Description
1. Commercial Introduction
The Absorbable Multifilament Suture PGAR RAPID from the PREMISUT RAPID range is the accelerated absorption reference for B2B supply to operating theatres, plastic and cosmetic surgery clinics, dermatology services and purchasing centres that need a synthetic absorbable thread with predictable handling, high knot security and complete disappearance within a few weeks, without the need for removal.
This configuration combines a braided strand of polyglycolic acid in USP 3/0 gauge, colourless and 75 cm long, with a ⅜ circle needle with triangular point in conventional precision cutting with square body of 16 mm (XNS16 profile): the ideal combination for tissue approximation, ligatures and precision work on skin, mucous membranes and subcutaneous tissue. Sterile single-use product, presented in a box of 12 units in individual sachets with batch traceability.
2. Technical Explanation
The PGAR suture (low molecular weight polyglycolic acid) is a synthetic polymer 100 % absorbable by hydrolysis, presented in a braided multifilament structure without grooves and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread sliding and facilitates firm and secure knotting, with excellent first knot retention.
Its rapid absorption profile is designed for short-duration healing: it retains 50 % of its tensile strength at 5 days and loses practically all its strength at 14 days, completing total absorption between 40 and 45 days through physiological hydrolysis, with minimal tissue reaction and independent of the patient’s enzymatic activity.
The XNS16 needle features a ⅜ circle curvature with a triangular point of conventional precision cutting (extra cutting, long edge) and square body: the edge cuts firm tissues cleanly generating a controlled passage, while the square section of the body optimises grip in the needle holder and prevents rotation during the surgical manoeuvre. The colourless colour of the thread facilitates visualisation of the field and offers maximum aesthetic discretion.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
|---|---|
| Factory reference | P222XNS1630 (Prime range) |
| Reference Cahnos (SKU) | AMP222XNS1630 |
| Brand / Range | PREMISUT RAPID |
| Thread material | 100 % polyglycolic acid (PGAR), absorbable by hydrolysis |
| Thread structure | Braided multifilament, non-ribbed |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric EP 2), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Colourless (violet on demand) |
| Needle shape | ⅜ circle (3/8 circular) |
| Tip type | Triangular, conventional precision cutting (extra cutting) |
| Needle body | Square |
| Needle size | XNS 16 mm, 1 needle per strand |
| Swaged | 1 needle / 1 suture per unit |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general and cosmetic surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Absorption profile and tensile strength retention
| Parameter | Value |
|---|---|
| Absorption type | Hydrolysis |
| Absorption time | 40 – 45 days |
| Tensile strength at 5 days | 50% of initial strength |
| Tensile strength at 14 days | 0% (total loss of tensile support) |
| Absorption rate | Predictable and uniform |
| Tissue reaction | Minimal |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The diameter of the strand complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
3.4. Reference code decoding
The official reference P222XNS1630 breaks down as follows: P (Prime range) + 222 (75 cm thread) + XNS16 (⅜ circle needle with triangular point of conventional precision cutting with square body of 16 mm) + 30 (USP 3/0 gauge). Cahnos markets the official reference without altering characters: no digit should be added, removed or reordered when ordering or bidding.
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (PGAR) is one of the most widespread families of braided absorbable suture on the market, comparable at clinical category level with the rapid absorption lines of major manufacturers. The reference P222XNS1630 shares low molecular weight absorbable material, USP 3/0 gauge, ⅜ circle curvature, 16 mm needle and colourless thread with the homologous lines of the segment. However, the comparison is declared at category level, not interchangeable 1:1, because the needle profile and body differ (XNS conventional cutting with square body versus standard reverse cutting profiles), the coating and the commercial presentations.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE™ (braided polyglactin 910, colourless) | W9932 – USP 3/0, 75 cm thread, PS reverse cutting Prime needle of 26 mm, ⅜ circle, colourless, box of 12 | Comparable by category (not interchangeable 1:1) | Same short-duration absorbable material, gauge, length and colour; the needle differs (PS reverse cutting of 26 mm versus XNS16 conventional precision cutting with square body). Code confirmed in official Ethicon EMEA catalogue and authorised distribution. |
| B. Braun / Aesculap | Safil Quick™ (low molecular weight polyglycolic acid, braided and coated) | C1046421 – USP 3/0, 70 cm thread, 3/8 circle cutting needle of 19 mm, undyed | Comparable by category (not interchangeable 1:1) | Same base material (PGAR), rapid absorption profile and gauge; the needle size differs (19 mm versus 16 mm), the point family (standard cutting versus conventional precision cutting with square body) and the thread length (70 cm versus 75 cm). Code confirmed in authorised distribution and GUDID/NHS. |
In the same therapeutic block are also the rapid absorption lines of DemeTECH (DemeCRYL™ Fast) and SMI (Surgicryl Fast Absorb), comparable by category by material and absorption profile, although their specific codes for this needle configuration must be verified against the manufacturer’s catalogue before any substitution. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference. There is no equivalence table validated by the manufacturer for this reference; consult the technical department before use in tendering.
5. Selection tips and variants of interest
USP 3/0 gauge (metric 2) is indicated for tissues with moderate tension: skin, subcutaneous tissue, mucous membranes and soft tissue closures. The XNS16 needle with conventional precision cutting and square body is especially recommended when a clean and controlled passage is needed in firm tissues, with excellent stability in the needle holder and minimal rotation during suturing.
| Variant / Reference | Description |
|---|---|
| AMP222XNS1630 (P222XNS1630) | PGAR RAPID Prime XNS16 – USP 3/0 colourless – 75 cm (this product) |
| AMP228XNS1640 (P228XNS1640) | PGAR RAPID XNS16 – USP 4/0 colourless – 70 cm (same needle, finer gauge for delicate tissue) |
| AMP2L22XNS1630 (P2L22XNS1630) | PGLAR Polyglactin 910 RAPID Prime XNS16 – USP 3/0 colourless – 75 cm (same needle geometry, PGLAR polymer) |
| AMP222XDS1930 (P222XDS1930) | PGAR RAPID Prime XDS19 – USP 3/0 colourless – 75 cm (precision reverse cutting with 19 mm square body for skin and cosmetic surgery) |
| AMP222DS1630 (P222DS1630) | PGAR RAPID Prime DS16 – USP 3/0 colourless – 75 cm (classic reverse cutting of 16 mm, economical alternative) |
If greater visibility of the suture on the surgical field is required, request the same configuration in violet colour on demand. Sterile single-use product: check the integrity of the packaging and the expiry date before each procedure, retain the batch number for traceability and store below 30 °C, away from moisture and direct sunlight. For special B2B conditions, volume orders or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.