Description
1. Commercial Introduction
The absorbable multifilament suture PGAR RAPID in Prime quality (PREMISUT RAPID range) is the reference for accelerated absorption surgical closure for hospitals, general surgery operating theatres, emergency services and B2B purchasing centres that need a braided, flexible thread with early and fully predictable strength loss. Its ½ circle needle with 17 mm reverse cutting triangular point (HS 17 profile) ensures clean, controlled penetration in fine and firm tissues, while the 45 cm undyed thread offers maximum discretion on the surgical field. A practical and cost-effective presentation, in a box of 12 units, designed for fast-healing tissues.
2. Technical Explanation
Rapid polyglycolic acid (PGAR) is a synthetic absorbable polymer spun into a braided multifilament structure, coated with polycaprolactone and calcium stearate to promote thread glide, smooth handling and a firm, secure knot. In the RAPID variant, the thread is formulated to lose tensile strength in an accelerated and programmed manner: it retains 50 % of initial strength at 5 days and 0 % at 14 days, with complete absorption by hydrolysis between 40 and 45 days. This profile makes it especially suitable for superficial tissues with rapid healing (skin closure, mucous membranes, subcutaneous tissue) and ligation of small-calibre vessels. The reverse cutting needle creates a clean entry hole by directing the cut away from the wound, preserving the skin edge and preventing the cut from widening at the puncture site. Sterile single-use product, not for re-sterilisation.
3. Technical Specifications
3.1. General characteristics (Ref. AMP221HS1730)
| Parameter | Specification |
| Material | 100 % rapid polyglycolic acid (PGAR), coated braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Quality | Prime (PREMISUT RAPID range) |
| Gauge | USP 3/0 (metric 2; diameter approx. 0.20 – 0.249 mm according to Pharmacopoeia) |
| Thread colour | Colourless |
| Thread length | 45 cm |
| Needle shape | Half circle (1/2, 160°) |
| Needle tip | Triangular, reverse cutting |
| Needle size | HS 17 mm |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial format | Box of 12 units (minimum order 30 boxes) |
| SKU Cahnos | AMP221HS1730 |
3.2. Absorption profile and tensile strength retention
| Post-implantation time | Retained tensile strength |
| 5 days | 50% of initial strength |
| 14 days | 0% (total loss of tensile support) |
| 40 – 45 days | Complete absorption by hydrolysis (predictable and uniform rate) |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 3/0 | 2 | 0.20 – 0.249 mm (indicative, according to Pharmacopoeia) |
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (low molecular weight) is the most widespread multifilament absorbable category in the surgical market. This reference shares synthetic absorbable material with rapid action, USP 3/0 gauge, 45 cm undyed thread and early strength loss profile (50 % at 5 days / 0 % at 14 days) with the homologous lines of the main manufacturers. Only codes whose attributes have been verified in official catalogues or authorised distributors are included:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (braided polyglactin 910, colourless) | MPVR4970H – USP 3/0, PS-2 needle 19 mm, ⅜ circle, reverse cutting, 45 cm | Comparable by category (not interchangeable 1:1) | Same rapid absorption profile (≈42 days), USP 3/0 gauge, undyed colour and length of 45 cm; the needle curvature (⅜ vs ½ circle) and size (19 mm vs HS 17 mm) differ. Code confirmed in catalogues of authorised distributors. Verify needle geometry on the official data sheet before replacing. |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (braided polyglactin 910, colourless) | W9931 – USP 3/0, 75 cm, colourless, FS-2 needle 19 mm, ⅜ circle, reverse cutting | Comparable by category (not interchangeable 1:1) | Same rapid absorption profile (≈42 days), USP 3/0 gauge and undyed colour; the thread length (75 cm vs 45 cm), curvature (⅜ vs ½ circle) and needle size (19 mm vs HS 17 mm) differ. Code confirmed in official Ethicon EMEA catalogue and authorised distributors. |
There is no equivalence table validated by the manufacturer; consult with the technical department before using in a tender. Homologous ranges in the rapid absorption segment include Safil Quick / Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category, material and absorption profile, although for the exact combination USP 3/0 / ½ circle needle HS 17 mm / 45 cm thread, their specific codes must be confirmed with each manufacturer. Standard VICRYL (absorption 56 – 72 days) is not comparable to this RAPID version.
5. Selection tips and variants of interest
USP 3/0 gauge (metric 2) is a common choice in fine and medium tissues where an optimal balance between strength and fineness is required: skin approximation, subcutaneous tissue, mucous membranes and ligation of small-calibre vessels. The 17 mm ½ circle reverse cutting needle is especially recommended when a firm and precise passage through skin or mucous membranes is needed without tearing the edge. Within the same PGAR RAPID Prime range, Cahnos offers complementary configurations: the same needle and gauge with 75 cm thread (SKU AMP222HS1730), the version in USP 4/0 gauge with 45 cm thread (SKU AMP221HS1740), the variant with 22 mm reverse cutting triangular needle (SKU AMP221HS2230), the same configuration in standard quality (SKU AM221HS1730) and the PGLAR Polyglactin 910 RAPID line with identical configuration (SKU AMP2L21HS1730) as a material alternative within the same absorption profile. The undyed version offers maximum discretion in the field; if greater visibility is required, request the same configuration in dyed thread on demand. Sterile single-use product: check the integrity of the packaging and the expiry date before each procedure. For special B2B conditions or volume orders (minimum 30 boxes), contact the technical-commercial team of Cahnos.