Description
1. Product Presentation
The absorbable multifilament suture PGA Prime is the reference solution for operating theatre services and purchasing departments looking for a high-performance synthetic reabsorbable thread with an excellent balance between handling, knot security and cost. In the double needle ¼ circle spatulated Micropoint 8 mm (2xVSP) configuration, USP 4/0 gauge and 45 cm violet thread, it is designed for precision procedures on delicate tissues: ophthalmic surgery, soft tissue approximation, subcutaneous closures, ligatures and general and specialised surgery. Each box contains 12 sterile units in individual sachets, ready for the operating theatre. Factory reference P111VSP840-2 · SKU Cahnos AMP111VSP840-2 · Minimum order 10 boxes.
2. Technical Explanation
The thread is made of 100% polyglycolic acid (PGA), a synthetic polymer absorbable by hydrolysis, structured as a braided multifilament coated with polycaprolactone and calcium stearate. This coating provides a uniform, non-grooved surface, smooth passage through tissue and excellent knot security, with minimal tissue reaction. Absorption occurs by physiological hydrolysis over a predictable period of 60 to 90 days, maintaining tensile strength during the critical healing phase: 75% at 2 weeks, 50% at 3 weeks and 40% at 4 weeks. The Micropoint spatulated needle has a flattened body that separates tissue fibres rather than cutting them, minimising trauma and the risk of leakage in delicate lamellar structures such as the cornea, conjunctiva and sclera; its double-needle format (one at each end of the thread) allows continuous sutures without rethreading and a balanced distribution of tension. The product is sterile by ethylene oxide, single-use and non-resterilizable.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Material | 100% polyglycolic acid (PGA) multifilament, coated with polycaprolactone and calcium stearate |
| Structure | Braided multifilament, non-grooved surface |
| Gauge | USP 4/0 (metric EP 1.5) |
| Needle | Double needle ¼ circle spatulated Micropoint (2xVSP), non-grooved |
| Needle size | 8 mm |
| Thread length | 45 cm |
| Colour | Violet |
| Absorption | By hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (EO) |
| Presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 10 boxes |
| Quality | Prime |
| Factory reference | P111VSP840-2 |
| SKU Cahnos | AMP111VSP840-2 |
| Indications | Tissue approximation, ligatures, general surgery, ophthalmology and microsurgery |
Tensile strength retention after implantation:
| Period | Percentage of retained strength |
|---|---|
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 40 % |
Absorption profile:
| Mechanism | Value |
|---|---|
| Absorption mechanism | Hydrolysis |
| Practically complete absorption | 60-90 days |
| Tissue reaction | Minimal |
| Use | Single-use · Do not re-sterilize |
4. Equivalences and Cross-References
Polyglycolic acid (PGA) is a multifilament absorbable material widely established in the market. The reference indicated below is comparable at category level (same needle configuration, gauge and length), but should not be considered interchangeable 1:1: the polymer composition and absorption profile differ. Always check the manufacturer’s official technical data sheet before substituting.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (J&J MedTech) | Coated VICRYL (Polyglactin 910), double needle S-14 spatulated 8 mm ¼ circle, USP 4/0, violet, 45 cm | J572G | Comparable by category (not interchangeable 1:1) | PGLA material (copolymer 90% glycolide / 10% L-lactide) versus pure PGA; absorption 56-70 days versus 60-90 days; strength retention at 4 weeks of 30% versus 40%. Same needle dimensions (8 mm, spatulated, ¼ circle, double), gauge 4/0 and length 45 cm. |
Other homologous PGA absorbable families present in the sector (without validated code cross-reference in this data sheet, pending confirmation with the manufacturer): Safil from B. Braun (braided PGA), Surgicryl PGA from SMI, Assut Sutures, Peters Surgical and Atramat, all with ophthalmic spatulated needle ranges that must be verified code by code in their official catalogue.
There is no equivalence table validated by the manufacturer for this exact reference (USP 4/0, violet, double spatulated needle 8 mm, 45 cm); the SKU AMP111VSP840-2 is Cahnos’s own code and the factory reference P111VSP840-2 follows the range’s coding pattern. Consult the technical department before using in tenders or direct substitution of another brand.
5. Selection tips and variants of interest
The ¼ circle spatulated needle is the standard configuration in anterior ophthalmology and in the approximation of fine lamellar tissues (cornea, conjunctiva, eyelids), due to its atraumatic passage and control of stitch depth. For finer microsurgery, the same format is available in smaller gauges: AMP111VSP850-2 (USP 5/0) and AMP111VSP860-2 (USP 6/0), both with an 8 mm needle; and AMP111VSP6,560-2 (USP 6/0 with a 6.5 mm needle). If another curvature is required, the PGA Prime range includes the version AMP111HSP860-2 (½ circle spatulated 8 mm, USP 6/0). As an alternative of the same gauge and needle in polyglactin 910, see reference P1L11VSP840-2; for surgeries requiring non-absorbable suture, the polyester version AMP871VSP840-2. Single-use product, sterile and non-resterilizable; store in a cool, dry place, protected from light. For large-volume orders or public procurement, contact our technical-commercial department.





Reviews