Description
The PGA Prime Multifilament Absorbable Suture from Cahnos is the choice of the professional who demands a resorbable thread with predictable handling, secure knotting and consistent performance in the operating theatre. This configuration combines a braided strand of 100% synthetic polyglycolic acid (PGA), violet, fluted and 75 cm, with a ½-circle needle with cylindrical body and Taperpoint conical point of 17 mm (HR17 profile): the ideal option for soft tissue approximation, ligatures and general surgery, with an excellent value-for-money ratio for hospital replenishment, purchasing centres and B2B tenders. Presentation in a box of 12 units with a minimum order of 30 boxes.
Technical product explanation
Polyglycolic acid (PGA) is a 100% synthetic polymer absorbable by hydrolysis, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating regularises the surface of the strand: it reduces friction when passing through tissues (less saw effect), improves glide, minimises tissue reactivity and facilitates firm and secure knot tying, even in fine gauges such as USP 5/0.
Its degradation profile is designed for short-to-medium duration healing: it retains 75% of tensile strength at 2 weeks, 50% at 3 weeks and 40% at 4 weeks, completing absorption by hydrolysis between 60 and 90 days, independently of the patient’s enzymatic activity. The HR17 needle has a ½-circle (160°) curvature with a cylindrical body and Taperpoint conical point: it passes through soft tissues atraumatically, without aggressive cutting effect, avoiding enlargement of the exit hole. The violet colour provides optimal visibility of the strand against the surgical field, highly valued in deep planes and fine gauges. Product sterile by ethylene oxide (EO), single-use and not re-sterilisable.
Technical specifications
Technical data sheet for the reference (Ref. P112HR1750 · SKU AMP112HR1750)
| Parameter | Specification |
|---|---|
| Factory reference | P112HR1750 (official PGA Prime nomenclature) |
| Reference Cahnos (SKU) | AMP112HR1750 (box of 12 units) |
| Brand / quality | Prime (premium range) |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided coated multifilament, fluted finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 5/0 (metric 1 / EP 1), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.10 – 0.149 mm |
| Thread colour | Violet (colourless version available in the family) |
| Thread length | 75 cm (free end, without loop) |
| Needle shape | ½ circle (1/2 circular, 160°), cylindrical body |
| Needle tip | Taperpoint conical (atraumatic) |
| Needle size | HR 17 (17 mm), 1 needle per suture |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 60 and 90 days |
| Tensile strength retention | 75 % at 2 weeks / 50 % at 3 weeks / 40 % at 4 weeks |
| Sterilisation | Ethylene oxide (EO); single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets, batch traceability) |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; medical device with CE marking |
Absorption and tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Complete absorption | 60 – 90 days by hydrolysis (predictable and uniform rate) |
The profile cited corresponds to the standard absorption formulation of the PGA line (60–90 days); it should not be confused with rapid-absorption PGA (RAPID line, 40–45 days) nor with standard polyglactin 910 (56–72 days).
Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 6/0 | 0.7 | 0.07 – 0.099 mm |
| 5/0 (this product) | 1 | 0.10 – 0.149 mm |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
Equivalences and cross-references
Coated braided multifilament polyglycolic acid (PGA) is the most widespread medium-duration synthetic absorbable in the surgical market, with an absorption profile of 60 to 90 days and a coating of polycaprolactone and calcium stearate. For the verified configuration of this datasheet (½-circle cylindrical needle with conical point of 17 mm, USP gauge 5/0, thread 75 cm) only the reference whose code and attributes have been checked against the manufacturer’s official catalogue is included:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Italian brand | TRUGLYDE (braided and coated PGA) | GM22972 – USP 5/0, ½ circle needle with round body of 17 mm, colourless suture of 75 cm | Comparable by category (not interchangeable 1:1) | They match in material (coated braided multifilament PGA), USP gauge 5/0 (metric 1), needle geometry (½-circle, cylindrical body, 17 mm), length of 75 cm and standard absorption profile (70–90 days); they differ in colour (undyed vs violet), commercial coating and presentation format. Code checked against the brand’s official catalogue. |
Homologous families in the 100% coated braided PGA segment include Safil (B. Braun), Surgicryl PGA (SMI), Dexon II (Medtronic/Covidien), DemeSORB PGA (DemeTECH), Assucryl (Assut) and Atramat (Resorba); as polyglactin 910 (PGLA) lines comparable by category —chemically different, absorption 56–72 days—, Coated VICRYL / VICRYL Plus (Ethicon), Novosyn (B. Braun) and Polysorb (Medtronic). In all cases, property-by-property verification must be carried out —exact material, USP gauge 5/0, multifilament structure, ½-circle curvature, conical point family, needle length of 17 mm, strand of 75 cm, colour, 1-needle attachment and CE/MDR regulatory classification— before any substitution; unverified cross-reference codes are not published. There is no equivalence table validated by the manufacturer that cross-references this reference with other brands in the exact configuration; consult with the technical department of Cahnos before use in tenders. Within the same system, the PGA Prime family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are fully interchangeable with each other respecting gauge and needle geometry. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
Selection tips and variants of interest
USP gauge 5/0 (metric 1) is a fine size indicated for low-tension tissues and rapid healing: fine skin, mucous membranes, subcutaneous tissue, superficial planes and ligatures of small-calibre vessels in general surgery and minor surgery. The HR17 ½-circle needle with Taperpoint conical point combines the sharp entry of the point with the non-traumatic passage of the cylindrical body: it is the recommended option for reduced surgical fields and soft tissues, with atraumatic passage, without broad cutting effect and with minimal bleeding. The violet colour offers excellent visibility of the strand against the surgical field; if maximum discretion is preferred, request the undyed version of the same family.
| Variant (SKU) | Description |
|---|---|
| AMP112HR1750 (P112HR1750) | PGA Prime – HR17 taperpoint – USP 5/0 violet – 75 cm (this product) |
| AM112HR1750 (112HR1750) | Standard PGA – HR17 taperpoint – USP 5/0 violet – 75 cm (same configuration, standard range; direct equivalent within the same manufacturer) |
| AMP112HR1740 | PGA Prime – HR17 taperpoint – USP 4/0 violet – 75 cm (immediately larger gauge) |
| AMP112HR1730 | PGA Prime – HR17 taperpoint – USP 3/0 violet – 75 cm (gauge for higher tension) |
| AMP112HR1720 | PGA Prime – HR17 taperpoint – USP 2/0 violet – 75 cm (higher strength) |
| AMP128HR1750 | PGA Prime – HR17 taperpoint – USP 5/0 undyed – 70 cm (same geometry, undyed thread) |
| P1L12HR1750 | PGLA polyglactin 910 – HR17 taperpoint – USP 5/0 violet – 75 cm (absorption 56–72 days) |
All references are supplied in boxes of 12 units with a minimum order of 30 boxes. If the procedure requires longer wound support, consider the polydioxanone (PDO, 180–210 days) or poliglecaprone PGCL (90–120 days) family; if a shorter absorption profile is required, consider PGA RAPID (40–45 days). Sterility and storage: sterile product by ethylene oxide and single-use; check the integrity of the pouch and the expiry date before each procedure, keep the reference P112HR1750 together with the batch number for traceability in the operating theatre and store below 30 °C, in a cool, dry place, protected from moisture and direct sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalence with other commercial brands, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.