Description
1. Commercial Introduction
The Absorbable Braided Multifilament PGA Prime Suture reference AMP123HRT3601 is the surgical closure solution your centre needs for general surgery procedures that require controlled reabsorption, excellent handling, and impeccable knot security. It combines a ½ circle cylindrical body needle with a trocar point (tapercut) of 36 mm with a braided polyglycolic acid (PGA) thread in USP 1 gauge, colourless fluted, and 90 cm length.
Presented in B2B format in a box of 12 units with a minimum order of 30 boxes, it offers an outstanding value for money for operating theatres, clinics, distributors, purchasing centres, and surgical material tenders.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100 % absorbable synthetic polymer that is presented in a braided multifilament structure with a fluted finish, coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread glide, and facilitates firm and secure knotting, with minimal tissue reaction.
Absorption occurs by hydrolysis, a predictable process independent of the patient’s enzymatic activity, completed between 60 and 90 days. Tensile strength retention maintains the necessary support during the first 2 to 4 weeks of healing (75 % at 2 weeks, 50 % at 3, and 40 % at 4), while the tissue recovers its own strength.
The HRT36 needle features a ½ circle curvature with a cylindrical body and trocar point (tapercut): a converging triangular section point that combines firm penetration in moderate-high tension tissues with contained trauma. Its 36 mm length makes it suitable for planes of some depth. The colourless colour of the thread avoids visual interference and is ideal in fields where a discreet suture is required. Product sterile by ethylene oxide, single-use and non-resterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference AMP123HRT3601
| Characteristic | Specification |
|---|---|
| Reference (SKU) | AMP123HRT3601 (manufacturer ref. P123HRT3601) |
| Range / Quality | Premisut – Prime |
| Material | 100 % Polyglycolic acid (PGA), absorbable homopolymer |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 1 (metric 4), conforming to USP/EP Pharmacopoeia |
| Thread colour | Colourless fluted |
| Thread length | 90 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Tapercut (converging triangular cross-section) |
| Needle size | HRT 36 mm, 1 needle per thread |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 1 | 4 | 0.40 – 0.499 mm (indicative according to Pharmacopoeia) |
3.4. Reference code decoding
Official reference P123HRT3601: 123 (90 cm thread) + HRT36 (½ circle cylindrical body needle with 36 mm trocar point) + 01 (USP 1 gauge). Cahnos markets the official reference without altering characters: no digit should be added, removed, or reordered when ordering or tendering.
4. Equivalences and Cross-References
Polyglycolic acid (PGA) with 60-90 days absorption is one of the most proven raw materials for synthetic absorbable multifilament suture on the market. There is no equivalence table validated by the manufacturer that cross-references this reference with other commercial houses in the same needle configuration (HRT 36 mm, USP 1 gauge, colourless 90 cm thread); therefore, no unverified cross-reference codes are published.
At the level of material and clinical category, this reference is comparable, without being 1:1 interchangeable, with international lines of 100 % braided coated PGA —Safil (B. Braun), Dexon II (Medtronic/Covidien), DemeSORB PGA (DemeTECH), Surgicryl PGA (SMI), Petcryl (Dolphin Sutures)— and, with differences in polymer and degradation profile, with polyglactin 910 (PGLA) lines: Vicryl / Vicryl Plus (Ethicon), Polysorb (Medtronic) and Novosyn (B. Braun). Before any substitution, it is mandatory to verify in the official data sheet of each house: exact polymer composition, USP 1 gauge, needle geometry (½ circle, cylindrical body, trocar point, 36 mm, 1 needle per thread), 90 cm length, and absorption profile.
There is no equivalence table validated by the manufacturer; consult with the technical department before using in tenders. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP 1 gauge (metric 4) is indicated for tissues that support moderate-high tension: fascia, aponeurosis, muscle planes, abdominal wall, and ligatures of medium-calibre vessels. The ½ circle trocar point (tapercut) HRT needle is the appropriate choice for firm tissues where clean penetration without tearing is required; if you need a fully atraumatic point for soft tissues, consider the conical point (HR) versions of the same PGA family.
| Reference | Variant |
|---|---|
| AMP123HRT3601 | PGA Prime USP 1, HRT36 trocar point needle, colourless, 90 cm (current product) |
| AMP113HRT3601 | PGA Prime USP 1, HRT36 trocar point needle, violet, 90 cm (same configuration in violet) |
| AMP113HRT3610 | PGA Prime USP 0, HRT36 trocar point needle, violet, 90 cm (smaller gauge) |
| AMP113HRT3620 | PGA Prime USP 2/0, HRT36 trocar point needle, violet, 90 cm (smaller gauge) |
| AMP113HRT3630 | PGA Prime USP 3/0, HRT36 trocar point needle, violet, 90 cm (fine gauge) |
| P1L13HRT3601 | PGLA Prime USP 1, HRT36 trocar point needle, violet, 90 cm (absorption 56-72 days) |
| P1L23HRT3601 | PGLA Prime USP 1, HRT36 trocar point needle, colourless, 90 cm (colourless PGLA version) |
All references are supplied in a box of 12 units; the minimum order is 30 boxes to guarantee continuous supply for your healthcare centre. If you need longer strength retention (180-210 days), consider the polydioxanone (PDO) family. As it is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, and keep the reference together with the batch number for traceability in the operating theatre. For B2B volume conditions or equivalences from other commercial houses, contact the technical-commercial team at Cahnos.