Description
1. Commercial Introduction
We offer operating theatres, general and digestive surgery services, gynaecology, urology, plastic surgery units and purchasing centres the absorbable monofilament suture of poliglecaprone 25 (PGCL / caprolactone) from the PREMISUT range, factory reference 322DS2440, marketed under SKU AM322DS2440.
This presentation combines a ⅜ circle needle with a 24 mm reverse cutting triangular point (profile DS24), USP gauge 4/0 and undyed 75 cm thread: a fine format, highly versatile with exceptional handling, secure knot, minimal tissue reaction and full batch traceability. Each unit is supplied sterile, ready for immediate use, in a box of 12 units in individual sachets.
2. Technical explanation of the material and procedure
Poliglecaprone 25 (PGCL) is a synthetic absorbable copolymer of glycolic acid (75%) and ε-caprolactone (25%) in an uncoated monofilament structure. Its single-strand design reduces friction when passing through tissue (less sawing effect), minimises capillarity and hinders bacterial adhesion compared to braided multifilament sutures, while maintaining high tensile strength and reliable knot security during the critical phases of healing.
Absorption occurs by controlled hydrolysis —a predictable mechanism independent of enzymatic activity— over a period of 90 to 120 days (medium absorption), with an approximate retention of original strength of 60% at 7 days and 30% at 14 days after implantation, comfortably covering the critical healing phase of most soft tissues. It is not recommended when prolonged tissue support is required (e.g., fascia).
The DS24 needle, 24 mm, with ⅜ circle curvature and reverse cutting triangular point, concentrates the cutting edge on the outer curve: it penetrates firm, non-distensible tissues (skin, aponeurosis, dense subcutaneous tissue) with minimal resistance and better aesthetic outcome, preventing the cut from widening in the puncture bed. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference 322DS2440
| Parameter | Specification |
| Factory reference | 322DS2440 (PREMISUT POLYGLECAPRONE 25 CAPROLACTONE PGCL) |
| Reference Cahnos (SKU) | AM322DS2440 |
| Brand / Quality | PREMISUT (standard quality; Prime version available with grooved needle, ref. P322DS2440 / SKU AMP322DS2440) |
| Thread material | Poliglecaprone 25 (PGCL), copolymer 75 % glycolic acid / 25 % ε-caprolactone |
| Thread structure | Synthetic absorbable monofilament, grooved surface |
| Coating | Uncoated |
| Gauge / size | USP 4/0 (EP metric 1.5), in accordance with USP Pharmacopoeia and European Pharmacopoeia |
| Approximate thread diameter | 0.15 – 0.199 mm |
| Thread length | 75 cm (1 strand / 1 needle per unit) |
| Thread colour | Colourless (grooved) |
| Needle shape | 3/8 circle (⅜ circular), body and point |
| Tip type | Triangular reverse cutting (Reverse Cutting) |
| Needle size | DS24 (24 mm) |
| Swaged | 1 needle / 1 thread of 75 cm per unit |
| Absorption mechanism | Hydrolysis |
| Absorption period | 90 – 120 days (average absorption) |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Soft tissue approximation and general ligatures |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforms to USP and European Pharmacopoeia (EP) gauges |
| Storage | Below 30 °C, away from moisture and direct sunlight |
3.2. Tensile strength retention profile
| Period after implantation | Tensile strength retention |
| 1 week | 60% of initial tensile strength |
| 2 weeks | 30% of initial tensile strength |
| Complete absorption | 90–120 days, by hydrolysis |
3.3. Gauge equivalence according to USP / EP Pharmacopoeia (absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 5/0 | 1 | 0.10 – 0.149 mm |
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
4. Equivalences and Cross-References
The reference 322DS2440 corresponds to the official nomenclature of the PREMISUT range (profile DS24: ⅜ circle needle with 24 mm reverse cutting triangular point, USP gauge 4/0, undyed fluted 75 cm thread, single-strand armed) and Cahnos markets it without altering characters: no digit may be added, removed or reordered when ordering or tendering.
Poliglecaprone 25 is one of the most standardised medium-absorption synthetic absorbable monofilaments on the international market, with the MONOCRYL® line from Ethicon (Johnson \u0026 Johnson MedTech) being the clinical reference in the category, along with homologous families such as MONOSYN® from B. Braun (based on glyconate, with different degradation kinetics) and the PGCL lines from DemeTECH (DemeCAPRONE), SMI (Caprofyl®), Healthium (Monoglyde®), Assut Medical, Dolphin Sutures (Petcryl Mono), WEGO and CP Medical. The following correspondences have been verified against the manufacturer’s official catalogue (EPC Ethicon) and authorised distribution, both comparable by category and not interchangeable 1:1:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | MONOCRYL® (poliglecaprone 25) undyed monofilament, USP 4/0, 70 cm, PS-1 Prime needle, 24 mm, ⅜ circle, reverse cutting | Y935H | Comparable by category (not interchangeable 1:1) | Confirmed in official Ethicon catalogue and authorised distribution: they match in polymer (poliglecaprone 25), monofilament structure, undyed colour, USP gauge 4/0, ⅜ circle geometry, needle size 24 mm and reverse cutting point family. They differ in thread length (70 cm vs. 75 cm). Validate in technical data sheet/IFU before substituting in tendering. |
| Ethicon (J&J MedTech) | MONOCRYL® Plus Antibacterial (poliglecaprone 25) undyed monofilament, USP 4/0, 70 cm, PS-1 Prime needle, 24 mm, ⅜ circle, reverse cutting | MCP935H | Comparable by category (not interchangeable 1:1) | Confirmed in official Ethicon catalogue and authorised distribution: same configuration 4/0 · 24 mm · 70 cm undyed with antibacterial coating (triclosan), absent in this reference. The antibacterial load represents a relevant functional difference. Validate IFU before substituting. |
There is no equivalence table validated by the manufacturer that cross-references this specific configuration (PGCL monofilament, DS24 needle, 24 mm reverse cutting, USP gauge 4/0, undyed 75 cm thread, ref. 322DS2440) with references from other houses in the exact configuration; consult the technical department before use in tendering. The homologous families mentioned must be reviewed property by property (material, gauge, curvature, point family, length and number of strands, colour, coating and current CE/MDR regulatory classification) before any substitution. MONOCRYL® codes with 19 mm (PS-2) or 26 mm needles are outside the ±1 mm tolerance relative to the DS24 24 mm needle and must not be used as a direct replacement. The factory reference of this article does not appear in public regulatory databases (GUDID/EUDAMED) and must be confirmed with the supplier before use in specifications. The brands cited belong to their respective manufacturers and are mentioned solely as a comparative reference.
5. Selection tips and variants of interest
Tissue to suture: USP gauge 4/0 (metric EP 1.5) with a 24 mm needle is a fine, high-precision thread, indicated for approximation of delicate soft tissues, fine skin, mucosa, subcutaneous tissue and ligation of small-calibre vessels. The reverse cutting point penetrates dense tissues without tearing the puncture bed and improves the aesthetic outcome.
Quality and presentation: this reference is single-strand (1×75 cm) in standard quality. If you prefer the premium range with a fluted needle, the Prime version of the same profile DS24 is available (AMP322DS2440, ref. P322DS2440); if the procedure requires continuous or multi-point sutures without re-arming, the double-strand variant (2×75 cm) reduces the opening of sachets per procedure.
Alternative gauges and needles in the same PGCL range: same DS24 profile in USP 2/0 (AM322DS2420), USP 3/0 (AM322DS2430) and USP 5/0 (AM322DS2450); DS26 profile in USP 4/0 (AM322DS2640), USP 3/0 (AM322DS2630) and USP 2/0 (AM322DS2620); shorter DS19 needle in USP 4/0 (AM322DS1940). All references are supplied in boxes of 12 units.
Absorption profile: if the procedure requires longer wound support, consider polydioxanone PDO (180–210 days); if a shorter absorption profile with braided suture is required, consider polyglactin 910 (56–72 days), PGA (60–90 days) or fast-absorbing PGAR (40–45 days).
Sterility verification: product sterilised by ethylene oxide and single-use; check the blister integrity and expiry date before each procedure, keep the reference 322DS2440 and batch number for traceability in the operating theatre, and store below 30 °C, away from moisture and direct sunlight. For special B2B volume conditions, hospital supply agreements or validation of equivalents from other manufacturers, contact the technical-commercial team at Cahnos.





Reviews