Description
1. Commercial Introduction
The Absorbable Multifilament PGA Prime Suture, reference AMP113HRT3610, is the surgical closure solution your facility needs for general surgery, gynaecology and urology procedures that require controlled reabsorption, excellent handling and impeccable knot security. It combines a ½ circle needle with a cylindrical body and a trocar (tapercut) point of 36 mm with a braided thread of 100% polyglycolic acid (PGA) in USP 0 calibre, violet fluted colour and 90 cm length.
Presented in B2B format in a box of 12 units with a minimum order of 30 boxes, it offers an outstanding price-quality ratio for operating theatres, clinics, distributors, purchasing centres and tenders for surgical material.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer that is presented in a braided multifilament structure with a fluted finish, coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread sliding and facilitates firm, secure knotting, with minimal tissue reaction.
Absorption occurs by hydrolysis, a predictable process independent of the patient’s enzymatic activity, completed between 60 and 90 days. Tensile strength retention maintains the necessary support during the first 2 to 4 weeks of healing (75% at 2 weeks, 50% at 3 and 40% at 4), while the tissue recovers its own resistance.
The HRT 36 needle features a ½ circle curvature with a cylindrical body and trocar (tapercut) point: a convergent triangular-section point that combines firm penetration in tissues of moderate tension or slightly fibrous tissues with contained trauma. Its 36 mm length makes it suitable for planes of certain depth. The violet colour of the thread guarantees optimal visibility on the surgical field. Product sterilised by ethylene oxide, single-use and non-reesterilizable.
3. Technical Specifications
3.1. Technical data sheet for reference AMP113HRT3610
| Characteristic | Specification |
|---|---|
| Reference (SKU) | AMP113HRT3610 |
| Range / Quality | Prime |
| Material | 100 % Polyglycolic acid (PGA), absorbable homopolymer |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 0 (metric 3.5), conforming to USP/EP Pharmacopoeia |
| Thread colour | Violet grooved |
| Thread length | 90 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Tapercut (converging triangular cross-section) |
| Needle size | HRT 36 mm, 1 needle per thread |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 0 | 3.5 | 0.35 – 0.399 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Polyglycolic acid (PGA) with 60-90 days of absorption is one of the most well-established synthetic absorbable multifilament suture raw materials on the market. There is no official equivalence table validated by the manufacturer for a 1:1 cross-reference with the exact configuration of this reference (HRT 36 mm needle, USP 0 calibre, 90 cm violet thread); therefore, the codes cited below correspond to catalogue comparisons verified against authorised distribution and are not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | Coated VICRYL® – braided coated polyglactin 910, violet, USP 0, 90 cm, CTD-1 taper cut needle 36 mm, ½ circle | JB9994G | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: USP 0 calibre, violet colour, 90 cm length, ½ circle shape, 36 mm needle size and tapercut point family match; the polymer differs (polyglactin 910 vs. 100% PGA) and the absorption interval (56-72 days vs. 60-90 days). |
| Ethicon (Johnson & Johnson MedTech) | VICRYL® PLUS Antibacterial – braided polyglactin 910 with triclosan, violet, USP 0, 90 cm, 36 mm tapercut needle, ½ circle | W9467 | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: USP 0 calibre, colour, 90 cm length, ½ circle shape, 36 mm needle size and tapercut point match; the polymer (polyglactin 910 with triclosan vs. 100% PGA) and the presentation (box of 36 units vs. 12 units) differ. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. As homologous market families to verify reference by reference in their official data sheet before substitution, the following lines are cited: 100% coated braided PGA —Safil (B. Braun), Dexon II (Medtronic/Covidien), Surgicryl PGA (SMI), DemeSORB PGA (DemeTECH)— and polyglactin 910 —Vicryl / Vicryl Plus (Ethicon), Polysorb (Medtronic), Novosyn (B. Braun)—, always checking USP 0 calibre, 36 mm needle size in ½ circle, trocar point, 90 cm thread length and violet colour. The cited brands belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
USP 0 calibre (metric 3.5) is indicated for tissues with moderate-medium tension: fascia, aponeurosis, abdominal wall, subcutaneous tissue and ligatures of medium-calibre vessels. The ½ circle trocar (tapercut) HRT needle is the appropriate choice for firm tissues where clean penetration without tearing is required; if you need a fully atraumatic point for soft tissues, consider the conical point (HR) versions of the same PGA family.
| Reference | Variant |
|---|---|
| AMP113HRT3610 | PGA Prime USP 0, HRT36 needle trocar point, violet, 90 cm (current product) |
| AMP113HRT3601 | PGA Prime USP 1, HRT36 tapercut needle, violet, 90 cm (larger gauge) |
| AMP113HRT3620 | PGA Prime USP 2/0, HRT36 trocar point needle, violet, 90 cm (smaller gauge) |
| AMP113HRT3630 | PGA Prime USP 3/0, HRT36 trocar point needle, violet, 90 cm (fine gauge) |
| AMP113HR3610 | PGA Prime USP 0, HR36 needle conical point, violet, 90 cm (atraumatic point) |
| AM113HRT3610 | PGA normal quality USP 0, HRT36 needle trocar point, violet, 90 cm (standard range) |
| AMP113HRN3610H | PGA Prime Heavy USP 0, HRN36 needle blunt point, violet, 90 cm (maximum atraumatic in friable tissues) |
| P1L13HRT3610 | PGLA Prime USP 0, HRT36 needle trocar point, violet, 90 cm (absorption 56-72 days) |
| 1L13HRT3610 | PGLA normal quality USP 0, HRT36 needle trocar point, violet, 90 cm (absorption 56-72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare facility. If you need longer strength retention (180-210 days), consider the polydioxanone (PDO) family. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, and keep the reference together with the batch number for traceability in the operating theatre. For B2B volume conditions or equivalences from other commercial houses, contact the technical-commercial team at Cahnos.





Reviews