP gauge 0, violet) are the version with 75 cm thread (1L12HR4810), the alternative in polyglycolic acid (AM113HR4810) and the equivalent in Prime quality (P1L13HR4810). The equivalences indicated are for guidance; for a formal substitution of references, please consult our technical-commercial department.
5. Selection tips and variants of interest
USP gauge 0 (metric 3.5) is very versatile: suitable for subcutaneous tissue, superficial muscle planes, approximation of soft tissues and ophthalmic surgery procedures, with an excellent balance between initial resistance and predictable absorption during healing of 2 to 4 weeks. The conical point (HR) separates the fibres without cutting them and is ideal for soft tissues; if the tissue is very friable, the most atraumatic option is the blunt point (HRN), while for firm and dense tissues the reverse cutting triangular point (HS) or the trocar point (HRT) is recommended.
| Reference | Variant |
| 1L13HR4810 | PGLA USP 0, conical point HR 48 mm, violet, 90 cm (current product) |
| 1L13HR4801 | PGLA USP 1, conical point HR 48 mm, violet, 90 cm |
| 1L13HR4802 | PGLA USP 2, conical point HR 48 mm, violet, 90 cm |
| 1L13HR4010 | PGLA USP 0, conical point HR 40 mm, violet, 90 cm |
| 1L12HR4810 | PGLA USP 0, conical point HR 48 mm, violet, 75 cm |
| 1L13HRN4810 | PGLA USP 0, blunt point HRN 48 mm, violet, 90 cm |
| 1L13HS4810 | PGLA USP 0, reverse cutting triangular point HS 48 mm, violet, 90 cm |
| 1L13HRT4820 | PGLA USP 2/0, trocar point HRT 48 mm, violet, 90 cm |
For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to guarantee continuous supply to your healthcare facility. For special B2B conditions and volume orders, contact our technical-commercial team.