Description
1. Commercial Introduction
The absorbable multifilament suture PREMISUT RAPID of polyglactic acid (PGAR) is the accelerated absorption surgical closure solution designed for general surgery operating theatres, ambulatory surgery, day hospitals and B2B purchasing centres that need a braided, flexible thread with an early and fully predictable loss of tensile strength. Its ½ circle needle with Tapercut trocar point of 26 mm penetrates medium-consistency tissues with precision and without a wide cutting effect, while the colourless 75 cm thread offers maximum discretion on the surgical field. A cost-effective presentation, in a box of 12 units, with a minimum order of 30 boxes.
2. Technical Explanation
The polyglactic acid (PGAR) is a synthetic absorbable polymer spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate, which provides flexibility, low shape memory, smooth passage through tissue and a firm, secure knot tie. In its RAPID variant (low molecular weight), absorption occurs by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days: it retains approximately 50% of the initial tensile strength at 5 days and loses it completely by 14 days. This short support window makes it ideal for rapidly healing tissues, minimising residual foreign material. The ½ circle needle with cylindrical body and trocar point (Tapercut) combines the strength of a round body with a sharp end that pierces somewhat firmer tissues than a conical point, without the cutting effect of a conventional cutting needle. Gamma radiation sterilisation ensures final sterility with minimal alteration of mechanical properties. Sterile single-use product, not re-sterilisable.
3. Technical Specifications
| Parameter | Specification |
| Material | 100% polyglactic acid (PGAR), synthetic and absorbable, rapid-absorption variant |
| Thread structure | Braided multifilament with polycaprolactone and calcium stearate coating |
| Gauge | USP 3/0 (metric 2) |
| Thread diameter | 0.20 – 0.249 mm (indicative, according to Pharmacopoeia) |
| Thread colour | Colourless |
| Thread length | 75 cm |
| Needle shape | ½ circle (1/2), cylindrical body |
| Needle tip | Trocar / Tapercut (triangular) |
| Needle size | 26 mm (HRT26) |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery |
| Manufacturer reference | 222HRT2630 |
| Commercial format | Box of 12 units (minimum order 30 boxes) |
Absorption and tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40–45 days, by hydrolysis (predictable and uniform rate) |
The retention profile mentioned corresponds to the rapid absorption formulation (RAPID line); it should not be confused with the standard absorption version of polyglactin 910 (56–72 days), which is not comparable in terms of tensile support window.
4. Equivalences and Cross-References
Rapid absorption polyglactic acid (low molecular weight) is one of the most widespread categories of absorbable multifilament suture in the surgical market. This reference shares rapid absorbable material, USP 3/0 gauge, ½ circle curvature, cylindrical needle and colourless 75 cm thread with the homologous line of the sector’s main manufacturer. Only the reference whose code and attributes have been verified in official catalogues and authorised distributors is included:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | W9932 – USP 3/0, 75 cm, colourless, PS needle | Comparable by category (not interchangeable 1:1) | They match in rapid absorption material (≈42 days), USP 3/0 gauge, colourless colour and 75 cm length. They differ in needle nomenclature and geometry (Ethicon’s PS versus this reference’s HRT26 Tapercut) and the braid coating. Code confirmed in Ethicon’s official catalogue and authorised distributors. Verify needle geometry in the official datasheet before substituting. |
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders. Homologous ranges in the rapid absorption segment include Safil Quick and NOVOSYN QUICK (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category, material and absorption profile, although for the exact combination USP 3/0 / 26 mm trocar point needle / 75 cm thread, their specific codes must be confirmed with the manufacturer. Standard VICRYL (56–72 day absorption) is not comparable to this RAPID version.
5. Selection tips and variants of interest
The USP 3/0 gauge (metric 2) offers an optimal balance between strength and fineness, indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels. The 26 mm needle with Tapercut trocar point is the recommended option when firm penetration without a wide cutting effect is needed in medium-consistency tissues. Within the same HRT26 configuration in superior quality, Cahnos offers the PRIME version with reference AMP222HRT2630. If the protocol prefers a larger radius needle, consider the HRT36 family (ref. AM222HRT3610) or HRT40; for finer tissues, the PGAR RAPID line offers ⅜ circle trocar point needles DRT16 (AMP222DRT1630) and ⅜ circle reverse cutting DS16 (AM222DS1630), as well as the ½ circle configuration with reverse cutting triangular point HS22 in 45 cm thread (AM221HS2230). In the PGLAR family of polyglactin 910 RAPID, the same HRT26 trocar needle exists in reference AM2L22HRT2630. The colourless version offers maximum discretion in the field. As this is a sterile single-use product, check the integrity of the packaging and the expiry date before each procedure and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or equivalences from other commercial houses, contact the technical-commercial team at Cahnos.





Reviews