Description
1. Commercial Introduction
We make available to operating theatres, general surgery departments, plastic and reconstructive surgery units, dermatology, emergency departments and purchasing centres the absorbable monofilament suture of poliglecaprone 25 (PGCL / caprolactone) from the PREMISUT range, factory reference 322DS1940 (SKU AM322DS1940). This presentation combines a ⅜ circle needle with a reverse cutting triangular body and point of 19 mm (DS 19 profile), USP 4/0 gauge and a 75 cm undyed thread.
It is a medium-absorption thread (90–120 days), with exceptional handling, high knot security and a very smooth passage through tissue, specifically designed for closing skin and tissues that require firm penetration with maximum precision. Synthetic absorbable medical device, sterile and single-use, presented in a box of 12 units, with a very competitive quality-performance ratio for hospital procurement.
2. Technical Explanation
Poliglecaprone 25 is a synthetic absorbable copolymer (glycolide and ε-caprolactone) with a monofilament structure that degrades by hydrolysis, i.e., by cleavage of the polymer chains on contact with organic fluids, over a predictable period of 90 to 120 days and independent of the patient’s enzymatic activity. Being a monofilament, the strand has a smooth grooved surface that reduces friction when passing through tissue, minimises the sawing effect and hinders bacterial adherence, while maintaining high tensile strength and reliable knot tying during the critical phases of wound healing.
The ⅜ circle needle (medium curvature, the most common in cutaneous surgery) with a reverse cutting triangular body and point (reverse cutting) penetrates the skin and fibrous tissues with immediate penetration, without the sawing effect of conventional cutting needles, making it ideal for closures of skin, subcutaneous tissue and fasciae. The USP 4/0 gauge (metric 1.5) provides an excellent balance between fineness, manageability and initial strength, and the undyed grooved colour guarantees maximum aesthetic discretion in superficial and subcutaneous planes. The product is sterilised by ethylene oxide (EO), is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference 322DS1940
| Parameter | Specification |
| Factory reference | 322DS1940 (PREMISUT POLIGLECAPRONE 25 CAPROLACTONE PGCL) |
| SKU Cahnos | AM322DS1940 |
| Range / Quality | PREMISUT, standard quality |
| Material | Polyglecaprone 25 (PGCL / caprolactone), synthetic absorbable monofilament |
| Thread structure | Grooved monofilament |
| Gauge | USP 4/0 (metric 1.5) |
| Approximate diameter | 0.15 – 0.199 mm (indicative according to USP/EP) |
| Thread colour | Colourless grooved |
| Needle shape | ⅜ circle (3/8 circular, approx. 135°) |
| Needle tip | Grooved reverse cutting triangular (Reverse Cutting), body and point |
| Needle size | DS 19 – 19 mm, 1 needle per thread |
| Thread length | 75 cm (single strand with needle) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 90 – 120 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, intradermal closures, ligatures; general, dermatological, plastic and cosmetic surgery |
| Presentation | Box of 12 sterile units |
| Regulatory standard | Conforms to USP and European Pharmacopoeia (EP) gauges |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 1 week | 60 % |
| 2 weeks | 30 % |
| Complete absorption | 90 – 120 days by hydrolysis |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate diameter |
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
4. Equivalences and Cross-References
Poliglecaprone 25 monofilament is one of the most well-established families of synthetic absorbable sutures with medium absorption on the market, with the MONOCRYL® line from Ethicon (Johnson & Johnson MedTech) being the clinical reference in the category. For this specific configuration (PGCL monofilament, USP 4/0 gauge, ⅜ circle reverse cutting needle of 19 mm, 75 cm undyed thread), the following correspondences have been verified against the manufacturer’s official catalogue and authorised distribution sources, with the caveat that the commercial assembly of each house presents differences in needle designation, thread length and presentation, so they are not interchangeable 1:1:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | MONOCRYL® – poliglecaprone 25 monofilament, undyed, USP 4/0, 70 cm, PS-2 needle of 19 mm, ⅜ circle, precision reverse cutting | Y426H | Comparable by category (not interchangeable 1:1) | Confirmed in the official Ethicon catalogue (EPC) and authorised distribution: they match in material (poliglecaprone 25), USP 4/0 gauge, undyed colour, ⅜ circle shape, needle size (19 mm) and reverse cutting point family; they differ in thread length (70 cm vs 75 cm), needle geometry (PS-2 vs DS 19) and commercial presentation. |
| Ethicon (Johnson & Johnson MedTech) | MONOCRYL® – poliglecaprone 25 monofilament, undyed, USP 4/0, 45 cm, PS-2 needle of 19 mm, ⅜ circle, precision reverse cutting | Y496G | Comparable by category (not interchangeable 1:1) | Confirmed in the official Ethicon catalogue and authorised distribution: they match in material, USP 4/0 gauge, undyed colour, PS-2 needle of 19 mm in ⅜ circle with precision reverse cutting; they differ in thread length (45 cm vs 75 cm) and needle geometry (PS-2 vs DS 19). |
There is no equivalence table validated by the manufacturer for an exact 1:1 cross-reference with a DS 19 reverse cutting triangular needle of 19 mm, USP 4/0 gauge and 75 cm undyed thread; consult the technical department before use in tenders. As homologous market families to verify in their official data sheet before substitution, the MONOCRYL® line from Ethicon (poliglecaprone 25), Monosyn from B. Braun (glyconate, different degradation kinetics) and medium-absorption absorbable monofilaments from other suppliers (DemeTECH – DemeCAPRONE –, SMI, Assut) are cited, checking one by one: exact material, USP 4/0 gauge, ⅜ circle curvature, needle size of 19 mm, reverse cutting point family, thread length of 75 cm, undyed colour and current CE/MDR regulatory classification. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
The USP 4/0 gauge with a 19 mm reverse cutting needle is a reference combination for skin closure in general surgery, plastic surgery, dermatology and emergency departments: the reverse cutting triangular point penetrates the epidermis firmly and without sawing, while the 75 cm undyed thread offers aesthetic discretion and sufficient working length for simple interrupted and continuous cutaneous sutures. If the procedure requires a reverse cutting needle with a larger radius, consider the higher DS profiles in the same range; if you prefer an atraumatic cylindrical needle for soft tissues, consider the conical point variants (HR profile) within the PGCL catalogue.
| Reference (SKU) | Variant of interest |
| AM322DS1940 (322DS1940) | PGCL monofilament, DS 19 mm reverse cutting needle, USP 4/0, undyed, 75 cm (product analysed) |
| AMP322DS1940 | PGCL Prime quality, same configuration (superior quality) |
| AM321DS1940 | PGCL, DS 19 mm needle, USP 4/0, undyed, 45 cm (shorter thread) |
| AM322DS1930 | PGCL, DS 19 mm needle, USP 3/0, undyed, 75 cm (larger gauge) |
| AM321DS1930 | PGCL, DS 19 mm needle, USP 3/0, colourless, 45 cm (larger gauge and shorter thread) |
All references are supplied in boxes of 12 units. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, store the suture in a cool, dry place below 30 °C and handle the thread with gloves to avoid altering its mechanical properties. For special B2B volume conditions, hospital supply contracts or validation of equivalence with other commercial brands, contact our technical-commercial team.