1. Commercial Introduction
At Cahnos we offer the absorbable multifilament suture of Polyglactin 910 (PGLA), reference 1L28HR1740 from the Premisut range, the professional reference solution for B2B supply to operating theatres, ophthalmic clinics, day surgery units and procurement centres. It combines a synthetic copolymer with predictable behaviour, exceptional handling and great knot security in the approximation of soft tissues, ligatures and ophthalmic surgery. Its ½ circle needle with cylindrical body and conical point of 17 mm ensures atraumatic penetration, while the colourless grooved 70 cm thread offers maximum discretion on the surgical field and smooth sliding during suturing. It is a synthetic absorbable medical device, sterile and single-use, with an excellent value for money for continuous supply contracts.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that favours the passage of the thread through tissue, smooth handling, low shape memory and firm, secure knot tying, even in fine gauges such as USP 4/0 (metric 1.5).
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days and independent of the patient’s enzymatic activity. This profile maintains the necessary tensile strength during the first 2 to 4 weeks of healing, while the tissue recovers its own strength.
The ½ circle needle with cylindrical body and conical point (atraumatic geometry) passes through tissues by separating the fibres without cutting effect, minimising trauma to delicate structures; it is therefore especially recommended in ophthalmic surgery and in the approximation of soft tissues. The product is sterilised by ethylene oxide, is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet for reference 1L28HR1740
| Characteristic | Specification |
|---|---|
| Reference (SKU) | 1L28HR1740 |
| Brand / Quality | Premisut – Normal quality |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Suture type | Synthetic absorbable multifilament |
| Thread structure | Braided multifilament with grooved finish |
| Gauge | USP 4/0 (metric 1.5) |
| Thread length | 70 cm |
| Thread colour | Colourless |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Atraumatic conical (Taperpoint) |
| Needle size | 17 mm (HR 17) |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide (EO), sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot security |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
3.2. Tensile strength retention and absorption profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 30% of initial tensile strength |
| Total absorption | 56-72 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units; minimum order of 30 boxes.
4. Equivalences and Cross-References
Polyglactin 910 is one of the most well-established synthetic absorbable multifilament suture raw materials on the market. This braided, medium-absorption suture is clinically equivalent, in composition, absorption profile (56-72 days) and mechanical behaviour, to the reference lines of the international market: Ethicon (Johnson & Johnson) Vicryl / Vicryl Plus (coated braided polyglactin 910), B. Braun Safil (braided polyglactin 910), Medtronic / Covidien Polysorb (absorbable braided range), SMI Surgicryl 910, Assut Assucryl 910, DemeTECH DemeCRYL / Vitacryl and the Polyglactin 910 line from Laboratorio Aragó (range 4xxxx, Premisut).
| Manufacturer / Company | Equivalent line |
|---|---|
| Ethicon (Johnson & Johnson) | Vicryl / Vicryl Plus (braided coated polyglactin 910) |
| B. Braun | Safil (braided polyglactin 910) |
| Medtronic (Covidien) | Polysorb (absorbable braided range) |
| SMI | Surgicryl 910 |
| Assut | Assucryl 910 |
| DemeTECH | DemeCRYL / Vitacryl |
| Laboratory Aragó | Polyglactin 910 line (range 4xxxx, Premisut) |
Within the Cahnos catalogue, its counterpart in violet with identical configuration (HR 17 mm needle, USP 4/0 gauge and 70 cm thread) is reference 1L18HR1740, and the higher quality Prime version is processed as P1L28HR1740. The indicated equivalences are for guidance only; for a formal substitution of references, please consult our technical-commercial department.
5. Selection tips and variants of interest
The USP 4/0 gauge (metric 1.5) is indicated for fine and delicate tissues with moderate tension: ophthalmic, plastic, paediatric surgery, ligatures and approximation of soft tissues. The ½ circle needle with conical point of 17 mm is the most atraumatic option for friable tissues and mucous membranes; if greater penetration in firm tissues is required, consider the trocar points of the HRT series (ref. 1L28HRT1740 colourless or 1L18HRT1740 violet).
When greater visibility of the thread in the field is required, request the violet variant (ref. 1L18HR1740); if you prefer a more prolonged retention of strength, consider the polydioxanone (PDO) family from Cahnos. Alternative references of the same material with HR17 needle: gauge 3/0 colourless (1L28HR1730), gauge 5/0 colourless (1L28HR1750), versions with 75 cm thread (1L12HR1730 violet) and an alternative in polyglycolic acid PGA with the same needle (AM128HR1740). All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare centre.
As it is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions by volume, institutional rates or equivalents from other commercial houses, contact our technical-commercial team.