Description
1. Commercial Introduction
Absorbable multifilament suture PREMISUT RAPID Prime (polyglactic acid PGAR) is the accelerated absorption closure solution for general and ambulatory surgery operating theatres, emergency departments, digestive surgery, gynaecology and B2B purchasing centres that require a braided, flexible thread with early and fully predictable loss of tensile strength. This reference combines a 75 cm colourless strand in USP 2/0 gauge with a ⅜ circle needle with a triangular reverse cutting point of 26 mm (DS26 Prime profile): clean penetration in skin and firm tissues with excellent aesthetic results and minimal tissue reaction. Sterile product, ready to use, presented in a box of 12 units in individual sachets with batch traceability; intended for hospital replenishment, clinics, distribution and tenders.
2. Technical Explanation
Polyglactic acid (PGAR) is a 100% absorbable synthetic polymer, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate: this coating reduces friction when passing through tissues, improves thread sliding and facilitates firm, secure knotting, with minimal tissue reaction. In its RAPID variant (low molecular weight), absorption occurs by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days: it retains approximately 50% of the initial tensile strength at 5 days and loses it completely at 14 days, suitable for tissues that heal quickly and where it is desired to minimise residual foreign material.
The DS26 needle features a ⅜ circle curvature with a triangular reverse cutting point: the cutting edge is oriented towards the outer face of the curvature, so it cuts in the opposite direction to the wound bed, preventing the cut from widening and improving the aesthetic result of the closure. Its 26 mm length is suitable for medium-depth planes, skin and subcutaneous tissue. Gamma radiation sterilisation guarantees final sterility with minimal alteration of mechanical properties; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
|---|---|
| Factory reference | P222DS2620 (Prime range) |
| Reference Cahnos (SKU) | AMP222DS2620 |
| Brand / quality | PREMISUT RAPID – PGAR Prime |
| Thread material | 100 % polyglactic acid (PGAR), rapid absorption |
| Thread structure | Braided multifilament, smooth |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 2/0 (metric 3 / EP 3), USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.30 – 0.339 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Colourless (violet on demand) |
| Needle shape | ⅜ circle (3/8 circular, 135°) |
| Needle tip | Triangular reverse cutting (Reverse Cutting) |
| Needle size | DS 26 mm (Prime), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption | Hydrolysis; complete between 40 and 45 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial format | Box of 12 units (individual sterile sachets); minimum order 30 boxes |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Decoding of the official reference
The official reference P222DS2620 breaks down as follows: P (Prime range) + 222 (75 cm thread) + DS26 (⅜ circle needle with triangular reverse cutting point of 26 mm) + 20 (USP 2/0 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
3.3. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40 – 45 days, by hydrolysis (predictable and uniform rate) |
The retention profile corresponds to the rapid absorption formulation (RAPID line); it should not be confused with standard absorption polyglactic acid (60–90 days) or standard polyglactin 910 (56–72 days), also available in our catalogue.
3.4. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 0 | 3.5 | 0.35 – 0.399 mm |
| 2/0 (this product) | 3 | 0.30 – 0.339 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding their limits.
4. Equivalences and Cross-References
The reference P222DS2620 corresponds to the official nomenclature of the PREMISUT RAPID range (rapid absorption polyglactic acid, multifilament, USP 2/0, 75 cm colourless thread, ⅜ circle needle with triangular reverse cutting point of 26 mm). Rapid absorption polyglactic acid (PGAR) belongs to the worldwide family of short-duration braided multifilament absorbable sutures, whose international clinical benchmark is the VICRYL RAPIDE line (polyglactin 910). Only the reference whose code and complete attributes have been verified in the manufacturer’s official catalogue (Ethicon EMEA) and in authorised distribution is included:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | W9933 – USP 2/0, 75 cm, colourless, Prime reverse cutting needle 26 mm, ⅜ circle, box of 12 | Comparable by category (not interchangeable 1:1) | They match on: short-duration absorbable material (≈42 days), USP 2/0 gauge, colourless, 75 cm length, ⅜ circle geometry, 26 mm reverse cutting needle and 12-unit format. They differ in the base polymer (polyglactin 910 versus polyglactic acid) and the braid coating. Code confirmed in Ethicon’s EMEA catalogue and authorised distributors (medicalsutures, Mistry, Aston Pharma); verify absorption profile in the official datasheet before substituting in a tender. |
The remaining homologous lines in the rapid absorption segment —Safil Quick and NOVOSYN QUICK (B. Braun / Aesculap), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI)— are comparable by category in terms of material and absorption profile, but must be reviewed property by property (material, USP 2/0 gauge, multifilament structure, ⅜ circle curvature, reverse cutting point family, needle length 26 mm, strand length 75 cm, colourless and CE/MDR regulatory classification) before any substitution; their codes for this exact configuration are pending confirmation with each manufacturer and are not published as definitive data. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference. There is no equivalence table validated by the manufacturer; consult the technical department before use in a tender. Within the PREMISUT system itself, the PGAR RAPID family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are fully interchangeable with each other respecting gauge and needle geometry.
5. Selection tips and variants of interest
The USP 2/0 (metric 3) gauge offers an optimal balance between strength and fineness, indicated for tissues under moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels. The DS26 needle, ⅜ circle with reverse cutting point, is the recommended option when firm penetration is required while preserving the skin edge and achieving a better aesthetic result, particularly valued in general surgery and skin closures.
| SKU | Variant |
|---|---|
| AM222DS2620 (222DS2620) | PGAR RÁPIDO DS26 – USP 2/0 undyed – 75 cm (same configuration, standard quality; direct equivalent within the same manufacturer) |
| AMP222HRT2620 (P222HRT2620) | PGAR RÁPIDO Prime HRT26 trocar point – USP 2/0 undyed – 75 cm (same configuration with trocar point for dense tissues) |
| AM222HRT2620 (222HRT2620) | PGAR RÁPIDO HRT26 trocar point – USP 2/0 undyed – 75 cm (standard quality, trocar needle) |
| AMP222HRT2630 (P222HRT2630) | PGAR RÁPIDO Prime HRT26 – USP 3/0 undyed – 75 cm (finer gauge, trocar point) |
| AMP228HRT2610 (P228HRT2610) | PGAR RÁPIDO Prime HRT26 – USP 0 undyed – 75 cm (larger gauge for higher tension) |
| AMP222XDS1930 (P222XDS1930) | PGAR RÁPIDO Prime XDS19 precision reverse cutting with square body – USP 3/0 undyed – 75 cm (plastic/cosmetic profile) |
| AMP222XNS1630 (P222XNS1630) | PGAR RÁPIDO Prime XNS16 precision conventional cutting – USP 3/0 undyed – 75 cm (conventional cutting for firm tissues) |
If the protocol prefers even more atraumatic penetration in lax tissues, consider the conical point series (HR / taperpoint) from the same range; for deep planes with a wider field, consult the HRT36/HRT40 needles. The undyed version offers maximum aesthetic discretion on the surgical field; if greater visibility is required, request the same configuration in dyed thread (violet) on demand. Sterility and storage: product sterile by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each use, keep the reference P222DS2620 together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For special B2B conditions by volume or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.