Description
1. Commercial Introduction
PREMISUT RAPID is the accelerated absorption line from Cahnos for the professional who needs a resorbable thread with predictable handling, secure knotting and complete disappearance from the body within a few weeks, without the need to remove stitches. This Prime quality configuration combines a braided multifilament strand of polyglycolic acid (PGAR), colourless and 75 cm long, with a ½ circle needle with cylindrical body and Taperpoint conical point of 17 mm (HR17 profile): the ideal choice for soft tissue approximation, ligatures and general surgery, with excellent value for money for hospital replenishment, purchasing centres and tenders. Sterile product, presented in a box of 12 units in individual sachets with batch traceability.
2. Technical Explanation
The PGAR suture (rapid-profile, low-molecular-weight polyglycolic acid) is a synthetic polymer that is 100 % absorbable by hydrolysis, braided into a multifilament and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves sliding and facilitates firm, secure knotting, with minimal tissue reaction and excellent initial tensile strength.
Its degradation profile is designed for short-term healing: it retains 50 % of tensile strength at 5 days and loses virtually all its resistance at 14 days, completing total absorption by hydrolysis between 40 and 45 days. The HR17 needle features a ½ circle curvature (160°) with a cylindrical body and Taperpoint conical point: the round body passes through soft tissues atraumatically and the sharp conical point penetrates without aggressive cutting effect, avoiding enlargement of the exit hole. Its 17 mm length makes it particularly suitable for reduced surgical fields, skin, subcutaneous tissue, mucous membranes and paediatric surgery. The USP 5/0 (metric 1) gauge provides the fineness needed for very delicate tissues under low tension. Sterilisation by radiation gamma; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics (Ref. P222HR1750 · SKU AMP222HR1750)
| Parameter | Specification |
| Factory reference | P222HR1750 (Prime range) |
| Reference Cahnos (SKU) | AMP222HR1750 (box of 12 units) |
| Brand / quality | PREMISUT RAPID (Prime range) |
| Thread material | 100% polyglycolic acid (PGAR), synthetic absorbable, low molecular weight variant (RAPID) |
| Thread structure | Coated braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 5/0 (metric 1 / EP 1), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.10 – 0.149 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Colourless (undyed) |
| Needle shape | ½ circle (1/2 circular, 160°), cylindrical body |
| Tip type | Taperpoint conical (atraumatic) |
| Needle size | HR 17 Prime (17 mm), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Tensile strength retention | 50 % at 5 days / 0 % at 14 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Absorption profile and tensile strength retention
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Complete absorption | 40 – 45 days by hydrolysis (predictable and uniform rate) |
The cited retention profile corresponds to the rapid absorption formulation (RAPID line); it must not be confused with standard absorption polyglycolic acid (PGA line, 60–90 days) or with standard polyglactin 910 (56–72 days).
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 6/0 | 0.7 | 0.07 – 0.099 mm |
| 5/0 (this product) | 1 | 0.10 – 0.149 mm |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
3.4. Decoding of the official reference
Official reference P222HR1750: P (Prime range) + 222 (75 cm thread) + HR17 (½ circle needle, cylindrical body, Taperpoint conical point of 17 mm) + 50 (USP 5/0 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (PGAR) belongs to the worldwide family of short-term braided multifilament absorbable sutures, whose international clinical benchmark is the VICRYL RAPIDE line (polyglactin 910), with which it shares the hydrolysis absorption mechanism and accelerated resistance loss profile (≈50 % at 5 days and total loss towards 14 days).
For this specific configuration (USP 5/0, ½ circle cylindrical needle with conical point of 17 mm, colourless 75 cm thread), no cross-reference code from another manufacturer whose attributes match one-to-one has been found in official sources: the HR17 Taperpoint needle geometry in gauge 5/0 is not reproduced identically by international houses. As homologous ranges in the rapid absorption segment, the following are listed: VICRYL RAPIDE (Ethicon / Johnson \u0026 Johnson MedTech), Safil Quick and Novosyn Quick (B. Braun / Aesculap), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category, material and absorption profile; their specific codes for the combination USP 5/0 / HR17 Taperpoint needle / 75 cm must be confirmed with each manufacturer, verifying property by property: material, USP 5/0 gauge (metric 1), multifilament structure, ½ circle curvature, conical point family, needle length 17 mm, 75 cm thread, colourless colour, 1 needle attachment and CE/MDR regulatory classification. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
There is no manufacturer-validated equivalence table cross-referencing this reference with other brands; consult the technical department of Cahnos before use in tenders. Within the PREMISUT system itself, the direct equivalent with identical needle geometry, gauge and length is the standard version (Ref. 222HR1750, SKU AM222HR1750), and the PGAR RAPID family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are fully interchangeable with each other, respecting gauge and needle geometry.
5. Selection tips and variants of interest
The USP 5/0 (metric 1) gauge is a fine size indicated for low-tension tissues and rapid healing: fine skin, mucous membranes, subcutaneous tissue, superficial planes and ligatures of small-calibre vessels in general surgery and minor surgery. The HR17 ½ circle needle with Taperpoint conical point combines the sharp entry of the point with the non-traumatic passage of the cylindrical body: it is the recommended option for reduced surgical fields and soft tissues where an atraumatic passage is sought, without cutting effect and with minimal bleeding. The colourless colour offers maximum discretion in the surgical bed.
| Variant (SKU) | Description |
| AM222HR1750 (222HR1750) | PGAR RAPID standard – HR17 taperpoint – USP 5/0 undyed – 75 cm (same geometry and gauge, standard range; direct equivalent within the same manufacturer) |
| AMP222HR1740 (P222HR1740) | PGAR RAPID Prime – HR17 taperpoint – USP 4/0 undyed – 75 cm (same needle, next gauge up) |
| AMP222HR1730 (P222HR1730) | PGAR RAPID Prime – HR17 taperpoint – USP 3/0 undyed – 75 cm (same needle, gauge for higher tension) |
| AMP222DS1250 (P222DS1250) | PGAR RAPID Prime – DS12 reverse cutting ⅜ circle – USP 5/0 undyed – 75 cm (same strand, triangular point for firmer tissues) |
| AMP222HR2240 (P222HR2240) | PGAR RAPID Prime – HR22 conical point – USP 4/0 undyed – 75 cm (22 mm needle for somewhat wider fields) |
If the protocol requires penetration into firm or poorly distensible tissues, consider the reverse cutting point (DS) series from the same range; for planes under higher tension, use USP 4/0 or 3/0 gauge with the same HR17 needle. Sterility and storage: sterile product by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each use and keep reference P222HR1750 together with the batch number for traceability in the operating theatre. Store below 30 °C, away from moisture and direct sunlight. For B2B volume conditions or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews